Next-generation digital polymerase chain reaction assay for BCR::ABL1 in chronic myeloid leukemia: comprehensive validation for standardized measurable residual disease detection

PURPOSE: The sensitivity of real-time quantitative reverse-transcription polymerase chain reaction (RT-qPCR) for detecting measurable residual disease (MRD) in patients with chronic myeloid leukemia (CML) who have achieved a deep molecular response is limited. We validated a digital PCR (dPCR) assay (LQ Check BCR::ABL1 Major %IS Kit) for MRD quantification. METHODS: Validation was performed according to CLSI guidelines EP17-A2, EP06, EP05-A3, and EP07 using BCR::ABL1-positive and -negative materials and the World Health Organization (WHO) 1st International Standard. A nonparametric analysis of the limit of blank was used to establish a positivity cutoff of ≥ 2 droplets. The limit of detection (LoD) was verified using a detection criterion of ≥ 95%, and the limit of quantitation (LoQ) was defined as the lowest level meeting ≥ 95% detection and a study-defined %CV of ≤ 76%. Clinical validation included 72 CML specimens for molecular response (MR) category comparisons and 67 specimens for paired quantitative comparisons. RESULTS: = 0.9916-0.9972), within-run %CV ≤ 1.52%, cross-condition %CV ≤ 2.76%, and WHO-traceable accuracy. In 67 paired specimens, the correlation coefficient was r = 0.9740, with a Passing-Bablok slope of 1.0189, intercept of -0.2552, and Bland-Altman mean bias of 0.2266. Among the 72 specimens, dPCR classified 20 as having a deeper MR than RT-qPCR. Reproducibility was characterized by r = 0.99 and a mean bias of 0.0967. Whole blood was preferably processed within 48 h, with processing at 72 h considered acceptable only when BCR QC criteria were met. CONCLUSION: The LQ Check BCR::ABL1 Major %IS Kit enabled sensitive, precise, and reproducible BCR::ABL1 quantification at MR4.6. The clinical significance of a deeper MR classification using dPCR, particularly for treatment-free remission, requires longitudinal validation.

Authors

Institutions

Publication Details

Journal
Blood Research
Published
2026-10-16
DOI
https://doi.org/10.1007/s44313-026-00188-5
Primary Topic
Chronic Myeloid Leukemia Treatments
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Next-generation digital polymerase chain reaction assay for BCR::ABL1 in chronic myeloid leukemia: comprehensive validation for standardized measurable residual disease detection

Kyung‐Mi Kee, Soohyun Kim, Se‐Heon Kim, Yong-Hwi Kim et al.
Blood Research
Chronic Myeloid Leukemia Treatments
article

Next-generation digital polymerase chain reaction assay for BCR::ABL1 in chronic myeloid leukemia: comprehensive validation for standardized measurable residual disease detection

Kyung‐Mi Kee, Soohyun Kim, Se‐Heon Kim, Yong-Hwi Kim, Bong-Suk Kim, In-Young Roh, Jae-Hoon Kim, Min-Sik Jang, Dong-Wook Kim
article en

Abstract

PURPOSE: The sensitivity of real-time quantitative reverse-transcription polymerase chain reaction (RT-qPCR) for detecting measurable residual disease (MRD) in patients with chronic myeloid leukemia (CML) who have achieved a deep molecular response is limited. We validated a digital PCR (dPCR) assay (LQ Check BCR::ABL1 Major %IS Kit) for MRD quantification. METHODS: Validation was performed according to CLSI guidelines EP17-A2, EP06, EP05-A3, and EP07 using BCR::ABL1-positive and -negative materials and the World Health Organization (WHO) 1st International Standard. A nonparametric analysis of the limit of blank was used to establish a positivity cutoff of ≥ 2 droplets. The limit of detection (LoD) was verified using a detection criterion of ≥ 95%, and the limit of quantitation (LoQ) was defined as the lowest level meeting ≥ 95% detection and a study-defined %CV of ≤ 76%. Clinical validation included 72 CML specimens for molecular response (MR) category comparisons and 67 specimens for paired quantitative comparisons. RESULTS: = 0.9916-0.9972), within-run %CV ≤ 1.52%, cross-condition %CV ≤ 2.76%, and WHO-traceable accuracy. In 67 paired specimens, the correlation coefficient was r = 0.9740, with a Passing-Bablok slope of 1.0189, intercept of -0.2552, and Bland-Altman mean bias of 0.2266. Among the 72 specimens, dPCR classified 20 as having a deeper MR than RT-qPCR. Reproducibility was characterized by r = 0.99 and a mean bias of 0.0967. Whole blood was preferably processed within 48 h, with processing at 72 h considered acceptable only when BCR QC criteria were met. CONCLUSION: The LQ Check BCR::ABL1 Major %IS Kit enabled sensitive, precise, and reproducible BCR::ABL1 quantification at MR4.6. The clinical significance of a deeper MR classification using dPCR, particularly for treatment-free remission, requires longitudinal validation.

Blood ResearchVol. 61(1)
Eulji University (KR)
Korea Medical Device Development Fund, Ministry of Trade, Industry and Energy, Ministry of Food and Drug Safety, Ministry of Science and ICT, South Korea
Good health and well-being
Openalex Percentile: Top 13%
Chronic Myeloid Leukemia Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.