A practical demand-based model for determining cyclotron irradiation time in routine 18F-FDG production

PURPOSE: The study aimed to develop and evaluate a practical model for determining cyclotron irradiation time based on planned fluorodeoxyglucose (FDG) supply in a hospital-based production setting. METHODS: Irradiation time was calculated using a composite coefficient of 2.8 min per GBq of planned clinical demand. The coefficient integrated a facility-specific conversion between required end-of-bombardment activity and irradiation time, a conservative synthesis yield of 55%, quality control activity consumption of 11.1%, radioactive decay during an approximately 30-min processing interval, and an operational safety factor of 1.2. RESULTS: During a fixed 24-month evaluation period, the model was applied on 516 consecutive scheduled FDG production days. Sufficient FDG activity was supplied for all scheduled examinations on all 516 days (100%), with no activity shortage attributable to underestimation of the required irradiation time. CONCLUSION: A demand-based irradiation model incorporating conservative yield assumptions, quality control consumption, and handling margins enables reliable FDG supply while maintaining an appropriate activity margin. This approach provides a practical framework for hospital-based radiopharmaceutical production.

Authors

Institutions

Publication Details

Journal
Nuclear Medicine Communications
Published
2026-10-13
DOI
https://doi.org/10.1097/mnm.0000000000002250
Primary Topic
Radiopharmaceutical Chemistry and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A practical demand-based model for determining cyclotron irradiation time in routine 18F-FDG production

Hiroshi Yamaguchi
Nuclear Medicine Communications
Radiopharmaceutical Chemistry and Applications
article

A practical demand-based model for determining cyclotron irradiation time in routine 18F-FDG production

Hiroshi Yamaguchi
article en

Abstract

PURPOSE: The study aimed to develop and evaluate a practical model for determining cyclotron irradiation time based on planned fluorodeoxyglucose (FDG) supply in a hospital-based production setting. METHODS: Irradiation time was calculated using a composite coefficient of 2.8 min per GBq of planned clinical demand. The coefficient integrated a facility-specific conversion between required end-of-bombardment activity and irradiation time, a conservative synthesis yield of 55%, quality control activity consumption of 11.1%, radioactive decay during an approximately 30-min processing interval, and an operational safety factor of 1.2. RESULTS: During a fixed 24-month evaluation period, the model was applied on 516 consecutive scheduled FDG production days. Sufficient FDG activity was supplied for all scheduled examinations on all 516 days (100%), with no activity shortage attributable to underestimation of the required irradiation time. CONCLUSION: A demand-based irradiation model incorporating conservative yield assumptions, quality control consumption, and handling margins enables reliable FDG supply while maintaining an appropriate activity margin. This approach provides a practical framework for hospital-based radiopharmaceutical production.

Nuclear Medicine Communications
Fujita Health University Hospital (JP)
Openalex Percentile: Top 13%
Radiopharmaceutical Chemistry and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.