Intraprocedural Embolization of a Tricuspid Edge-to-Edge Repair Device and Its Percutaneous Retrieval: A Case Report

Background and Clinical Significance: Tricuspid transcatheter edge-to-edge repair (T-TEER) is an established therapy for severe tricuspid regurgitation (TR) in selected patients. Single-leaflet device attachment (SLDA) is the commonest device-related complication, but complete intraprocedural detachment with embolization is rare, and percutaneous retrieval of an embolized tricuspid device has been reported only exceptionally. Case Presentation: An inoperable elderly patient with previous transcatheter aortic valve implantation (TAVI) and torrential functional TR underwent T-TEER. After implantation of the first clip at the anteroseptal commissure with apparently adequate leaflet capture, anterior SLDA was identified and progressed to complete detachment and right atrial embolization of the device. Despite this, the T-TEER was completed by successful implantation of the second clip, leaving only mild residual central TR. In the same session, the embolized device was retrieved percutaneously from the femoral venous system using a 15 mm gooseneck snare advanced through a 22 Fr sheath, with a safety wire and en-bloc removal, without forcing ensheathing and without venous injury. The patient was discharged on post-procedure day 2 with durable TR reduction and clinical improvement. Conclusions: Complete leaflet detachment is a possible complication of T-TEER. Confirming deep, symmetric insertion in multiple echocardiographic windows before every release is essential. When embolization occurs, knowledge of device dimensions and immediate availability of a complete snare kit is important. Although this approach may be considered in selected patients, particularly those at high surgical risk, the available evidence is limited to isolated case reports and does not allow general conclusions on its safety or efficacy.

Authors

Institutions

Publication Details

Journal
Reports — Medical Cases Images and Videos
Published
2026-10-16
DOI
https://doi.org/10.3390/reports9040359
Primary Topic
Cardiac Valve Diseases and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Intraprocedural Embolization of a Tricuspid Edge-to-Edge Repair Device and Its Percutaneous Retrieval: A Case Report

Frane Runjić, Andrija Matetić, Darija Baković, Ivona Mustapić
Reports — Medical Cases Images and Videos
Cardiac Valve Diseases and Treatments
article

Intraprocedural Embolization of a Tricuspid Edge-to-Edge Repair Device and Its Percutaneous Retrieval: A Case Report

Frane Runjić, Andrija Matetić, Darija Baković, Ivona Mustapić
article en

Abstract

Background and Clinical Significance: Tricuspid transcatheter edge-to-edge repair (T-TEER) is an established therapy for severe tricuspid regurgitation (TR) in selected patients. Single-leaflet device attachment (SLDA) is the commonest device-related complication, but complete intraprocedural detachment with embolization is rare, and percutaneous retrieval of an embolized tricuspid device has been reported only exceptionally. Case Presentation: An inoperable elderly patient with previous transcatheter aortic valve implantation (TAVI) and torrential functional TR underwent T-TEER. After implantation of the first clip at the anteroseptal commissure with apparently adequate leaflet capture, anterior SLDA was identified and progressed to complete detachment and right atrial embolization of the device. Despite this, the T-TEER was completed by successful implantation of the second clip, leaving only mild residual central TR. In the same session, the embolized device was retrieved percutaneously from the femoral venous system using a 15 mm gooseneck snare advanced through a 22 Fr sheath, with a safety wire and en-bloc removal, without forcing ensheathing and without venous injury. The patient was discharged on post-procedure day 2 with durable TR reduction and clinical improvement. Conclusions: Complete leaflet detachment is a possible complication of T-TEER. Confirming deep, symmetric insertion in multiple echocardiographic windows before every release is essential. When embolization occurs, knowledge of device dimensions and immediate availability of a complete snare kit is important. Although this approach may be considered in selected patients, particularly those at high surgical risk, the available evidence is limited to isolated case reports and does not allow general conclusions on its safety or efficacy.

Reports — Medical Cases Images and VideosVol. 9(4)
London School of Economics and Political Science (GB), University of Split (HR)
Good health and well-being
Openalex Percentile: Top 12%
Cardiac Valve Diseases and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.