Pupil size measurements with two multifunctional devices based on different technologies and an infrared based pupillometer

Abstract Purpose To compare inter- and intra-device agreement of pupil size (PS) measurements of an aberrometer/coherence interferometer/tomographer (Pentacam AXL Wave (OCULUS)), optical coherence tomographer/placido topographer (MS-39, (CSO/Schwind)) and an infrared based pupillometer (PupilX (Albomed)). Methods In this prospective study, PS of healthy eyes was measured with Pentacam without (WO) and with glare (WG), PupilX in 0, 0.06, 0.12, 4 and 64 lx and MS-39 in 0.04, 4 and 50 lx three times each. Scotopic conditions were defined as WO and ≤ 0.12 lx, mesopic as 4 lx and photopic as WG, 50 and 64 lx. Measurements were compared per lighting condition. Results Thirty-seven eyes of 37 participants (mean age 30.8 years ± 9.1 years) were included. Pentacam WG and PupilX at 64 lx showed the lowest inter-device difference and had the only non-significant difference ( p = 0.204). The second lowest mean difference was between Pentacam WO and PupilX at 0.06 lx where significance was also not met ( p = 0.05). Pentacam WO and PupilX at 0 lx showed the highest inter-device difference ( p < 0.001). The lowest intra-device difference (IDD) for scotopic conditions was 0.19 mm (standard deviation (SD): 0.17 mm) with PupilX at 0 lx. In mesopic conditions, the lowest IDD was 0.2 mm (SD: 0.15 mm) with MS-39, in photopic conditions 0.16 mm (SD: 0.11 mm) with PupilX. Conclusion Pentacam showed close-to-acceptable agreement with PupilX in photopic conditions. In the other conditions significant differences were found, which applied for MS-39 and PupilX for all illumination levels. All three devices showed excellent intra-device repeatability for all illuminations levels.

Authors

Institutions

Publication Details

Journal
Graefe s Archive for Clinical and Experimental Ophthalmology
Published
2026-09-29
DOI
https://doi.org/10.1007/s00417-026-07509-5
Primary Topic
Glaucoma and retinal disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
article

Pupil size measurements with two multifunctional devices based on different technologies and an infrared based pupillometer

Petra Dávidová, Julian Bucur, Klemens Paul Kaiser, Thomas Kohnen et al.
Graefe s Archive for Clinical and Experimental Ophthalmology
Glaucoma and retinal disorders
article

Pupil size measurements with two multifunctional devices based on different technologies and an infrared based pupillometer

Petra Dávidová, Julian Bucur, Klemens Paul Kaiser, Thomas Kohnen, Jakob Wend, Asma Achourak
article en

Abstract

Abstract Purpose To compare inter- and intra-device agreement of pupil size (PS) measurements of an aberrometer/coherence interferometer/tomographer (Pentacam AXL Wave (OCULUS)), optical coherence tomographer/placido topographer (MS-39, (CSO/Schwind)) and an infrared based pupillometer (PupilX (Albomed)). Methods In this prospective study, PS of healthy eyes was measured with Pentacam without (WO) and with glare (WG), PupilX in 0, 0.06, 0.12, 4 and 64 lx and MS-39 in 0.04, 4 and 50 lx three times each. Scotopic conditions were defined as WO and ≤ 0.12 lx, mesopic as 4 lx and photopic as WG, 50 and 64 lx. Measurements were compared per lighting condition. Results Thirty-seven eyes of 37 participants (mean age 30.8 years ± 9.1 years) were included. Pentacam WG and PupilX at 64 lx showed the lowest inter-device difference and had the only non-significant difference ( p = 0.204). The second lowest mean difference was between Pentacam WO and PupilX at 0.06 lx where significance was also not met ( p = 0.05). Pentacam WO and PupilX at 0 lx showed the highest inter-device difference ( p < 0.001). The lowest intra-device difference (IDD) for scotopic conditions was 0.19 mm (standard deviation (SD): 0.17 mm) with PupilX at 0 lx. In mesopic conditions, the lowest IDD was 0.2 mm (SD: 0.15 mm) with MS-39, in photopic conditions 0.16 mm (SD: 0.11 mm) with PupilX. Conclusion Pentacam showed close-to-acceptable agreement with PupilX in photopic conditions. In the other conditions significant differences were found, which applied for MS-39 and PupilX for all illumination levels. All three devices showed excellent intra-device repeatability for all illuminations levels.

Graefe s Archive for Clinical and Experimental Ophthalmology
Goethe University Frankfurt (DE)
Openalex Percentile: Top 9%
Glaucoma and retinal disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.