Comparative In Vitro Assessment of Original Reciproc® Blue Instruments vs. Two Replicas: Mechanical Characteristics, Safety, and Efficacy Analysis

Background/Objectives: To compare mechanical behavior, shaping performance, and safety of an original reciprocating endodontic instrument (Reciproc® Blue) with two visually similar replicas obtained from an online marketplace. Methods: The original Reciproc® Blue and two replica (designated Replica 1 and Replica 2) instruments purchased from AliExpress were tested. Evaluations included torque resistance, bending resistance, cyclic fatigue resistance (all according to ISO 3630-1 standards, n = 10 per group per test), and an in vitro shaping procedure using artificial root canals in plastic blocks (5 canals per instrument, n = 10 instruments per group) were carried out. Preparation time, number of pecks, procedural errors (blockages, ledges, deformations), and instrument fractures were recorded. Surface characteristics of unused (n = 2) and used instruments (n = 2) were examined using scanning electron microscopy (SEM). Statistical analysis used one-way ANOVA, Welch ANOVA, or Kruskal–Wallis tests according to the distribution and variance pattern of each endpoint, followed by the corresponding post hoc test. Repeated use was analyzed with a linear mixed effects model and checked with a robust generalized estimating equation sensitivity analysis. Deformation-free and fracture-free event curves were compared with exploratory log rank tests. Results: The original instrument showed the highest Mf/D3 and Mt/D3 values. Overall group differences were significant for Mf/D3 (Welch p < 0.001) and Mt/D3 (Welch p = 0.001), with significant Games Howell contrasts between the original and Replica 2. Replicas showed longer cyclic fatigue life (Original 407.1 s, Replica 1 490.2 s, Replica 2 607.8 s; ANOVA p < 0.001). The original instrument had a lower average preparation time for each canal than both replicas (72.8 s vs. 95.8 s and 92.3 s; Kruskal–Wallis p < 0.001) and a lower average number of pecks (8.64 vs. 9.71 and 10.00; p = 0.008). Exploratory log rank comparisons also showed group differences in the deformation-free and fracture-free event curves (Holm adjusted p = 0.005 and p = 0.024, respectively). Both replicas failed to reach working length in 4 out of 10 cases each, with high rates of blockages (Replica 1: 7/10; Replica 2: 2/10), ledges (Replica 1 only), and deformations. The limited SEM subsample revealed severe wear of cutting edges in Replica 1. Conclusions: Both tested replicas exhibited significant deviations in mechanical behavior and markedly inferior shaping safety compared to the original Reciproc® Blue instrument, despite superficial visual similarity. Therefore, dental practitioners should be aware that low-cost imitations may compromise clinical safety.

Authors

Institutions

Publication Details

Journal
Oral
Published
2026-10-09
DOI
https://doi.org/10.3390/oral6050130
Primary Topic
Endodontics and Root Canal Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparative In Vitro Assessment of Original Reciproc® Blue Instruments vs. Two Replicas: Mechanical Characteristics, Safety, and Efficacy Analysis

Sebastian Fitzek, Maria Lessani, Jörg Philipp Tchorz, Yalda Amir Sayfadini et al.
Oral
Endodontics and Root Canal Treatments
article

Comparative In Vitro Assessment of Original Reciproc® Blue Instruments vs. Two Replicas: Mechanical Characteristics, Safety, and Efficacy Analysis

Sebastian Fitzek, Maria Lessani, Jörg Philipp Tchorz, Yalda Amir Sayfadini, Karl-Thomas Wrbas, Johan Ripplinger
article en

Abstract

Background/Objectives: To compare mechanical behavior, shaping performance, and safety of an original reciprocating endodontic instrument (Reciproc® Blue) with two visually similar replicas obtained from an online marketplace. Methods: The original Reciproc® Blue and two replica (designated Replica 1 and Replica 2) instruments purchased from AliExpress were tested. Evaluations included torque resistance, bending resistance, cyclic fatigue resistance (all according to ISO 3630-1 standards, n = 10 per group per test), and an in vitro shaping procedure using artificial root canals in plastic blocks (5 canals per instrument, n = 10 instruments per group) were carried out. Preparation time, number of pecks, procedural errors (blockages, ledges, deformations), and instrument fractures were recorded. Surface characteristics of unused (n = 2) and used instruments (n = 2) were examined using scanning electron microscopy (SEM). Statistical analysis used one-way ANOVA, Welch ANOVA, or Kruskal–Wallis tests according to the distribution and variance pattern of each endpoint, followed by the corresponding post hoc test. Repeated use was analyzed with a linear mixed effects model and checked with a robust generalized estimating equation sensitivity analysis. Deformation-free and fracture-free event curves were compared with exploratory log rank tests. Results: The original instrument showed the highest Mf/D3 and Mt/D3 values. Overall group differences were significant for Mf/D3 (Welch p < 0.001) and Mt/D3 (Welch p = 0.001), with significant Games Howell contrasts between the original and Replica 2. Replicas showed longer cyclic fatigue life (Original 407.1 s, Replica 1 490.2 s, Replica 2 607.8 s; ANOVA p < 0.001). The original instrument had a lower average preparation time for each canal than both replicas (72.8 s vs. 95.8 s and 92.3 s; Kruskal–Wallis p < 0.001) and a lower average number of pecks (8.64 vs. 9.71 and 10.00; p = 0.008). Exploratory log rank comparisons also showed group differences in the deformation-free and fracture-free event curves (Holm adjusted p = 0.005 and p = 0.024, respectively). Both replicas failed to reach working length in 4 out of 10 cases each, with high rates of blockages (Replica 1: 7/10; Replica 2: 2/10), ledges (Replica 1 only), and deformations. The limited SEM subsample revealed severe wear of cutting edges in Replica 1. Conclusions: Both tested replicas exhibited significant deviations in mechanical behavior and markedly inferior shaping safety compared to the original Reciproc® Blue instrument, despite superficial visual similarity. Therefore, dental practitioners should be aware that low-cost imitations may compromise clinical safety.

OralVol. 6(5)
Danube Private University (AT), University of Freiburg (DE), University Medical Center Freiburg (DE)
Openalex Percentile: Top 9%
Endodontics and Root Canal Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.