Scoliosis Induction in the Porcine Spine Using a Growth-Modulating Posterior Tether: Finite Element Model Development and Calibration

Abstract Purpose Predictable correction of early onset scoliosis through mechanical modulation of endochondral growth requires understanding vertebral stress–growth relationships. Here, we develop image-based finite element models (FEMs) to estimate vertebral stress–growth relationships under prescribed loading conditions using a porcine model of spine deformity induced mechanically via a posterolateral tether, assuming that, after initial tether-induced elastic deformation, subsequent deformity develops solely from vertebral endochondral growth against the tether and elastic anatomical structures. Methods Image-based FEMs were generated to simulate juvenile Yorkshire pigs instrumented with a posterolateral tether implant which applied asymmetric force across the spine. FEM growth parameters were calibrated to best match in vivo data using landmark points obtained from 3D imaging for each pig. Results Asymmetric stresses generated by loading of the posterolateral offset cable produced multi-planar spine deformity. Calibrated FEMs achieved moderate agreement with frontal plane projections of the deformity: Cobb angle (MAE = 4.6–8.5° across all pigs and timepoints) and anatomic landmark point correspondence (MAE = 2.1–2.9 mm), but sagittal plane agreement was more limited, partly attributable to variability in positioning during imaging. FEM-derived vertebral body growth rates were reduced for vertebrae along the concavity of the deformity. Conclusion Image-based FEMs estimated vertebral stress–growth relationships and reproduced deformity with comparable Cobb angle error to an existing parametric FE model (current 2.5 ± 1.6° vs. reported ~ 2.0 ± 2.4° in a single-pig matched comparison) with realistic baseline IVD stresses (current ~ 0.45 MPa vs. reported 0.48 MPa in the growth plate). With expanded validation, such models may support preclinical evaluation of implant designs and surgical configurations prior to in vivo testing.

Authors

Institutions

Publication Details

Journal
Annals of Biomedical Engineering
Published
2026-10-06
DOI
https://doi.org/10.1007/s10439-026-04403-3
Primary Topic
Scoliosis diagnosis and treatment
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Scoliosis Induction in the Porcine Spine Using a Growth-Modulating Posterior Tether: Finite Element Model Development and Calibration

Thomas P. Schaer, Patrick J. Cahill, Sriram Balasubramanian, Brian D. Snyder et al.
Annals of Biomedical Engineering
Scoliosis diagnosis and treatment
article

Scoliosis Induction in the Porcine Spine Using a Growth-Modulating Posterior Tether: Finite Element Model Development and Calibration

Thomas P. Schaer, Patrick J. Cahill, Sriram Balasubramanian, Brian D. Snyder, Edward J. Vresilovic, Mattan R. Orbach, Christian Reynold D'Andrea, Alessandra Fusco
article en

Abstract

Abstract Purpose Predictable correction of early onset scoliosis through mechanical modulation of endochondral growth requires understanding vertebral stress–growth relationships. Here, we develop image-based finite element models (FEMs) to estimate vertebral stress–growth relationships under prescribed loading conditions using a porcine model of spine deformity induced mechanically via a posterolateral tether, assuming that, after initial tether-induced elastic deformation, subsequent deformity develops solely from vertebral endochondral growth against the tether and elastic anatomical structures. Methods Image-based FEMs were generated to simulate juvenile Yorkshire pigs instrumented with a posterolateral tether implant which applied asymmetric force across the spine. FEM growth parameters were calibrated to best match in vivo data using landmark points obtained from 3D imaging for each pig. Results Asymmetric stresses generated by loading of the posterolateral offset cable produced multi-planar spine deformity. Calibrated FEMs achieved moderate agreement with frontal plane projections of the deformity: Cobb angle (MAE = 4.6–8.5° across all pigs and timepoints) and anatomic landmark point correspondence (MAE = 2.1–2.9 mm), but sagittal plane agreement was more limited, partly attributable to variability in positioning during imaging. FEM-derived vertebral body growth rates were reduced for vertebrae along the concavity of the deformity. Conclusion Image-based FEMs estimated vertebral stress–growth relationships and reproduced deformity with comparable Cobb angle error to an existing parametric FE model (current 2.5 ± 1.6° vs. reported ~ 2.0 ± 2.4° in a single-pig matched comparison) with realistic baseline IVD stresses (current ~ 0.45 MPa vs. reported 0.48 MPa in the growth plate). With expanded validation, such models may support preclinical evaluation of implant designs and surgical configurations prior to in vivo testing.

Annals of Biomedical Engineering
Boston Children's Hospital (US), Beth Israel Deaconess Medical Center (US), Children's Hospital of Philadelphia (US), Harvard University (US), Thomas Jefferson University (US), Drexel University (US), University of Pennsylvania (US), University of Delaware (US)
Openalex Percentile: Top 9%
Scoliosis diagnosis and treatment
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.