Analytical Performance of the Sysmex XF ‐1600 Flow Cytometer in Routine Clinical Practice

INTRODUCTION: Flow cytometry systems are increasingly used for the identification and quantification of lymphocyte subpopulations, supporting the diagnosis and monitoring of hematological and immunological conditions. This study aimed to evaluate the analytical performance of the novel Sysmex XF-1600 flow cytometer for lymphocyte subpopulation quantification in peripheral blood, including precision, carryover, and agreement with established laboratory methods. METHODS: Peripheral blood samples from routine clinical workload were analysed in the hematology laboratory. Analytical performance was assessed by evaluating precision (repeatability and reproducibility) and carryover. Absolute lymphocyte and leukocyte counts obtained using the XF-1600 automated sample preparation and volumetric counting system were compared with the Sysmex XN-series hematology analyser, while lymphocyte subset quantification was compared with the Navios flow cytometer (Beckman Coulter). RESULTS: The XF-1600 showed coefficients of variation below 6% for major lymphocyte populations, including CD3+, CD4+, CD8+, CD19+, and CD16+/CD56+ cells. No carryover was detected. Comparison with the Sysmex XN analyser showed strong correlation for leukocyte and lymphocyte counts (Spearman's ρ = 0.986 and 0.993, respectively). Comparison with the Navios flow cytometer demonstrated high correlation across lymphocyte subpopulations (Spearman's ρ > 0.9 for most parameters). Bland-Altman and Passing-Bablok analyses showed good agreement between methods, although some parameters were not fully interchangeable. CONCLUSION: The Sysmex XF-1600 flow cytometer demonstrated good analytical performance for lymphocyte subset quantification, with high precision and strong agreement with routinely used laboratory methods. The integration of automated sample preparation and volumetric counting may support standardization of immunophenotyping workflows in clinical laboratories.

Authors

Institutions

Publication Details

Journal
International Journal of Laboratory Hematology
Published
2026-10-26
DOI
https://doi.org/10.1111/ijlh.70267
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Analytical Performance of the Sysmex XF ‐1600 Flow Cytometer in Routine Clinical Practice

Natàlia Lloveras, Maite Serrando, Meritxell Deulofeu, Leire Saiz Sierra et al.
International Journal of Laboratory Hematology
Clinical Laboratory Practices and Quality Control
article

Analytical Performance of the Sysmex XF ‐1600 Flow Cytometer in Routine Clinical Practice

Natàlia Lloveras, Maite Serrando, Meritxell Deulofeu, Leire Saiz Sierra, Anna Marull, Orlando José Jimenez-Jarava, David Cruz García
article en

Abstract

INTRODUCTION: Flow cytometry systems are increasingly used for the identification and quantification of lymphocyte subpopulations, supporting the diagnosis and monitoring of hematological and immunological conditions. This study aimed to evaluate the analytical performance of the novel Sysmex XF-1600 flow cytometer for lymphocyte subpopulation quantification in peripheral blood, including precision, carryover, and agreement with established laboratory methods. METHODS: Peripheral blood samples from routine clinical workload were analysed in the hematology laboratory. Analytical performance was assessed by evaluating precision (repeatability and reproducibility) and carryover. Absolute lymphocyte and leukocyte counts obtained using the XF-1600 automated sample preparation and volumetric counting system were compared with the Sysmex XN-series hematology analyser, while lymphocyte subset quantification was compared with the Navios flow cytometer (Beckman Coulter). RESULTS: The XF-1600 showed coefficients of variation below 6% for major lymphocyte populations, including CD3+, CD4+, CD8+, CD19+, and CD16+/CD56+ cells. No carryover was detected. Comparison with the Sysmex XN analyser showed strong correlation for leukocyte and lymphocyte counts (Spearman's ρ = 0.986 and 0.993, respectively). Comparison with the Navios flow cytometer demonstrated high correlation across lymphocyte subpopulations (Spearman's ρ > 0.9 for most parameters). Bland-Altman and Passing-Bablok analyses showed good agreement between methods, although some parameters were not fully interchangeable. CONCLUSION: The Sysmex XF-1600 flow cytometer demonstrated good analytical performance for lymphocyte subset quantification, with high precision and strong agreement with routinely used laboratory methods. The integration of automated sample preparation and volumetric counting may support standardization of immunophenotyping workflows in clinical laboratories.

International Journal of Laboratory Hematology
Universitat de Girona (ES), Institut Català d'Oncologia (ES), Hospital Universitari de Girona Doctor Josep Trueta (ES)
Openalex Percentile: Top 13%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon