A passive thin-film mechanical device for visual respiratory monitoring during oxygen mask use: development and bench validation

Objective Continuous respiratory monitoring during transfer from the operating theatre to the post-anaesthesia care unit (PACU) is frequently unavailable despite this being a period of recognised patient vulnerability. Existing electronic respiratory monitors introduce cost, complexity and set-up that may limit use in fast-paced clinical environments. We developed Flow-Bag, a fully passive mechanical breath-indicating device consisting of a thin heat-sealed high-density polyethylene (HDPE) film bag that attaches externally to an oxygen mask. Methods Preliminary film-thickness screening was undertaken using a programmable breathing simulator (PEMDx Pty Ltd, Australia). A 6 µm (per ply) prototype was assessed across respiratory rates of 5-60 breaths min⁻¹ and tidal volumes of 50-600 mL. Three blinded observers positioned 5 m from the device counted visible inflation-deflation cycles over 60-s intervals, with counts compared against simulator-delivered breaths. Performance was additionally assessed across four commercially available oxygen mask models. Results In screening, the lowest detected tidal volume fell from 400 mL with 20 µm film to 150 mL with 6 µm film, the thinnest film that could be reproducibly manufactured using the processes available to us. The final prototype produced observable breath indications in 42 of 56 tested conditions (all tidal volumes ≥150 mL) but not at 50 or 100 mL. Within these detectable conditions, observer counts agreed 100% with simulator-delivered breaths. Performance was maintained across four oxygen mask models. However, sensitivity was reduced by incomplete sealing at either the mask-to-face or device-to-mask interface. Conclusion Flow-Bag is a fully passive mechanical breath indicator. Under controlled bench conditions using a mechanical breathing simulator, it provided reliable visual indication of simulated breathing across respiratory rates of 5-60 breaths min⁻¹ and tidal volumes of 150-600 mL. Subject to clinical validation and safety testing, this device may offer a potentially low-cost adjunct alongside standard monitoring. .

Authors

Institutions

Publication Details

Journal
Medical Engineering & Physics
Published
2026-10-15
DOI
https://doi.org/10.1088/1873-4030/aeb4bd
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A passive thin-film mechanical device for visual respiratory monitoring during oxygen mask use: development and bench validation

Mariam Rowe, Dylan Rowe, Phillip Melksham
Medical Engineering & Physics
Non-Invasive Vital Sign Monitoring
article

A passive thin-film mechanical device for visual respiratory monitoring during oxygen mask use: development and bench validation

Mariam Rowe, Dylan Rowe, Phillip Melksham
article en

Abstract

Objective Continuous respiratory monitoring during transfer from the operating theatre to the post-anaesthesia care unit (PACU) is frequently unavailable despite this being a period of recognised patient vulnerability. Existing electronic respiratory monitors introduce cost, complexity and set-up that may limit use in fast-paced clinical environments. We developed Flow-Bag, a fully passive mechanical breath-indicating device consisting of a thin heat-sealed high-density polyethylene (HDPE) film bag that attaches externally to an oxygen mask. Methods Preliminary film-thickness screening was undertaken using a programmable breathing simulator (PEMDx Pty Ltd, Australia). A 6 µm (per ply) prototype was assessed across respiratory rates of 5-60 breaths min⁻¹ and tidal volumes of 50-600 mL. Three blinded observers positioned 5 m from the device counted visible inflation-deflation cycles over 60-s intervals, with counts compared against simulator-delivered breaths. Performance was additionally assessed across four commercially available oxygen mask models. Results In screening, the lowest detected tidal volume fell from 400 mL with 20 µm film to 150 mL with 6 µm film, the thinnest film that could be reproducibly manufactured using the processes available to us. The final prototype produced observable breath indications in 42 of 56 tested conditions (all tidal volumes ≥150 mL) but not at 50 or 100 mL. Within these detectable conditions, observer counts agreed 100% with simulator-delivered breaths. Performance was maintained across four oxygen mask models. However, sensitivity was reduced by incomplete sealing at either the mask-to-face or device-to-mask interface. Conclusion Flow-Bag is a fully passive mechanical breath indicator. Under controlled bench conditions using a mechanical breathing simulator, it provided reliable visual indication of simulated breathing across respiratory rates of 5-60 breaths min⁻¹ and tidal volumes of 150-600 mL. Subject to clinical validation and safety testing, this device may offer a potentially low-cost adjunct alongside standard monitoring. .

Medical Engineering & Physics
The University of Queensland (AU), Wesley Hospital (AU)
Good health and well-being
Openalex Percentile: Top 24%
Non-Invasive Vital Sign Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.