Body Size, Radiation Dose and Exposure Index in Conventional Radiography: A Multicentre Prospective Study

Background/Objectives: Exposure index (EI) reflects detector exposure rather than patient dose, but its behaviour as patient size increases remains incompletely characterised across radiographic examinations. We evaluated continuous associations of body mass index (BMI) and regional body thickness proxy with dose–area product (DAP), entrance surface dose (ESD), and EI, and tested whether associations differed by examination and institution. Methods: This prospective multicentre observational study included 572 exposures from 331 adults undergoing nine projections at three institutions. Generalised estimating equations with a Gamma distribution and log link accounted for repeated exposures. Total-association models adjusted for examination, institution, age, and sex; technical-adjusted models additionally included log mAs and kVp, with SSD included only in BMI-based models. Results: In total-association models adjusted for examination, institution, age, and sex, each 1 kg/m2 increase in BMI was associated with 4.31% higher DAP, 3.25% higher estimated ESD, and 3.45% lower EI (all p < 0.001). Each 1 cm increase in geometrically derived regional-thickness proxy was associated with 3.03% higher DAP, 2.97% higher estimated ESD, and 2.51% lower EI (all p < 0.001). Technical adjustment substantially attenuated the associations with DAP, whereas inverse associations with EI persisted. Conclusions: Body-size associations were outcome- and examination-specific. Direct EI–dose associations were generally positive but substantially attenuated after accounting for exposure technique and regional thickness, indicating that their relationship is strongly influenced by patient attenuation and technical exposure factors. These findings support further examination- and system-specific evaluation of EI behaviour alongside body size, technical factors, and patient-dose indicators.

Authors

Publication Details

Journal
Diagnostics
Published
2026-11-11
DOI
https://doi.org/10.3390/diagnostics16223723
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Body Size, Radiation Dose and Exposure Index in Conventional Radiography: A Multicentre Prospective Study

Manuel Duarte Lobo, Rui Pedro de Almeida, Sergio Miravent
1 links
Diagnostics
Radiation Dose and Imaging
article

Body Size, Radiation Dose and Exposure Index in Conventional Radiography: A Multicentre Prospective Study

Manuel Duarte Lobo, Rui Pedro de Almeida, Sergio Miravent
article en

Abstract

Background/Objectives: Exposure index (EI) reflects detector exposure rather than patient dose, but its behaviour as patient size increases remains incompletely characterised across radiographic examinations. We evaluated continuous associations of body mass index (BMI) and regional body thickness proxy with dose–area product (DAP), entrance surface dose (ESD), and EI, and tested whether associations differed by examination and institution. Methods: This prospective multicentre observational study included 572 exposures from 331 adults undergoing nine projections at three institutions. Generalised estimating equations with a Gamma distribution and log link accounted for repeated exposures. Total-association models adjusted for examination, institution, age, and sex; technical-adjusted models additionally included log mAs and kVp, with SSD included only in BMI-based models. Results: In total-association models adjusted for examination, institution, age, and sex, each 1 kg/m2 increase in BMI was associated with 4.31% higher DAP, 3.25% higher estimated ESD, and 3.45% lower EI (all p < 0.001). Each 1 cm increase in geometrically derived regional-thickness proxy was associated with 3.03% higher DAP, 2.97% higher estimated ESD, and 2.51% lower EI (all p < 0.001). Technical adjustment substantially attenuated the associations with DAP, whereas inverse associations with EI persisted. Conclusions: Body-size associations were outcome- and examination-specific. Direct EI–dose associations were generally positive but substantially attenuated after accounting for exposure technique and regional thickness, indicating that their relationship is strongly influenced by patient attenuation and technical exposure factors. These findings support further examination- and system-specific evaluation of EI behaviour alongside body size, technical factors, and patient-dose indicators.

DiagnosticsVol. 16(22)
Good health and well-being
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon