White Blood Cell Counting in Severely Leukopenic Samples on a Sysmex XN Hematology Analyzer: Analytical Performance Evaluation of Whole Blood Mode and Low White Blood Cell Mode

Background: /L. Methods: Precision of WBC quantification in the two modes was evaluated using four samples according to the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) criterion (CV% 5.6). Linearity was evaluated at five concentrations. The two modes were compared in 650 leukopenic samples. Two samples showing discrepancies between the two modes were reviewed. Results: =0.90) were noted. Discrepant samples were interpreted as clinically consistent in WB and LW modes in one case each. Conclusions: /L and discrepancies between the two modes were rare, careful attention is recommended.

Authors

Institutions

Publication Details

Journal
Annals of Laboratory Medicine
Published
2026-10-16
DOI
https://doi.org/10.3343/alm.2026.0225
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

White Blood Cell Counting in Severely Leukopenic Samples on a Sysmex XN Hematology Analyzer: Analytical Performance Evaluation of Whole Blood Mode and Low White Blood Cell Mode

Hanah Kim, Jong-Ho Yi, Gun-Hyuk Lee, Mina Hur et al.
Annals of Laboratory Medicine
Clinical Laboratory Practices and Quality Control
article

White Blood Cell Counting in Severely Leukopenic Samples on a Sysmex XN Hematology Analyzer: Analytical Performance Evaluation of Whole Blood Mode and Low White Blood Cell Mode

Hanah Kim, Jong-Ho Yi, Gun-Hyuk Lee, Mina Hur, Hyejung Lee, SeungWan Kim
article en

Abstract

Background: /L. Methods: Precision of WBC quantification in the two modes was evaluated using four samples according to the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) criterion (CV% 5.6). Linearity was evaluated at five concentrations. The two modes were compared in 650 leukopenic samples. Two samples showing discrepancies between the two modes were reviewed. Results: =0.90) were noted. Discrepant samples were interpreted as clinically consistent in WB and LW modes in one case each. Conclusions: /L and discrepancies between the two modes were rare, careful attention is recommended.

Annals of Laboratory Medicine
Konkuk University (KR), Samsung Medical Center (KR), Konkuk University Medical Center (KR), Sungkyunkwan University (KR)
Good health and well-being
Openalex Percentile: Top 14%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.