Improving tumor marker comparability: A case series on harmonizing carbohydrate antigen 19‐9 and carbohydrate antigen 15‐3 measurements across diverse measurement systems

Abstract Serum tumor markers play a critical role in the early screening, treatment monitoring, and prognostic assessment of malignant tumors, yet poor comparability across different measurement systems frequently undermines their clinical value. We evaluated the comparability of carbohydrate antigen 19‐9 (CA19‐9) and carbohydrate antigen 15‐3 (CA15‐3) across seven contemporary measurement systems—Abbott Alinity I, Autobio A2000PLUS, Beckman Access Dxl 800, Mindray CL6000i, Roche Cobas E801, Snibe Maglumi X8, and Sysmex HISCL‐5000 using 464 native serum samples for CA19‐9 and 499 for CA15‐3. Spearman correlations were higher for CA15‑3 ( r : 0.927‒0.997) than for CA19‑9 ( r : 0.777‒0.985). With the all‐method median as a pragmatic comparator, Passing‒Bablok regression and difference plots were generated for each system. Against analytical performance specifications derived from biological variation, all systems except Abbott Alinity (22.5%) met the minimum bias criterion of ±21.6% for CA19‐9, whereas for CA15‐3 only Abbott Alinity (‒10.159%), Roche Cobas (13.01%), and Snibe Maglumi (−3.923%) fell within the minimum bias criterion of ±13.9%. CA19‐9 and CA15‐3 concentrations were recalibrated using three mathematical approaches: principal component analysis and two nonlinear regression models. After recalibration, the variability of CA19‐9 and CA15‐3 values decreased significantly. Our findings indicate that CA19‐9 and CA15‐3 results are not directly interchangeable across different systems, which poses a real‐world challenge for longitudinal patient follow‐up. However, mathematical models derived from individual patient serum samples can effectively reduce system‐related discrepancies, thereby enabling more consistent serial interpretations and supporting more informed clinical decisions in routine practice.

Authors

Institutions

Publication Details

Journal
Open Research (University of Surrey)
Published
2026-09-29
DOI
https://doi.org/10.1002/viw2.70207
Primary Topic
Glycosylation and Glycoproteins Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
article

Improving tumor marker comparability: A case series on harmonizing carbohydrate antigen 19‐9 and carbohydrate antigen 15‐3 measurements across diverse measurement systems

Zongdan Wang, Xiaoying Lou, Meng Li, Jing Zhang et al.
Open Research (University of Surrey)
Glycosylation and Glycoproteins Research
article

Improving tumor marker comparability: A case series on harmonizing carbohydrate antigen 19‐9 and carbohydrate antigen 15‐3 measurements across diverse measurement systems

Zongdan Wang, Xiaoying Lou, Meng Li, Jing Zhang, Jia Li, Wei Cui, Mengdi Wu, Huishuang Gao, Kuo Zhang, Bing Wu, Dongmei Ma
article en

Abstract

Abstract Serum tumor markers play a critical role in the early screening, treatment monitoring, and prognostic assessment of malignant tumors, yet poor comparability across different measurement systems frequently undermines their clinical value. We evaluated the comparability of carbohydrate antigen 19‐9 (CA19‐9) and carbohydrate antigen 15‐3 (CA15‐3) across seven contemporary measurement systems—Abbott Alinity I, Autobio A2000PLUS, Beckman Access Dxl 800, Mindray CL6000i, Roche Cobas E801, Snibe Maglumi X8, and Sysmex HISCL‐5000 using 464 native serum samples for CA19‐9 and 499 for CA15‐3. Spearman correlations were higher for CA15‑3 ( r : 0.927‒0.997) than for CA19‑9 ( r : 0.777‒0.985). With the all‐method median as a pragmatic comparator, Passing‒Bablok regression and difference plots were generated for each system. Against analytical performance specifications derived from biological variation, all systems except Abbott Alinity (22.5%) met the minimum bias criterion of ±21.6% for CA19‐9, whereas for CA15‐3 only Abbott Alinity (‒10.159%), Roche Cobas (13.01%), and Snibe Maglumi (−3.923%) fell within the minimum bias criterion of ±13.9%. CA19‐9 and CA15‐3 concentrations were recalibrated using three mathematical approaches: principal component analysis and two nonlinear regression models. After recalibration, the variability of CA19‐9 and CA15‐3 values decreased significantly. Our findings indicate that CA19‐9 and CA15‐3 results are not directly interchangeable across different systems, which poses a real‐world challenge for longitudinal patient follow‐up. However, mathematical models derived from individual patient serum samples can effectively reduce system‐related discrepancies, thereby enabling more consistent serial interpretations and supporting more informed clinical decisions in routine practice.

Open Research (University of Surrey)
Beijing University of Chinese Medicine (CN), Chinese Academy of Medical Sciences & Peking Union Medical College (CN), Shandong Provincial Hospital (CN), Cancer Hospital of Chinese Academy of Medical Sciences (CN), National Center for Clinical Laboratories (CN), Beijing Chuiyangliu Hospital (CN)
Openalex Percentile: Top 19%
Glycosylation and Glycoproteins Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.