Comparison of Treatment Zone Decentration Between VISUMAX 500 and VISUMAX 800 Keratorefractive Lenticule Extraction Using Anterior Segment Optical Coherence Tomography

Background/Objectives: To compare treatment-zone decentration after small-incision lenticule extraction (SMILE) performed with VISUMAX 500 and VISUMAX 800 using corneal thickness difference maps derived from anterior segment optical coherence tomography. Methods: This retrospective comparative study included 44 eyes of 22 participants who underwent SMILE with VISUMAX 500 or VISUMAX 800 by a single surgeon and had preoperative and 3-month postoperative CASIA2 examinations. Treatment-zone centers were primarily identified manually relative to the corneal vertex. Between-platform differences were analyzed using generalized estimating equations (GEE) accounting for inter-eye correlation, with additional adjustment for preoperative cylinder. Participant-level analyses and semi-automated radial color-symmetry localization were performed as sensitivity analyses. Results: Twenty-two eyes of 11 participants were included in each group. Mean radial decentration was 0.219 ± 0.134 mm with VISUMAX 500 and 0.225 ± 0.171 mm with VISUMAX 800. The mean between-group difference was 0.006 mm (95% confidence interval [CI], −0.074 to 0.085; p = 0.890), and remained similar after adjustment for preoperative cylinder (0.006 mm; 95% CI, −0.082 to 0.094; p = 0.888). Participant-level analysis likewise showed no evidence of a between-group difference (mean difference, 0.0056 mm; 95% CI, −0.083 to 0.094; exact permutation p = 0.907). Reanalysis using semi-automated localization yielded a similarly small between-platform difference (0.0114 mm; 95% CI, −0.063 to 0.086; p = 0.763). Conclusions: In this small single-surgeon cohort, treatment-zone decentration did not differ significantly between VISUMAX 500 and VISUMAX 800. The findings were consistent across participant-level and semi-automated sensitivity analyses. Because the study was not designed as an equivalence or noninferiority trial, these results should be interpreted as an absence of evidence for a between-platform difference rather than evidence of equivalent centration performance.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Medicine
Published
2026-10-12
DOI
https://doi.org/10.3390/jcm15207851
Primary Topic
Corneal surgery and disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of Treatment Zone Decentration Between VISUMAX 500 and VISUMAX 800 Keratorefractive Lenticule Extraction Using Anterior Segment Optical Coherence Tomography

Sang Mok Lee, Tae Keun Yoo, Si-Hoon Park
Journal of Clinical Medicine
Corneal surgery and disorders
article

Comparison of Treatment Zone Decentration Between VISUMAX 500 and VISUMAX 800 Keratorefractive Lenticule Extraction Using Anterior Segment Optical Coherence Tomography

Sang Mok Lee, Tae Keun Yoo, Si-Hoon Park
article en

Abstract

Background/Objectives: To compare treatment-zone decentration after small-incision lenticule extraction (SMILE) performed with VISUMAX 500 and VISUMAX 800 using corneal thickness difference maps derived from anterior segment optical coherence tomography. Methods: This retrospective comparative study included 44 eyes of 22 participants who underwent SMILE with VISUMAX 500 or VISUMAX 800 by a single surgeon and had preoperative and 3-month postoperative CASIA2 examinations. Treatment-zone centers were primarily identified manually relative to the corneal vertex. Between-platform differences were analyzed using generalized estimating equations (GEE) accounting for inter-eye correlation, with additional adjustment for preoperative cylinder. Participant-level analyses and semi-automated radial color-symmetry localization were performed as sensitivity analyses. Results: Twenty-two eyes of 11 participants were included in each group. Mean radial decentration was 0.219 ± 0.134 mm with VISUMAX 500 and 0.225 ± 0.171 mm with VISUMAX 800. The mean between-group difference was 0.006 mm (95% confidence interval [CI], −0.074 to 0.085; p = 0.890), and remained similar after adjustment for preoperative cylinder (0.006 mm; 95% CI, −0.082 to 0.094; p = 0.888). Participant-level analysis likewise showed no evidence of a between-group difference (mean difference, 0.0056 mm; 95% CI, −0.083 to 0.094; exact permutation p = 0.907). Reanalysis using semi-automated localization yielded a similarly small between-platform difference (0.0114 mm; 95% CI, −0.063 to 0.086; p = 0.763). Conclusions: In this small single-surgeon cohort, treatment-zone decentration did not differ significantly between VISUMAX 500 and VISUMAX 800. The findings were consistent across participant-level and semi-automated sensitivity analyses. Because the study was not designed as an equivalence or noninferiority trial, these results should be interpreted as an absence of evidence for a between-platform difference rather than evidence of equivalent centration performance.

Journal of Clinical MedicineVol. 15(20)
Onnuri Smile Eye Clinic (KR), Catholic Kwandong University (KR)
Openalex Percentile: Top 12%
Corneal surgery and disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.