Revisiting Artifact Suppression in Electrospray Ionization Deconvolution
Abstract Charge-state assignment is a critical step in many workflows for intact protein analysis by mass spectrometry. Computational deconvolution algorithms are increasingly used to automate data analysis, but conventional approaches can introduce artifacts. Here, we explore several approaches to removing artifacts by limiting the charge-state assignments for each data point. One approach removes satellite charge states surrounding the local maxima. Another removes harmonics at double the more probable charge assignment. The last two limit either the total number of allowed charge-state assignments or the minimum relative charge-state assignment probability for a given data point. These new features are as effective as the prior SoftMax artifact removal in UniDec, but they enable easier tuning by the user. Although the new set of features introduces additional complexity and an expanded number of parameters, these tools enable users to more precisely reduce artifacts in the deconvolution results.