Patient and Database Shift in Lightweight ECG Arrhythmia Classification: A Controlled Evaluation of Noise, Quantization, and Embedded Inference

Efficient microcontroller execution does not ensure reliable electrocardiogram classification across patients and databases. We conducted a controlled benchmark of three convolutional architectures under a corrected subject-disjoint MIT-BIH split, external INCART testing, Noise Stress Test Database artifacts, 16-bit floating-point (FP16) and full-integer 8-bit (INT8) conversion, and classifier-only invocation on an ESP32 without external RAM. Five fixed seeds were used. On the MIT-BIH test cohort (DS2), INT8 four-class macro-F1 was 0.474-0.498 and supraventricular (S) recall was 0.150-0.180. Mean S-class area under the receiver operating characteristic curve (AUROC) was below 0.5 in all six architecture-regime groups, although all corresponding record-cluster intervals included 0.5. Within-record ranking could remain high despite poor pooled ranking. On INCART, S-class ranking was stronger, with all six patient-cluster AUROC intervals above 0.5, but the representative mixed-noise depthwise-separable network (DS-CNN) achieved 0.797 recall with 0.074 precision. Including record 201, which shares a subject with test record 202, increased DS2 S-recall point estimates by 2.49-14.68 percentage points. In the primary architecture panel, FP16 and INT8 met the engineering tolerances in all 120 comparisons. DS-CNN required 9.49 ms per classifier invocation on the ESP32 using preprocessed, annotation-centered inputs. Acquisition, QRS detection, preprocessing, current, and energy were not measured. Patient-dependent S-score behavior and external false-positive predictions limited this panel more than numerical precision or microcontroller resources.

Authors

Publication Details

Journal
Medical Engineering & Physics
Published
2026-10-15
DOI
https://doi.org/10.1088/1873-4030/aeb4bc
Primary Topic
ECG Monitoring and Analysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Patient and Database Shift in Lightweight ECG Arrhythmia Classification: A Controlled Evaluation of Noise, Quantization, and Embedded Inference

Woojin Cho, Hyun Mok Park, Seok-Oh Bang, Jin-Sung Roh et al.
Medical Engineering & Physics
ECG Monitoring and Analysis
article

Patient and Database Shift in Lightweight ECG Arrhythmia Classification: A Controlled Evaluation of Noise, Quantization, and Embedded Inference

Woojin Cho, Hyun Mok Park, Seok-Oh Bang, Jin-Sung Roh, Hyun-Seok Choi, Ki-Tae Kwon, Jong-Min Lim, Sang-Wuk Shin
article en

Abstract

Efficient microcontroller execution does not ensure reliable electrocardiogram classification across patients and databases. We conducted a controlled benchmark of three convolutional architectures under a corrected subject-disjoint MIT-BIH split, external INCART testing, Noise Stress Test Database artifacts, 16-bit floating-point (FP16) and full-integer 8-bit (INT8) conversion, and classifier-only invocation on an ESP32 without external RAM. Five fixed seeds were used. On the MIT-BIH test cohort (DS2), INT8 four-class macro-F1 was 0.474-0.498 and supraventricular (S) recall was 0.150-0.180. Mean S-class area under the receiver operating characteristic curve (AUROC) was below 0.5 in all six architecture-regime groups, although all corresponding record-cluster intervals included 0.5. Within-record ranking could remain high despite poor pooled ranking. On INCART, S-class ranking was stronger, with all six patient-cluster AUROC intervals above 0.5, but the representative mixed-noise depthwise-separable network (DS-CNN) achieved 0.797 recall with 0.074 precision. Including record 201, which shares a subject with test record 202, increased DS2 S-recall point estimates by 2.49-14.68 percentage points. In the primary architecture panel, FP16 and INT8 met the engineering tolerances in all 120 comparisons. DS-CNN required 9.49 ms per classifier invocation on the ESP32 using preprocessed, annotation-centered inputs. Acquisition, QRS detection, preprocessing, current, and energy were not measured. Patient-dependent S-score behavior and external false-positive predictions limited this panel more than numerical precision or microcontroller resources.

Medical Engineering & Physics
Good health and well-being
Openalex Percentile: Top 12%
ECG Monitoring and Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.