Spot-wise compensation of proton beam deflection and radiation field deformation in the transverse magnetic field of an in-beam MR scanner using a research version of a commercial treatment planning system
Objective . Magnetic resonance-integrated proton therapy (MRiPT) offers the potential for high-precision, image-guided particle therapy but is challenged by magnetic field (MF)-induced proton beam deflection and radiation field deformation. Unless compensated for, these effects compromise dose conformity and prohibit clinical implementation. This work aims to develop and experimentally validate a spot-wise compensation method capable of restoring proton dose distributions in the presence of a measured three-dimensional magnetic vector field map within a clinical treatment planning system (TPS). Approach . A spot-wise compensation framework was implemented in a research version of RayStation, a TPS commercially available for clinical use. The method compensates both lateral beam deflection and MF-induced radiation field deformation by shifting individual pencil-beam spots based on displacement vectors derived from dose calculations in water phantoms using the measured MF of an in-beam MR scanner integrated with a horizontal proton beamline. Two complementary strategies were implemented: a patient-specific approach based on plan-specific displacement calculations, and a generic approach using precomputed displacement maps derived from a reference phantom. Compensation performance was evaluated in representative clinical cases. Dose distributions were experimentally validated using 2D ionisation chamber array measurements. Main results . Both compensation methods effectively restored target coverage, dose homogeneity, and organ-at-risk sparing to the level of no-MF reference plans. The patient-specific method generally achieved a higher performance, whereas the generic compensation method enabled rapid automated correction with sufficient accuracy for clinical workflows involving deep-seated tumours. Local γ -analysis demonstrated excellent agreement between measured and calculated doses, with pass rates exceeding 98.5% for 3%/3 mm criteria. Significance . This work demonstrates that compensation of MF-induced proton beam deflection and field deformation can be achieved entirely within a research version of the RayStation TPS using measured, MR scanner-specific MF data. The proposed method provides a scalable, robust framework for future MRiPT clinical applications.
Authors
- Aswin Louis Hoffmann (ORCID: https://orcid.org/0000-0002-5821-3135)
- Petra Trnková (ORCID: https://orcid.org/0000-0002-0347-2981)
- Marisa Cobanaj (ORCID: https://orcid.org/0000-0003-2904-267X)
- Franciska Lebbink (ORCID: https://orcid.org/0000-0001-6555-9487)
- Felix Horst (ORCID: https://orcid.org/0000-0003-0707-0856)
- Stefan Menkel
- Erik Tranéus (ORCID: https://orcid.org/0000-0002-1850-7382)
- Krishna Godino Padre
Institutions
- Helmholtz-Zentrum Dresden-Rossendorf (DE)
- National Center for Tumor Diseases (DE)
- OncoRay (DE)
- RaySearch Laboratories (Sweden) (SE)
- Czech Technical University in Prague (CZ)
- Technische Universität Dresden (DE)
Publication Details
- Journal
- Physics in Medicine and Biology
- Published
- 2026-11-19
- DOI
- https://doi.org/10.1088/1361-6560/aeb84d
- Primary Topic
- Radiation Therapy and Dosimetry
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Helmholtz-Zentrum Dresden-Rossendorf