Assessment of Heavy Metal Concentrations and Water Quality Parameters in Hemodialysis Water in Kirkuk, Iraq

Objective: To assess physicochemical parameters and selected trace/heavy metals in water before and after the reverse osmosis (RO) treatment system at the main dialysis center in Kirkuk, Iraq, and to compare the results with WHO reference values and applicable AAMI/ANSI/ISO requirements. Methods: Water was sampled at the RO inlet (In) and outlet (Out) at two-week intervals from the beginning of April to mid-May 2024. Samples were collected in 0.5-L polyethylene bottles and stored at approximately 8 °C until analysis. pH, total dissolved solids (TDS), electrical conductivity (EC), Na, K, Ca, Mg, and selected trace metals were measured using pH/TDS/EC meters, flame photometry, flame atomic absorption spectrometry, and ICP-OES. Results: pH decreased from 7.60 ± 0.0493 to 6.50 ± 0.0950. TDS decreased from 688.3 ± 1.528 to 6.00 ± 0.800 mg/L, and EC from 1459 ± 2.646 to 13.00 ± 1.00 µS/cm (P < 0.0001). Na remained below its applicable limit. K decreased from 9.439 ± 0.1063 to 4.053 ± 0.02292 mg/L (P = 0.0002). Ca and Mg decreased to 39.08 ± 0.02235 and 21.70 ± 0.1865 mg/L, respectively, but remained above their applicable limits (P < 0.0001). Zn, Pb, and Cd decreased after RO but remained above their respective limits; Cr was 0.0200 ± 0.0006 mg/L at the inlet (P = 0.004) and was not detected after RO at the reported analytical level. Fe remained detectable at 0.1044 ± 0.0001 mg/L before RO and 0.09550 ± 0.0002 mg/L after RO. Al, Be, Ba, As, Ag, Sb, Se, Tl, Co, Ni, Mn, and Cu were not detected at the reported analytical level. Conclusion: RO treatment substantially reduced several measured parameters but did not achieve chemical compliance for Ca, Mg, Zn, Pb, and Cd. The findings support routine monitoring and systematic evaluation of pretreatment, membrane condition, storage, and distribution components.

Authors

Institutions

Publication Details

Journal
Science Forum (Journal of Pure and Applied Sciences)
Published
2026-10-06
DOI
https://doi.org/10.70882/hck1rx09
Primary Topic
Heavy Metal Pollution Remediation
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Assessment of Heavy Metal Concentrations and Water Quality Parameters in Hemodialysis Water in Kirkuk, Iraq

Sarhan Ali Salman, Firas Taher Maher, A. M. (Ameen) Aljuraisy
Science Forum (Journal of Pure and Applied Sciences)
Heavy Metal Pollution Remediation
article

Assessment of Heavy Metal Concentrations and Water Quality Parameters in Hemodialysis Water in Kirkuk, Iraq

Sarhan Ali Salman, Firas Taher Maher, A. M. (Ameen) Aljuraisy
article en

Abstract

Objective: To assess physicochemical parameters and selected trace/heavy metals in water before and after the reverse osmosis (RO) treatment system at the main dialysis center in Kirkuk, Iraq, and to compare the results with WHO reference values and applicable AAMI/ANSI/ISO requirements. Methods: Water was sampled at the RO inlet (In) and outlet (Out) at two-week intervals from the beginning of April to mid-May 2024. Samples were collected in 0.5-L polyethylene bottles and stored at approximately 8 °C until analysis. pH, total dissolved solids (TDS), electrical conductivity (EC), Na, K, Ca, Mg, and selected trace metals were measured using pH/TDS/EC meters, flame photometry, flame atomic absorption spectrometry, and ICP-OES. Results: pH decreased from 7.60 ± 0.0493 to 6.50 ± 0.0950. TDS decreased from 688.3 ± 1.528 to 6.00 ± 0.800 mg/L, and EC from 1459 ± 2.646 to 13.00 ± 1.00 µS/cm (P < 0.0001). Na remained below its applicable limit. K decreased from 9.439 ± 0.1063 to 4.053 ± 0.02292 mg/L (P = 0.0002). Ca and Mg decreased to 39.08 ± 0.02235 and 21.70 ± 0.1865 mg/L, respectively, but remained above their applicable limits (P < 0.0001). Zn, Pb, and Cd decreased after RO but remained above their respective limits; Cr was 0.0200 ± 0.0006 mg/L at the inlet (P = 0.004) and was not detected after RO at the reported analytical level. Fe remained detectable at 0.1044 ± 0.0001 mg/L before RO and 0.09550 ± 0.0002 mg/L after RO. Al, Be, Ba, As, Ag, Sb, Se, Tl, Co, Ni, Mn, and Cu were not detected at the reported analytical level. Conclusion: RO treatment substantially reduced several measured parameters but did not achieve chemical compliance for Ca, Mg, Zn, Pb, and Cd. The findings support routine monitoring and systematic evaluation of pretreatment, membrane condition, storage, and distribution components.

Science Forum (Journal of Pure and Applied Sciences)
University of Kirkuk (IQ), University of Tikrit (IQ)
Openalex Percentile: Top 23%
Heavy Metal Pollution Remediation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Assessment of Heavy Metal Concentrations and Water Quality Parameters in Hemodialysis Water in Kirkuk, Iraq — Sarhan Ali Salman, Firas Taher Maher, et al. · Science Forum (Journal of Pure and Applied Sciences) (2026) | TGRS Research Map | TGRS