Evaluation of a novel body temperature sensor (CORE) in dogs under general anesthesia following premedication with dexmedetomidine or acepromazine: A pilot study

Core body temperature monitoring in anesthetized dogs is a vital part of physiological evaluation. Depending on the procedure and patient positioning, consistent core temperature measurements can be difficult to obtain. The objective of this study was to evaluate the reliability and clinical validity of a wireless, non-invasive, superficial heat flux-based thermometric device (CORE, greenTEG AG, Rümlang, Switzerland). Three healthy beagle dogs were premedicated with either dexmedetomidine or acepromazine prior to induction with intravenous propofol and isoflurane maintenance. They were monitored every five minutes with pulse oximetry, non-invasive blood pressure, electrocardiography, and capnometry. Rectal thermometry and temperatures from the CORE sensor placed on clipped and unclipped skin over the jugular veins were measured every three minutes for up to 120 minutes following induction. Based on the correlation coefficient for repeated measures (rrm) and the Shrout-Fleiss intraclass correlation coefficient (ICC), in dogs administered acepromazine, there was good, statistically significant correlation and reliability between rectal and clipped CORE temperatures (rrm = 0.81, P < 0.0001; ICC = 0.78) and rectal and unclipped CORE temperatures (rrm = 0.79, P < 0.0001; ICC = 0.85). In dogs administered dexmedetomidine, although values were correlated (rrm = 0.30, P < 0.0001), there was poor reliability (ICC = 0.28) between rectal and clipped CORE temperatures which improved if dogs were unclipped (rrm = 0.48, P < 0.0001; ICC = 0.88). Clinically acceptable Bland-Altman agreements were found between rectal and unclipped CORE methods. These preliminary results suggest that the CORE sensor may provide an alternative method for assessing core body temperatures in unclipped, anesthetized dogs.

Authors

Institutions

Publication Details

Journal
PLoS ONE
Published
2026-10-15
DOI
https://doi.org/10.1371/journal.pone.0360808
Primary Topic
Veterinary Pharmacology and Anesthesia
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of a novel body temperature sensor (CORE) in dogs under general anesthesia following premedication with dexmedetomidine or acepromazine: A pilot study

Qianqian Zhao, Rebecca A. Johnson, Jacob P. Rastas, Chloe R. Dupleix et al.
PLoS ONE
Veterinary Pharmacology and Anesthesia
article

Evaluation of a novel body temperature sensor (CORE) in dogs under general anesthesia following premedication with dexmedetomidine or acepromazine: A pilot study

Qianqian Zhao, Rebecca A. Johnson, Jacob P. Rastas, Chloe R. Dupleix, Robert Young
article en

Abstract

Core body temperature monitoring in anesthetized dogs is a vital part of physiological evaluation. Depending on the procedure and patient positioning, consistent core temperature measurements can be difficult to obtain. The objective of this study was to evaluate the reliability and clinical validity of a wireless, non-invasive, superficial heat flux-based thermometric device (CORE, greenTEG AG, Rümlang, Switzerland). Three healthy beagle dogs were premedicated with either dexmedetomidine or acepromazine prior to induction with intravenous propofol and isoflurane maintenance. They were monitored every five minutes with pulse oximetry, non-invasive blood pressure, electrocardiography, and capnometry. Rectal thermometry and temperatures from the CORE sensor placed on clipped and unclipped skin over the jugular veins were measured every three minutes for up to 120 minutes following induction. Based on the correlation coefficient for repeated measures (rrm) and the Shrout-Fleiss intraclass correlation coefficient (ICC), in dogs administered acepromazine, there was good, statistically significant correlation and reliability between rectal and clipped CORE temperatures (rrm = 0.81, P < 0.0001; ICC = 0.78) and rectal and unclipped CORE temperatures (rrm = 0.79, P < 0.0001; ICC = 0.85). In dogs administered dexmedetomidine, although values were correlated (rrm = 0.30, P < 0.0001), there was poor reliability (ICC = 0.28) between rectal and clipped CORE temperatures which improved if dogs were unclipped (rrm = 0.48, P < 0.0001; ICC = 0.88). Clinically acceptable Bland-Altman agreements were found between rectal and unclipped CORE methods. These preliminary results suggest that the CORE sensor may provide an alternative method for assessing core body temperatures in unclipped, anesthetized dogs.

PLoS ONEVol. 21(10)
University of Wisconsin System (US), Waldorf College (US)
University of Wisconsin-Madison
Good health and well-being
Openalex Percentile: Top 10%
Veterinary Pharmacology and Anesthesia
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.