Improved assessment of the cochlear implant and temporal bone structures with ultra-high resolution photon-counting detector CT compared to conventional CT systems

PURPOSE: This study evaluates the image quality of photon-counting detector CT (PCD-CT) compared to conventional CT systems in cochlear implant imaging and aims to determine optimal scan parameters. METHODS: Three cochlear implant models were inserted into three post-mortem human specimen heads. Each head was scanned on three high-end CT systems: SOMATOM Edge, SOMATOM Force and NAEOTOM Alpha. A total of 192 reconstructions were obtained, including various tube voltages (120-150 kV), tin filtering, and slice thicknesses (0.2-0.6 mm). For each scanner, multiple reconstruction kernels were applied. Contrast-to-noise ratios (CNR) for bone-to-soft-tissue were calculated. Two radiologists independently assessed anatomy, implant visibility, and overall image quality on a 5-point Likert scale and scored diagnostic confidence for detecting anatomic dehiscence. RESULTS: PCD-CT showed significantly higher image quality scores across implant, anatomy, and general image quality compared to the Edge and Force (p < 0.001). PCD-CT also produced significantly higher CNR (mean 23.5 vs. 13.8 [Edge] and 12.7 [Force]) and greater diagnostic confidence. For PCD-CT, optimal image quality was achieved using 0.2 mm slice thickness and the Hr89 kernel at 140 kV. PCD-CT scans acquired with half dose (27.5 mGy) showed no significant difference in CNRs but led to slightly lower subjective scores compared to standard-dose imaging. CONCLUSION: PCD-CT provides superior image quality and diagnostic confidence in cochlear implant imaging compared to the Edge and Force. Optimal results are achieved using 140 kV, 0.2 mm slices, and the Hr89 kernel. Although dose reduction is feasible, it slightly compromises subjective image quality.

Authors

Institutions

Publication Details

Journal
European Archives of Oto-Rhino-Laryngology
Published
2026-10-13
DOI
https://doi.org/10.1007/s00405-026-10653-7
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Improved assessment of the cochlear implant and temporal bone structures with ultra-high resolution photon-counting detector CT compared to conventional CT systems

Marcel L. Dijkshoorn, Marcel van Straten, Ronald Booij, Judith van der Bie et al.
European Archives of Oto-Rhino-Laryngology
Advanced X-ray and CT Imaging
article

Improved assessment of the cochlear implant and temporal bone structures with ultra-high resolution photon-counting detector CT compared to conventional CT systems

Marcel L. Dijkshoorn, Marcel van Straten, Ronald Booij, Judith van der Bie, Hylke F. E. van der Toom, Robert Jan Pauw, Joel de Groen, Floris V.W.J. van Zijl, Aaike S. van den Berg
article en

Abstract

PURPOSE: This study evaluates the image quality of photon-counting detector CT (PCD-CT) compared to conventional CT systems in cochlear implant imaging and aims to determine optimal scan parameters. METHODS: Three cochlear implant models were inserted into three post-mortem human specimen heads. Each head was scanned on three high-end CT systems: SOMATOM Edge, SOMATOM Force and NAEOTOM Alpha. A total of 192 reconstructions were obtained, including various tube voltages (120-150 kV), tin filtering, and slice thicknesses (0.2-0.6 mm). For each scanner, multiple reconstruction kernels were applied. Contrast-to-noise ratios (CNR) for bone-to-soft-tissue were calculated. Two radiologists independently assessed anatomy, implant visibility, and overall image quality on a 5-point Likert scale and scored diagnostic confidence for detecting anatomic dehiscence. RESULTS: PCD-CT showed significantly higher image quality scores across implant, anatomy, and general image quality compared to the Edge and Force (p < 0.001). PCD-CT also produced significantly higher CNR (mean 23.5 vs. 13.8 [Edge] and 12.7 [Force]) and greater diagnostic confidence. For PCD-CT, optimal image quality was achieved using 0.2 mm slice thickness and the Hr89 kernel at 140 kV. PCD-CT scans acquired with half dose (27.5 mGy) showed no significant difference in CNRs but led to slightly lower subjective scores compared to standard-dose imaging. CONCLUSION: PCD-CT provides superior image quality and diagnostic confidence in cochlear implant imaging compared to the Edge and Force. Optimal results are achieved using 140 kV, 0.2 mm slices, and the Hr89 kernel. Although dose reduction is feasible, it slightly compromises subjective image quality.

European Archives of Oto-Rhino-Laryngology
Erasmus MC (NL), Erasmus University Rotterdam (NL)
Siemens Healthineers
Openalex Percentile: Top 25%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.