ICP-OES determination of lanthanum in polysiloxanes: comparison of direct organic introduction and single reaction chamber microwave digestion
, respectively. The relative standard deviations (RSDs) were below 3.4%, and the spiked recoveries ranged from 95.1% to 102.7%. Relative Bland-Altman analysis showed a small average difference between the two workflows for the four polysiloxane samples (mean relative difference: 0.75%). The overall AGREEprep scores of 0.47 and 0.20 for the DOI and SRC methods, respectively, indicate superior greenness for the DOI workflow. Featuring simpler operation, higher sample throughput, a lower validated MDL, and better environmental greenness, the DOI method is highly suitable for routine batch determinations of La in polysiloxanes. Conversely, SRC digestion serves as an effective complementary approach for cross-validation and for complex samples requiring organic matrix decomposition. Together, these two methods provide a reliable basis for selecting an appropriate sample preparation protocol.