Surface and intrapulpal temperature changes in teeth using different violet LED devices under continuous or pulsed irradiation protocols: an in vitro study

Abstract Purpose To evaluate intrapulpal and tooth surface temperature increases (ΔT) produced by different violet LED devices (405 nm) under continuous and pulsed irradiation protocols in an in vitro model. Methods Three human mandibular incisors were prepared and instrumented. A 35% hydrogen peroxide gel was applied to the enamel surface and replaced at each measurement. Three violet LED devices with different configurations were tested: Device #1 (300 mW/LED), Device #2 (375 mW/LED), and Device #3 (250 mW/LED). A 2 cm distance from the enamel surface was maintained, and two irradiation protocols were applied: continuous irradiation for 30 min and a pulsed protocol (1-minute irradiation followed by a 30-second pause, totaling 30 min). Temperature was recorded in real time, and temperature increase (ΔT, °C) relative to baseline was used for the analyses. Data were analyzed using peak temperature increase (ΔT) at each 5-minute interval over the 30-minute experimental period, employing two-way repeated-measures ANOVA (α = 5%). Results No measured intrapulpal temperature increase exceeded the historical 5.5 °C reference value. Device #2 frequently showed numerically higher temperature increases; however, pairwise differences between individual devices were not statistically significant after correction for multiple comparisons. Conclusion Within the limitations of this in vitro study, numerical differences in thermal behavior were observed among the tested violet LED devices. Under the tested experimental conditions, no measured intrapulpal temperature increase exceeded the historical 5.5 °C reference value. These findings describe the thermal behavior observed in the present in vitro model and should not be interpreted as evidence of clinical safety.

Authors

Institutions

Publication Details

Journal
Lasers in Dental Science
Published
2026-10-15
DOI
https://doi.org/10.1007/s41547-026-00362-z
Primary Topic
Laser Applications in Dentistry and Medicine
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Surface and intrapulpal temperature changes in teeth using different violet LED devices under continuous or pulsed irradiation protocols: an in vitro study

Anna Carolina Ratto Tempestini Horliana, Karen Müller Ramalho, Thaís Gimenez, Christian Ferreira Bernardi et al.
Lasers in Dental Science
Laser Applications in Dentistry and Medicine
article

Surface and intrapulpal temperature changes in teeth using different violet LED devices under continuous or pulsed irradiation protocols: an in vitro study

Anna Carolina Ratto Tempestini Horliana, Karen Müller Ramalho, Thaís Gimenez, Christian Ferreira Bernardi, Sophia Alcantara Ishikawa, Enzo Llobregat Ducatti
article en

Abstract

Abstract Purpose To evaluate intrapulpal and tooth surface temperature increases (ΔT) produced by different violet LED devices (405 nm) under continuous and pulsed irradiation protocols in an in vitro model. Methods Three human mandibular incisors were prepared and instrumented. A 35% hydrogen peroxide gel was applied to the enamel surface and replaced at each measurement. Three violet LED devices with different configurations were tested: Device #1 (300 mW/LED), Device #2 (375 mW/LED), and Device #3 (250 mW/LED). A 2 cm distance from the enamel surface was maintained, and two irradiation protocols were applied: continuous irradiation for 30 min and a pulsed protocol (1-minute irradiation followed by a 30-second pause, totaling 30 min). Temperature was recorded in real time, and temperature increase (ΔT, °C) relative to baseline was used for the analyses. Data were analyzed using peak temperature increase (ΔT) at each 5-minute interval over the 30-minute experimental period, employing two-way repeated-measures ANOVA (α = 5%). Results No measured intrapulpal temperature increase exceeded the historical 5.5 °C reference value. Device #2 frequently showed numerically higher temperature increases; however, pairwise differences between individual devices were not statistically significant after correction for multiple comparisons. Conclusion Within the limitations of this in vitro study, numerical differences in thermal behavior were observed among the tested violet LED devices. Under the tested experimental conditions, no measured intrapulpal temperature increase exceeded the historical 5.5 °C reference value. These findings describe the thermal behavior observed in the present in vitro model and should not be interpreted as evidence of clinical safety.

Lasers in Dental ScienceVol. 10(1)
Universidade de São Paulo (BR), Universidade Nove de Julho (BR)
Fundação de Amparo à Pesquisa do Estado de São Paulo
Good health and well-being
Openalex Percentile: Top 13%
Laser Applications in Dentistry and Medicine
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.