Continuous monitoring of endotracheal tube obstructions using naturally occurring pressure and flow oscillations

Abstract Background The endotracheal tube resistance dominates the total airway resistance in most intubated patients. Mucus deposition and biofilm formation can rapidly increase tube resistance and thereby contribute to serious ventilatory impairments, including dynamic hyperinflation, intrinsic PEEP build-up, added work of breathing, and patient-ventilator asynchrony. During controlled mechanical ventilation, an increased tube resistance can be inferred from the difference between peak and plateau pressure, but this approach fails during pressure-supported spontaneous breathing. Methods We present a method that estimates the linear and nonlinear components of tube resistance from naturally occurring airway pressure and flow fluctuations at the airway opening, without a tracheal pressure sensor and without imposed forced oscillations. This is achieved by solving the equation of motion using band-pass filtered airway pressure and flow signals. Band-pass filtering isolates the relevant resistive and inertive pressure losses across the tube by removing slow contributions from muscle pressure and lung elastance as well as high-frequency noise. Results Across different ventilation modes, ventilator settings, tube diameters, filters, and breathing patterns, the method recovers both linear and nonlinear tube resistance parameters with < 10% error and < 2% bias. The method can also track progressive tube obstruction on a breath-by-breath basis. Conclusions Continuous estimation of the endotracheal tube resistance from naturally occurring, ventilator-induced airway pressure and flow fluctuations enables breath-by-breath detection of tube obstruction and provides the parameters required for automatic tube resistance compensation during spontaneous breathing. This approach may improve patient safety, reduce patient-ventilator asynchrony, and facilitate weaning.

Authors

Institutions

Publication Details

Journal
Intensive Care Medicine Experimental
Published
2026-09-29
DOI
https://doi.org/10.1186/s40635-026-00982-6
Primary Topic
Respiratory Support and Mechanisms
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
article

Continuous monitoring of endotracheal tube obstructions using naturally occurring pressure and flow oscillations

Ben Fabry, Roland C. E. Francis, Christian Kuster
Intensive Care Medicine Experimental
Respiratory Support and Mechanisms
article

Continuous monitoring of endotracheal tube obstructions using naturally occurring pressure and flow oscillations

Ben Fabry, Roland C. E. Francis, Christian Kuster
article en

Abstract

Abstract Background The endotracheal tube resistance dominates the total airway resistance in most intubated patients. Mucus deposition and biofilm formation can rapidly increase tube resistance and thereby contribute to serious ventilatory impairments, including dynamic hyperinflation, intrinsic PEEP build-up, added work of breathing, and patient-ventilator asynchrony. During controlled mechanical ventilation, an increased tube resistance can be inferred from the difference between peak and plateau pressure, but this approach fails during pressure-supported spontaneous breathing. Methods We present a method that estimates the linear and nonlinear components of tube resistance from naturally occurring airway pressure and flow fluctuations at the airway opening, without a tracheal pressure sensor and without imposed forced oscillations. This is achieved by solving the equation of motion using band-pass filtered airway pressure and flow signals. Band-pass filtering isolates the relevant resistive and inertive pressure losses across the tube by removing slow contributions from muscle pressure and lung elastance as well as high-frequency noise. Results Across different ventilation modes, ventilator settings, tube diameters, filters, and breathing patterns, the method recovers both linear and nonlinear tube resistance parameters with < 10% error and < 2% bias. The method can also track progressive tube obstruction on a breath-by-breath basis. Conclusions Continuous estimation of the endotracheal tube resistance from naturally occurring, ventilator-induced airway pressure and flow fluctuations enables breath-by-breath detection of tube obstruction and provides the parameters required for automatic tube resistance compensation during spontaneous breathing. This approach may improve patient safety, reduce patient-ventilator asynchrony, and facilitate weaning.

Intensive Care Medicine ExperimentalVol. 14(1)
Friedrich-Alexander-Universität Erlangen-Nürnberg (DE), Universitätsklinikum Erlangen (DE)
Openalex Percentile: Top 12%
Respiratory Support and Mechanisms
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.