Reel Syndrome in Subcutaneous Implantable Cardioverter-Defibrillator Causing Lead Dislodgement and Inappropriate Shock: Successful Rescue with Intermuscular Two-Incision Technique and Serratus Anterior Plane Block

Background: Reel syndrome, rotation of the generator on its transverse axis with coiling of the lead, is a rare cause of cardiac implantable electronic device failure. While described for transvenous systems, it is exceptionally rare in subcutaneous ICDs (S-ICD). Case presentation: A 63-year-old woman with dilated cardiomyopathy and S-ICD implanted in 2022 for secondary prevention in a standard subcutaneous pocket presented with a shock during exertion. Remote monitoring (LATITUDE) had not been activated. Device interrogation showed an inappropriate 80J shock due to myopotential oversensing after caudal retraction of the lead. Chest X-ray documented Reel syndrome with proximal coiling around the can and distal tip retracted to the subcostal region, >10 cm from the manubrio-sternal junction, with horizontalization of the shock coil, incompatible with effective defibrillation. Management: The system was explanted and a new S-ICD re-implanted using the intermuscular two-incision technique in the virtual space between serratus anterior and latissimus dorsi. The procedure was performed under ultrasound-guided serratus anterior plane block combined with parasternal block. The generator was anchored to the fascia with non-absorbable sutures. Defibrillation test was effective at 40J. Remote monitoring was activated. Conclusion: To our knowledge, this is one of the few documented cases of Reel syndrome in S-ICD with both radiographic and stored electrogram evidence of inappropriate 80J shock, rescued with intermuscular two-incision technique under ultrasound-guided serratus anterior plane block. Intermuscular fixation with non-absorbable sutures, combined with regional anesthesia and systematic remote monitoring, ensures device stability and maintains continuous defibrillation protection.

Authors

Publication Details

Journal
Preprints.org
Published
2026-10-28
DOI
https://doi.org/10.20944/preprints202610.2816.v1
Primary Topic
Cardiac pacing and defibrillation studies
Type
preprint
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
preprint

Reel Syndrome in Subcutaneous Implantable Cardioverter-Defibrillator Causing Lead Dislodgement and Inappropriate Shock: Successful Rescue with Intermuscular Two-Incision Technique and Serratus Anterior Plane Block

Carlo Uran, Ambra Uccello
1 links
Preprints.org
Cardiac pacing and defibrillation studies
preprint

Reel Syndrome in Subcutaneous Implantable Cardioverter-Defibrillator Causing Lead Dislodgement and Inappropriate Shock: Successful Rescue with Intermuscular Two-Incision Technique and Serratus Anterior Plane Block

Carlo Uran, Ambra Uccello
preprint en

Abstract

Background: Reel syndrome, rotation of the generator on its transverse axis with coiling of the lead, is a rare cause of cardiac implantable electronic device failure. While described for transvenous systems, it is exceptionally rare in subcutaneous ICDs (S-ICD). Case presentation: A 63-year-old woman with dilated cardiomyopathy and S-ICD implanted in 2022 for secondary prevention in a standard subcutaneous pocket presented with a shock during exertion. Remote monitoring (LATITUDE) had not been activated. Device interrogation showed an inappropriate 80J shock due to myopotential oversensing after caudal retraction of the lead. Chest X-ray documented Reel syndrome with proximal coiling around the can and distal tip retracted to the subcostal region, >10 cm from the manubrio-sternal junction, with horizontalization of the shock coil, incompatible with effective defibrillation. Management: The system was explanted and a new S-ICD re-implanted using the intermuscular two-incision technique in the virtual space between serratus anterior and latissimus dorsi. The procedure was performed under ultrasound-guided serratus anterior plane block combined with parasternal block. The generator was anchored to the fascia with non-absorbable sutures. Defibrillation test was effective at 40J. Remote monitoring was activated. Conclusion: To our knowledge, this is one of the few documented cases of Reel syndrome in S-ICD with both radiographic and stored electrogram evidence of inappropriate 80J shock, rescued with intermuscular two-incision technique under ultrasound-guided serratus anterior plane block. Intermuscular fixation with non-absorbable sutures, combined with regional anesthesia and systematic remote monitoring, ensures device stability and maintains continuous defibrillation protection.

Preprints.org
Good health and well-being
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon