Mechanical stabilization of non-vascular catheters using a low-profile suture-anchored device

To evaluate the mechanical performance of AnchorDoc, a novel, low-profile, two-piece, suture-anchored device, against standard securement methods. We hypothesized that AnchorDoc would provide superior mechanical stability without compromising catheter patency. Using a custom anatomically-informed gelatin phantom of the human flank with porcine-skin overlay, we compared the performance of AnchorDoc against a standard adhesive device (StatLock) and a suture-tape technique (Roman sandal) for 8.5, 10.2, and 12 French (Fr) catheters. We conducted: (1) static tensile testing under dry and wet conditions to measure peak force to catheter slip (≥ 5 mm); (2) dynamic gait simulation to assess time-to-failure; and (3) static flow testing to evaluate for lumen compromise. In tensile testing, AnchorDoc required a significantly higher dislodgement force than both StatLock and Roman Sandal under dry (e.g., 10.2 Fr: 31.1 ± 1.4 N vs. 20.5 ± 2.4 N and 18.8 ± 2.3 N, respectively; p < 0.0001) and wet conditions (25.5 ± 1.0 N vs. 10.9 ± 1.3 N and 18.7 ± 1.0 N; p < 0.0001). The adhesive device’s retention strength degraded by nearly 50% when wet, while AnchorDoc maintained its superiority. In dynamic gait simulation, Kaplan-Meier analysis showed that AnchorDoc had a significantly higher survival probability (fewer failures) while tested over 12,000 cycles compared to both comparators (p < 0.05). Flow testing confirmed that AnchorDoc did not cause a statistically significant reduction in flow rate or lumen diameter across all catheter sizes (p > 0.05). The AnchorDoc device provides significantly greater resistance to both static and dynamic dislodgement forces than common adhesive and suture-tape techniques, particularly in a laboratory-simulated wet environment, without impairing catheter function. This suture-anchored, two-part platform represents a promising solution to mitigate the risk of non-vascular catheter dislodgement.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2026-10-13
DOI
https://doi.org/10.1038/s41598-026-68117-1
Primary Topic
Central Venous Catheters and Hemodialysis
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Mechanical stabilization of non-vascular catheters using a low-profile suture-anchored device

Qiyue Ma, Chang Zhou, Hanzhou Li, John Moon et al.
Scientific Reports
Central Venous Catheters and Hemodialysis
article

Mechanical stabilization of non-vascular catheters using a low-profile suture-anchored device

Qiyue Ma, Chang Zhou, Hanzhou Li, John Moon, Yue Chen, Zachary Bercu, Shreyas Naidu
article en

Abstract

To evaluate the mechanical performance of AnchorDoc, a novel, low-profile, two-piece, suture-anchored device, against standard securement methods. We hypothesized that AnchorDoc would provide superior mechanical stability without compromising catheter patency. Using a custom anatomically-informed gelatin phantom of the human flank with porcine-skin overlay, we compared the performance of AnchorDoc against a standard adhesive device (StatLock) and a suture-tape technique (Roman sandal) for 8.5, 10.2, and 12 French (Fr) catheters. We conducted: (1) static tensile testing under dry and wet conditions to measure peak force to catheter slip (≥ 5 mm); (2) dynamic gait simulation to assess time-to-failure; and (3) static flow testing to evaluate for lumen compromise. In tensile testing, AnchorDoc required a significantly higher dislodgement force than both StatLock and Roman Sandal under dry (e.g., 10.2 Fr: 31.1 ± 1.4 N vs. 20.5 ± 2.4 N and 18.8 ± 2.3 N, respectively; p < 0.0001) and wet conditions (25.5 ± 1.0 N vs. 10.9 ± 1.3 N and 18.7 ± 1.0 N; p < 0.0001). The adhesive device’s retention strength degraded by nearly 50% when wet, while AnchorDoc maintained its superiority. In dynamic gait simulation, Kaplan-Meier analysis showed that AnchorDoc had a significantly higher survival probability (fewer failures) while tested over 12,000 cycles compared to both comparators (p < 0.05). Flow testing confirmed that AnchorDoc did not cause a statistically significant reduction in flow rate or lumen diameter across all catheter sizes (p > 0.05). The AnchorDoc device provides significantly greater resistance to both static and dynamic dislodgement forces than common adhesive and suture-tape techniques, particularly in a laboratory-simulated wet environment, without impairing catheter function. This suture-anchored, two-part platform represents a promising solution to mitigate the risk of non-vascular catheter dislodgement.

Scientific Reports
Georgia Institute of Technology (US), Emory University (US), University of Florida Health (US), The Wallace H. Coulter Department of Biomedical Engineering (US), University of Florida (US)
National Institutes of Health, National Center for Advancing Translational Sciences
Openalex Percentile: Top 8%
Central Venous Catheters and Hemodialysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.