Point-of-care troponin I assay has acceptable performance detecting clinically important canine myocardial injury but reduced precision at low concentrations

Objective: To perform a partial validation of a point-of-care assay (Vcheck cardiac troponin I [VC-cTnI; cTnI]) for measurement of cTnI in canine serum and to perform a method comparison between the VC-cTnI and an established chemiluminescent assay (ADVIA Centaur XP High-Sensitivity Troponin I [AC-cTnI-HS]). Methods: From September 4, 2024, through August 21, 2025, surplus canine serum was collected from a diagnostic laboratory, and within-day variability, dilutional parallelism, spiking recovery, and the effect of common interferents (hemolysis, icterus, and lipemia) were analyzed for the VC-cTnI assay. A method comparison of the VC-cTnI and AC-cTnI-HS assays was performed with 121 canine serum samples. Results: The within-day coefficient of variation ranged from 8.1% to 43.8%, decreasing as cTnI concentrations increased. Dilutional parallelism and spiking recovery assessments had mean ± SD observed-to-expected ratios of 80.5 ± 33.2% (range, 21.1% to 122.2%) and 112.8 ± 35.6% (range, 46.0% to 173.9%), respectively. There was no interference from icterus, hemolysis, or lipemia. There was a positive correlation between the VC-cTnI and AC-cTnI-HS assays (Spearman ρ, 0.742), though 16 of 121 samples (13.2%) were below the VC-cTnI detectable limit. Conclusions: The VC-cTnI assay lacked precision for measurement of concentrations below 0.2 ng/mL, but within-day variability was acceptable for higher concentrations. Clinical Relevance: The VC-cTnI assay is more precise when measuring higher cTnI concentrations and can detect myocardial injury in dogs. Results in the normal and equivocal range of the manufacturer's reference interval may be more challenging to interpret because of suboptimal precision in this range.

Authors

Institutions

Publication Details

Journal
American Journal of Veterinary Research
Published
2026-10-14
DOI
https://doi.org/10.2460/ajvr.26.08.0336
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Point-of-care troponin I assay has acceptable performance detecting clinically important canine myocardial injury but reduced precision at low concentrations

Jonathan A. Lidbury, Sonya R. Wesselowski, Ashley B. Saunders, Bobbie Ditzler et al.
American Journal of Veterinary Research
Clinical Laboratory Practices and Quality Control
article

Point-of-care troponin I assay has acceptable performance detecting clinically important canine myocardial injury but reduced precision at low concentrations

Jonathan A. Lidbury, Sonya R. Wesselowski, Ashley B. Saunders, Bobbie Ditzler, Jacob W. Dunham, Philip Guadiano
article en

Abstract

Objective: To perform a partial validation of a point-of-care assay (Vcheck cardiac troponin I [VC-cTnI; cTnI]) for measurement of cTnI in canine serum and to perform a method comparison between the VC-cTnI and an established chemiluminescent assay (ADVIA Centaur XP High-Sensitivity Troponin I [AC-cTnI-HS]). Methods: From September 4, 2024, through August 21, 2025, surplus canine serum was collected from a diagnostic laboratory, and within-day variability, dilutional parallelism, spiking recovery, and the effect of common interferents (hemolysis, icterus, and lipemia) were analyzed for the VC-cTnI assay. A method comparison of the VC-cTnI and AC-cTnI-HS assays was performed with 121 canine serum samples. Results: The within-day coefficient of variation ranged from 8.1% to 43.8%, decreasing as cTnI concentrations increased. Dilutional parallelism and spiking recovery assessments had mean ± SD observed-to-expected ratios of 80.5 ± 33.2% (range, 21.1% to 122.2%) and 112.8 ± 35.6% (range, 46.0% to 173.9%), respectively. There was no interference from icterus, hemolysis, or lipemia. There was a positive correlation between the VC-cTnI and AC-cTnI-HS assays (Spearman ρ, 0.742), though 16 of 121 samples (13.2%) were below the VC-cTnI detectable limit. Conclusions: The VC-cTnI assay lacked precision for measurement of concentrations below 0.2 ng/mL, but within-day variability was acceptable for higher concentrations. Clinical Relevance: The VC-cTnI assay is more precise when measuring higher cTnI concentrations and can detect myocardial injury in dogs. Results in the normal and equivocal range of the manufacturer's reference interval may be more challenging to interpret because of suboptimal precision in this range.

American Journal of Veterinary Research
Texas A&M University (US)
Good health and well-being
Openalex Percentile: Top 14%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.