Direct Determination of Readily Accessible Quercetin-Reactive Aluminum and Ascorbate-Reducible, Ferrozine-Reactive Iron in Cosmetic Emulsions

Aluminum in antiperspirants is relevant to formulation control and compliance, whereas reactive iron can promote oxidative instability in emulsions. We developed smartphone-based colorimetric assays relying on metal–reagent complexation directly in oil-in-water (O/W) emulsions, without digestion or extraction. The aluminum assay measures quercetin-reactive aluminum from aluminum chlorohydrate (ACH). Calibration data were fitted by linear regression (R2 ≥ 0.971). Acceptable accuracy (relative bias within ± 8.5%) and intermediate precision (CV ≤ 9.4%) were obtained upon quality control (QC) analysis (above the LLOQ). Relative bias in commercial antiperspirants and corresponding in-house preparations was within ± 15.2%. The iron assay measures ascorbate-reducible, ferrozine-reactive iron. Its calibration data were also fitted by linear regression (R2 ≥ 0.981), while QC analysis showed acceptable accuracy (relative bias within ± 13.7%) and intermediate precision (CV ≤ 12.6%). Comparison with portable energy-dispersive X-ray fluorescence (EDXRF) yielded a regression slope of 1.14 and an intercept of 0.40 μg/g, suggesting a proportional difference and a possible constant component. Common chelators and certain metals interfered with both assays. The effect of interferents in commercially relevant concentration was, however, minimal (relative bias within ± 15.2% for Al and at 4.5% for a Co/Cu/Fe-fortified model emulsion) after matrix matching or dilution. Although mostly inferior in terms of technical characteristics compared to established instrumental methods, this methodology is sufficiently accurate, precise and selective under the defined conditions.

Authors

Institutions

Publication Details

Journal
Analytica—A Journal of Analytical Chemistry and Chemical Analysis
Published
2026-10-12
DOI
https://doi.org/10.3390/analytica7040078
Primary Topic
Analytical Methods in Pharmaceuticals
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Direct Determination of Readily Accessible Quercetin-Reactive Aluminum and Ascorbate-Reducible, Ferrozine-Reactive Iron in Cosmetic Emulsions

Georgia Eleni Tsotsou, M. Pilakouta, Georgia Bakara, Dimitra Dremetsika
Analytica—A Journal of Analytical Chemistry and Chemical Analysis
Analytical Methods in Pharmaceuticals
article

Direct Determination of Readily Accessible Quercetin-Reactive Aluminum and Ascorbate-Reducible, Ferrozine-Reactive Iron in Cosmetic Emulsions

Georgia Eleni Tsotsou, M. Pilakouta, Georgia Bakara, Dimitra Dremetsika
article en

Abstract

Aluminum in antiperspirants is relevant to formulation control and compliance, whereas reactive iron can promote oxidative instability in emulsions. We developed smartphone-based colorimetric assays relying on metal–reagent complexation directly in oil-in-water (O/W) emulsions, without digestion or extraction. The aluminum assay measures quercetin-reactive aluminum from aluminum chlorohydrate (ACH). Calibration data were fitted by linear regression (R2 ≥ 0.971). Acceptable accuracy (relative bias within ± 8.5%) and intermediate precision (CV ≤ 9.4%) were obtained upon quality control (QC) analysis (above the LLOQ). Relative bias in commercial antiperspirants and corresponding in-house preparations was within ± 15.2%. The iron assay measures ascorbate-reducible, ferrozine-reactive iron. Its calibration data were also fitted by linear regression (R2 ≥ 0.981), while QC analysis showed acceptable accuracy (relative bias within ± 13.7%) and intermediate precision (CV ≤ 12.6%). Comparison with portable energy-dispersive X-ray fluorescence (EDXRF) yielded a regression slope of 1.14 and an intercept of 0.40 μg/g, suggesting a proportional difference and a possible constant component. Common chelators and certain metals interfered with both assays. The effect of interferents in commercially relevant concentration was, however, minimal (relative bias within ± 15.2% for Al and at 4.5% for a Co/Cu/Fe-fortified model emulsion) after matrix matching or dilution. Although mostly inferior in terms of technical characteristics compared to established instrumental methods, this methodology is sufficiently accurate, precise and selective under the defined conditions.

Analytica—A Journal of Analytical Chemistry and Chemical AnalysisVol. 7(4)
University of West Attica (GR)
Openalex Percentile: Top 18%
Analytical Methods in Pharmaceuticals
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Direct Determination of Readily Accessible Quercetin-Reactive Aluminum and Ascorbate-Reducible, Ferrozine-Reactive Iron in Cosmetic Emulsions — Georgia Eleni Tsotsou, M. Pilakouta, et al. · Analytica—A Journal of Analytical Chemistry and Chemical Analysis (2026) | TGRS Research Map | TGRS