Twenty-Year Wear Progression and Long-Term Outcomes of Conventional Polyethylene Liners After Cementless Total Hip Arthroplasty

Background Polyethylene (PE) wear debris can cause periprosthetic osteolysis and late failure after total hip arthroplasty, but long-term wear behavior of conventional non–cross-linked PE remains unclear. Methods We reviewed 85 consecutive cementless total hip arthroplasties performed between 2002 and 2005 using ArCom non–cross-linked PE liners (Biomet Inc., Warsaw, IN). After excluding 31 hips lost to follow-up and 5 revised hips, 49 hips (45 patients) with ≥20 years of follow-up were analyzed. Femoral head penetration was measured using PolyWare (Draftware Inc., Vevay, IN) by three independent observers. The first postoperative year was defined as the bedding-in phase, and subsequent penetration as the steady-state phase (SSP). Associations between wear and patient, implant, surgical, and clinical factors were assessed. Results The mean Japanese Orthopaedic Association score improved from 39.6 ± 10.6 preoperatively to 84.4 ± 12.9 at 20 years ( P < .001). Osteolysis was observed in 18 hips. Mean bedding-in phase and SSP wear rates were 0.235 ± 0.155 mm/y and 0.031 ± 0.023 mm/y, respectively. SSP wear correlated linearly with postoperative time (R 2 = 0.99). Hips with higher SSP wear (>0.025 mm/y) had higher 20-year Japanese Orthopaedic Association gait ( P = .033), activities of daily living ( P = .018), and total scores ( P = .013). Charnley category A had higher SSP wear than category B ( P = .024), whereas osteolysis did not affect wear rates. Conclusions ArCom conventional PE showed a nearly constant SSP wear rate over 20 years. Exploratory analyses suggested associations with functional status and Charnley category, but not with other patient, implant, or surgical factors.

Authors

Institutions

Publication Details

Journal
Arthroplasty Today
Published
2026-10-22
DOI
https://doi.org/10.1016/j.artd.2026.102187
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Twenty-Year Wear Progression and Long-Term Outcomes of Conventional Polyethylene Liners After Cementless Total Hip Arthroplasty

Hisatoshi Ishikura, Takeyuki Tanaka, 徹 茂呂, Sakae Tanaka et al.
Arthroplasty Today
Orthopaedic implants and arthroplasty
article

Twenty-Year Wear Progression and Long-Term Outcomes of Conventional Polyethylene Liners After Cementless Total Hip Arthroplasty

Hisatoshi Ishikura, Takeyuki Tanaka, 徹 茂呂, Sakae Tanaka, Hirofumi Oshima, Yasutaka Watanabe
article en

Abstract

Background Polyethylene (PE) wear debris can cause periprosthetic osteolysis and late failure after total hip arthroplasty, but long-term wear behavior of conventional non–cross-linked PE remains unclear. Methods We reviewed 85 consecutive cementless total hip arthroplasties performed between 2002 and 2005 using ArCom non–cross-linked PE liners (Biomet Inc., Warsaw, IN). After excluding 31 hips lost to follow-up and 5 revised hips, 49 hips (45 patients) with ≥20 years of follow-up were analyzed. Femoral head penetration was measured using PolyWare (Draftware Inc., Vevay, IN) by three independent observers. The first postoperative year was defined as the bedding-in phase, and subsequent penetration as the steady-state phase (SSP). Associations between wear and patient, implant, surgical, and clinical factors were assessed. Results The mean Japanese Orthopaedic Association score improved from 39.6 ± 10.6 preoperatively to 84.4 ± 12.9 at 20 years ( P < .001). Osteolysis was observed in 18 hips. Mean bedding-in phase and SSP wear rates were 0.235 ± 0.155 mm/y and 0.031 ± 0.023 mm/y, respectively. SSP wear correlated linearly with postoperative time (R 2 = 0.99). Hips with higher SSP wear (>0.025 mm/y) had higher 20-year Japanese Orthopaedic Association gait ( P = .033), activities of daily living ( P = .018), and total scores ( P = .013). Charnley category A had higher SSP wear than category B ( P = .024), whereas osteolysis did not affect wear rates. Conclusions ArCom conventional PE showed a nearly constant SSP wear rate over 20 years. Exploratory analyses suggested associations with functional status and Charnley category, but not with other patient, implant, or surgical factors.

Arthroplasty TodayVol. 42
NTT Medical Center (JP), The University of Tokyo (JP)
Openalex Percentile: Top 10%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon