Extended-focus second-harmonic generation microscopy of biological tissues using a two-axicon Bessel-beam shaper

Conventional Gaussian-beam second-harmonic generation (SHG) microscopy requires time-consuming axial scanning to image across tissue depth. Here, we integrate a two-axicon Bessel-beam shaper into a stage-scanning SHG microscope and compare imaging across 2° and 5° axicon pairs. Characterization of SHG from a quartz slab gives depths of focus (DOFs) that are 2.37 and 1.56 times the Gaussian DOF, with continuous tuning via inter-axicon translation. Evaluation across tendon and zebrafish specimens quantifies structural visibility and contrast. Bessel-beam performance was further tested on week-10 mouse mammary glands, where a single Bessel image captures collagen structures also visible in a z-projection of 15 Gaussian planes (∼28 µm span). Relative to this protocol, one Bessel acquisition results in an estimated 15-fold speedup in acquisition time. Additional equal-input-power quartz measurements and temporal image analyses quantify the signal-to-noise trade-offs associated with extended-focus imaging.

Authors

Institutions

Publication Details

Journal
Biomedical Optics Express
Published
2026-11-25
DOI
https://doi.org/10.1364/boe.612810
Primary Topic
Advanced Fluorescence Microscopy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Extended-focus second-harmonic generation microscopy of biological tissues using a two-axicon Bessel-beam shaper

Arash Aghigh, Isabelle Plante, Jaskaran Singh, Shunmoogum A. Patten et al.
1 links
Biomedical Optics Express
Advanced Fluorescence Microscopy Techniques
article

Extended-focus second-harmonic generation microscopy of biological tissues using a two-axicon Bessel-beam shaper

Arash Aghigh, Isabelle Plante, Jaskaran Singh, Shunmoogum A. Patten, François Légaré, Jysiane Cardot, Bhawna Bhawna, Fariborz Ehteshami, Eva Gierski, Cameron Brown
article en

Abstract

Conventional Gaussian-beam second-harmonic generation (SHG) microscopy requires time-consuming axial scanning to image across tissue depth. Here, we integrate a two-axicon Bessel-beam shaper into a stage-scanning SHG microscope and compare imaging across 2° and 5° axicon pairs. Characterization of SHG from a quartz slab gives depths of focus (DOFs) that are 2.37 and 1.56 times the Gaussian DOF, with continuous tuning via inter-axicon translation. Evaluation across tendon and zebrafish specimens quantifies structural visibility and contrast. Bessel-beam performance was further tested on week-10 mouse mammary glands, where a single Bessel image captures collagen structures also visible in a z-projection of 15 Gaussian planes (∼28 µm span). Relative to this protocol, one Bessel acquisition results in an estimated 15-fold speedup in acquisition time. Additional equal-input-power quartz measurements and temporal image analyses quantify the signal-to-noise trade-offs associated with extended-focus imaging.

Biomedical Optics ExpressVol. 17(12)
Queensland University of Technology (AU), Armand Frappier Museum (CA)
Advanced Fluorescence Microscopy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon