Spot-wise compensation of proton beam deflection and radiation field deformation in the transverse magnetic field of an in-beam MR scanner using a research version of a commercial treatment planning system

Objective . Magnetic resonance-integrated proton therapy (MRiPT) offers the potential for high-precision, image-guided particle therapy but is challenged by magnetic field (MF)-induced proton beam deflection and radiation field deformation. Unless compensated for, these effects compromise dose conformity and prohibit clinical implementation. This work aims to develop and experimentally validate a spot-wise compensation method capable of restoring proton dose distributions in the presence of a measured three-dimensional magnetic vector field map within a clinical treatment planning system (TPS). Approach . A spot-wise compensation framework was implemented in a research version of RayStation, a TPS commercially available for clinical use. The method compensates both lateral beam deflection and MF-induced radiation field deformation by shifting individual pencil-beam spots based on displacement vectors derived from dose calculations in water phantoms using the measured MF of an in-beam MR scanner integrated with a horizontal proton beamline. Two complementary strategies were implemented: a patient-specific approach based on plan-specific displacement calculations, and a generic approach using precomputed displacement maps derived from a reference phantom. Compensation performance was evaluated in representative clinical cases. Dose distributions were experimentally validated using 2D ionisation chamber array measurements. Main results . Both compensation methods effectively restored target coverage, dose homogeneity, and organ-at-risk sparing to the level of no-MF reference plans. The patient-specific method generally achieved a higher performance, whereas the generic compensation method enabled rapid automated correction with sufficient accuracy for clinical workflows involving deep-seated tumours. Local γ -analysis demonstrated excellent agreement between measured and calculated doses, with pass rates exceeding 98.5% for 3%/3 mm criteria. Significance . This work demonstrates that compensation of MF-induced proton beam deflection and field deformation can be achieved entirely within a research version of the RayStation TPS using measured, MR scanner-specific MF data. The proposed method provides a scalable, robust framework for future MRiPT clinical applications.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2026-11-19
DOI
https://doi.org/10.1088/1361-6560/aeb84d
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Spot-wise compensation of proton beam deflection and radiation field deformation in the transverse magnetic field of an in-beam MR scanner using a research version of a commercial treatment planning system

Aswin Louis Hoffmann, Petra Trnková, Marisa Cobanaj, Franciska Lebbink et al.
1 links
Physics in Medicine and Biology
Radiation Therapy and Dosimetry
article

Spot-wise compensation of proton beam deflection and radiation field deformation in the transverse magnetic field of an in-beam MR scanner using a research version of a commercial treatment planning system

Aswin Louis Hoffmann, Petra Trnková, Marisa Cobanaj, Franciska Lebbink, Felix Horst, Stefan Menkel, Erik Tranéus, Krishna Godino Padre
article en

Abstract

Objective . Magnetic resonance-integrated proton therapy (MRiPT) offers the potential for high-precision, image-guided particle therapy but is challenged by magnetic field (MF)-induced proton beam deflection and radiation field deformation. Unless compensated for, these effects compromise dose conformity and prohibit clinical implementation. This work aims to develop and experimentally validate a spot-wise compensation method capable of restoring proton dose distributions in the presence of a measured three-dimensional magnetic vector field map within a clinical treatment planning system (TPS). Approach . A spot-wise compensation framework was implemented in a research version of RayStation, a TPS commercially available for clinical use. The method compensates both lateral beam deflection and MF-induced radiation field deformation by shifting individual pencil-beam spots based on displacement vectors derived from dose calculations in water phantoms using the measured MF of an in-beam MR scanner integrated with a horizontal proton beamline. Two complementary strategies were implemented: a patient-specific approach based on plan-specific displacement calculations, and a generic approach using precomputed displacement maps derived from a reference phantom. Compensation performance was evaluated in representative clinical cases. Dose distributions were experimentally validated using 2D ionisation chamber array measurements. Main results . Both compensation methods effectively restored target coverage, dose homogeneity, and organ-at-risk sparing to the level of no-MF reference plans. The patient-specific method generally achieved a higher performance, whereas the generic compensation method enabled rapid automated correction with sufficient accuracy for clinical workflows involving deep-seated tumours. Local γ -analysis demonstrated excellent agreement between measured and calculated doses, with pass rates exceeding 98.5% for 3%/3 mm criteria. Significance . This work demonstrates that compensation of MF-induced proton beam deflection and field deformation can be achieved entirely within a research version of the RayStation TPS using measured, MR scanner-specific MF data. The proposed method provides a scalable, robust framework for future MRiPT clinical applications.

Physics in Medicine and BiologyVol. 71(22)
Helmholtz-Zentrum Dresden-Rossendorf (DE), National Center for Tumor Diseases (DE), OncoRay (DE), RaySearch Laboratories (Sweden) (SE), Czech Technical University in Prague (CZ), Technische Universität Dresden (DE)
Helmholtz-Zentrum Dresden-Rossendorf
Good health and well-being
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon