Diagnostic performance of Sysmex XN-31 automated hematology analyzer compared to microscopy and PCR for detecting and quantifying malaria parasites in clinical settings of Oromia Regional State, Ethiopia

Malaria remains a major public health challenge in Ethiopia, and improved clinical care supported by innovative diagnostic tools is essential for elimination efforts. The Sysmex XN-31 automated hematology analyzer, which detects and quantifies malaria-infected red blood cells, was evaluated in clinical settings for malaria diagnosis in Ethiopia. A hospital-based cross-sectional study was conducted from April to September 2024 at two public hospitals in Ethiopia. Febrile outpatients suspected of malaria were enrolled after providing consent. Capillary blood was collected for routine microscopy, and 4 mL of venous EDTA blood was obtained for rapid diagnostic tests (RDTs), Sysmex XN-31 analysis, and qPCR. Sysmex XN-31 performance was assessed against expert microscopy and qPCR. Statistical analyses were performed using SPSS 27 and MedCalc v23. Diagnostic parameters (sensitivity, specificity, and predictive values) were calculated with 95% confidence intervals, and hypothetical community prevalence rates (5%, 25%, and 50%) were modeled. Usability of the Sysmex XN-31 was evaluated at both sites, with statistical significance assessed using Fisher’s exact test. Agreement between tests was measured using kappa statistics. Parasitemia quantification was compared using Bland–Altman and Passing-Bablok regression. Among 400 suspected febrile cases, malaria positivity was 22.2% (n = 89), 21% (n = 84), 19.2% (n = 77), 18.8% (n = 75), and 16.2% (n = 65) by PCR, Sysmex XN-31, expert microscopy, field microscopy, and RDTs, respectively. Sysmex XN-31 sensitivity and specificity were 93.5% (95% CI 88.0, 99.0) and 96.6% (95% CI 94.2, 98.4), compared with expert microscopy. Using PCR as the reference, sensitivity decreased to 84.3% (95% CI 80.7, 87.8), while specificity increased slightly to 97% (95% CI 95.3, 98.7). Under hypothetical community prevalence scenarios, the Sysmex-XN-31 demonstrated high negative and positive predictive values relative to field microscopy and RDTs. Substantial agreement was observed between Sysmex XN-31 and field microscopy [89.8%, kappa = 0.68 (0.61–0.80)]. The analyzer also showed strong near-perfect agreement with RDTs [92.5%, kappa = 0.74 (0.61–0.80)] and PCR (93.8% agreement, kappa = 0.81 [0.81–1.00]). Bland–Altman and Passing-Bablok regressions showed no significant differences ( p > 0.05) in parasite quantification between Sysmex XN-31 and microscopy. Usability testing showed that the analyzer can be operated with minimal training, with no statistically significant differences across educational backgrounds or professions (Fisher’s exact test, p > 0.05). The Sysmex XN-31 automated hematology analyzer demonstrated high sensitivity and specificity compared to expert microscopy and PCR, offering substantial value for malaria surveillance and elimination efforts. Although further calibration was needed for parasite quantification, the analyzer provided rapid malaria detection alongside routine CBC results.

Authors

Institutions

Publication Details

Journal
Malaria Journal
Published
2026-10-03
DOI
https://doi.org/10.1186/s12936-026-06174-7
Primary Topic
Malaria Research and Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Diagnostic performance of Sysmex XN-31 automated hematology analyzer compared to microscopy and PCR for detecting and quantifying malaria parasites in clinical settings of Oromia Regional State, Ethiopia

Mahlet Belachew, Hailemariam Difabachew, Hiwot Solomon, Bokretsion Gidey Brhane et al.
Malaria Journal
Malaria Research and Control
article

Diagnostic performance of Sysmex XN-31 automated hematology analyzer compared to microscopy and PCR for detecting and quantifying malaria parasites in clinical settings of Oromia Regional State, Ethiopia

Mahlet Belachew, Hailemariam Difabachew, Hiwot Solomon, Bokretsion Gidey Brhane, Yonas Wuletaw, Dawit Wolday, Getachew Tollera, Abnet Abebe, Daniel Melese, Aster Tsegaye, Abeba G Tsadik, Geremew Tasew, Mesay Hailu, Gemechu Tadese, Adugna Woyessa, Adugna Abera, Selam Yirga, Gudissa Asefa
article en

Abstract

Malaria remains a major public health challenge in Ethiopia, and improved clinical care supported by innovative diagnostic tools is essential for elimination efforts. The Sysmex XN-31 automated hematology analyzer, which detects and quantifies malaria-infected red blood cells, was evaluated in clinical settings for malaria diagnosis in Ethiopia. A hospital-based cross-sectional study was conducted from April to September 2024 at two public hospitals in Ethiopia. Febrile outpatients suspected of malaria were enrolled after providing consent. Capillary blood was collected for routine microscopy, and 4 mL of venous EDTA blood was obtained for rapid diagnostic tests (RDTs), Sysmex XN-31 analysis, and qPCR. Sysmex XN-31 performance was assessed against expert microscopy and qPCR. Statistical analyses were performed using SPSS 27 and MedCalc v23. Diagnostic parameters (sensitivity, specificity, and predictive values) were calculated with 95% confidence intervals, and hypothetical community prevalence rates (5%, 25%, and 50%) were modeled. Usability of the Sysmex XN-31 was evaluated at both sites, with statistical significance assessed using Fisher’s exact test. Agreement between tests was measured using kappa statistics. Parasitemia quantification was compared using Bland–Altman and Passing-Bablok regression. Among 400 suspected febrile cases, malaria positivity was 22.2% (n = 89), 21% (n = 84), 19.2% (n = 77), 18.8% (n = 75), and 16.2% (n = 65) by PCR, Sysmex XN-31, expert microscopy, field microscopy, and RDTs, respectively. Sysmex XN-31 sensitivity and specificity were 93.5% (95% CI 88.0, 99.0) and 96.6% (95% CI 94.2, 98.4), compared with expert microscopy. Using PCR as the reference, sensitivity decreased to 84.3% (95% CI 80.7, 87.8), while specificity increased slightly to 97% (95% CI 95.3, 98.7). Under hypothetical community prevalence scenarios, the Sysmex-XN-31 demonstrated high negative and positive predictive values relative to field microscopy and RDTs. Substantial agreement was observed between Sysmex XN-31 and field microscopy [89.8%, kappa = 0.68 (0.61–0.80)]. The analyzer also showed strong near-perfect agreement with RDTs [92.5%, kappa = 0.74 (0.61–0.80)] and PCR (93.8% agreement, kappa = 0.81 [0.81–1.00]). Bland–Altman and Passing-Bablok regressions showed no significant differences ( p > 0.05) in parasite quantification between Sysmex XN-31 and microscopy. Usability testing showed that the analyzer can be operated with minimal training, with no statistically significant differences across educational backgrounds or professions (Fisher’s exact test, p > 0.05). The Sysmex XN-31 automated hematology analyzer demonstrated high sensitivity and specificity compared to expert microscopy and PCR, offering substantial value for malaria surveillance and elimination efforts. Although further calibration was needed for parasite quantification, the analyzer provided rapid malaria detection alongside routine CBC results.

Malaria Journal
Government of Ethiopia (ET), Population Health Research Institute (CA), Ethiopian Public Health Institute (ET), Federal Ministry of Health (ET), Addis Ababa University (ET), McMaster University (CA)
Openalex Percentile: Top 9%
Malaria Research and Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.