Unintended Right Atrial Lesions After Monopolar Biphasic Lattice-tip Pulsed-Field Ablation: A Case Series

Structured Abstract Background Monopolar biphasic pulsed-field ablation (PFA) delivered by a lattice-tip catheter may create deeper lesions than conventional bipolar systems. The impact on adjacent right atrial structures remains poorly characterized. Case summary We report four patients undergoing pulmonary vein isolation using a lattice-tip PFA catheter. Post-procedural electroanatomic mapping revealed unintended right atrial lesions involving the posterior right atrium, interatrial septum, or superior vena cava. In patients with short anatomical distances between right pulmonary veins and adjacent right atrial structures, lesions developed despite apparently successful pulmonary vein isolation. Catheter instability related to coughing under deep sedation appeared to contribute to broader lesion formation in two cases. Conclusion These cases suggest that monopolar biphasic lattice-tip PFA may create unintended lesions in adjacent right atrial tissue when anatomical proximity is short or catheter stability is compromised. Awareness of these mechanisms may help optimize procedural safety and lesion delivery.

Authors

Institutions

Publication Details

Journal
European Heart Journal - Case Reports
Published
2026-09-29
DOI
https://doi.org/10.1093/ehjcr/ytag730
Primary Topic
Cardiac Arrhythmias and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
article

Unintended Right Atrial Lesions After Monopolar Biphasic Lattice-tip Pulsed-Field Ablation: A Case Series

Masato Fukunaga, Jun Hirokami, Michio Nagashima, Kengo Korai et al.
European Heart Journal - Case Reports
Cardiac Arrhythmias and Treatments
article

Unintended Right Atrial Lesions After Monopolar Biphasic Lattice-tip Pulsed-Field Ablation: A Case Series

Masato Fukunaga, Jun Hirokami, Michio Nagashima, Kengo Korai, Kenji Andò, 廣島 謙一
article en

Abstract

Structured Abstract Background Monopolar biphasic pulsed-field ablation (PFA) delivered by a lattice-tip catheter may create deeper lesions than conventional bipolar systems. The impact on adjacent right atrial structures remains poorly characterized. Case summary We report four patients undergoing pulmonary vein isolation using a lattice-tip PFA catheter. Post-procedural electroanatomic mapping revealed unintended right atrial lesions involving the posterior right atrium, interatrial septum, or superior vena cava. In patients with short anatomical distances between right pulmonary veins and adjacent right atrial structures, lesions developed despite apparently successful pulmonary vein isolation. Catheter instability related to coughing under deep sedation appeared to contribute to broader lesion formation in two cases. Conclusion These cases suggest that monopolar biphasic lattice-tip PFA may create unintended lesions in adjacent right atrial tissue when anatomical proximity is short or catheter stability is compromised. Awareness of these mechanisms may help optimize procedural safety and lesion delivery.

European Heart Journal - Case Reports
Kokura Memorial Hospital (JP)
Good health and well-being
Openalex Percentile: Top 11%
Cardiac Arrhythmias and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Unintended Right Atrial Lesions After Monopolar Biphasic Lattice-tip Pulsed-Field Ablation: A Case Series — Masato Fukunaga, Jun Hirokami, et al. · European Heart Journal - Case Reports (2026) | TGRS Research Map | TGRS