Development of a prospective clinical workflow for proton beam therapy of ventricular tachycardia

Abstract Purpose This study presents a prospective preclinical workflow for proton beam therapy of refractory ventricular tachycardia (VT), covering target transfer, treatment planning (TP), and beam gating. Methods Photon-based stereotactic arrhythmia radioablation (STAR) protocols were adapted to proton beam therapy, encompassing (i) target transfer from electroanatomical mapping (EAM) onto planning computed tomography (CT), (ii) simulation of three TP cases from the Standardized Treatment and Outcome Platform for Stereotactic Therapy of Re-Entrant Tachycardia by a Multidisciplinary (STOPSTORM.eu) benchmark, and (iii) evaluation of trigger events and delays for synchrotron beam delivery control using gating. Results CARDIO-RT enabled electroanatomical mapping to computed tomography (EAM-to-CT) target transfer; two experienced cardiologists qualitatively judged the transferred targets as clinically acceptable by visual inspection. Treatment plans demonstrated effective organ at risk (OAR) sparing and clinical adequacy. The planning target volume (PTV) D 95% ranged from 23.3 to 23.9 Gy relative biological effectiveness (RBE) and the clinical target volume electrophysiology (CTV [EP]) D 95% from 30.9 to 31.6 Gy (RBE); underdosage was permitted where anatomical compromise was unavoidable. No CTV(EP) hotspot exceeded 33.5 Gy (RBE). Optical trigger signals are supported for beam gating; the total cardiac gating chain latency was below 30 ms. Conclusion This technical feasibility study demonstrates a prospective preclinical workflow for proton beam STAR, covering EAM-to-CT target transfer, TP across three benchmark clinical cases, and accelerator-side optical gating integration at a synchrotron-based facility. The workflow demonstrates technical feasibility; however, full clinical translation still requires patient positioning and verification, motion characterization, and end-to-end validation. Future work will employ a dynamic anthropomorphic phantom to characterize motion-induced dosimetric changes and assess plan robustness for proton and carbon ion beam delivery.

Authors

Institutions

Publication Details

Journal
Strahlentherapie und Onkologie
Published
2026-09-28
DOI
https://doi.org/10.1007/s00066-026-02599-w
Primary Topic
Cardiac Arrhythmias and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
article

Development of a prospective clinical workflow for proton beam therapy of ventricular tachycardia

Oliver Blanck, Martina Fuß, Melanie Grehn, Jingyang Xie et al.
Strahlentherapie und Onkologie
Cardiac Arrhythmias and Treatments
article

Development of a prospective clinical workflow for proton beam therapy of ventricular tachycardia

Oliver Blanck, Martina Fuß, Melanie Grehn, Jingyang Xie, Valeria Trojani, Markus Stock, Ariadna Cherit Hernández, Dietmar Georg, Piero Fossati, Lukas Fiedler, Günter Gruber-Hammer, Claus Schmitzer
article en

Abstract

Abstract Purpose This study presents a prospective preclinical workflow for proton beam therapy of refractory ventricular tachycardia (VT), covering target transfer, treatment planning (TP), and beam gating. Methods Photon-based stereotactic arrhythmia radioablation (STAR) protocols were adapted to proton beam therapy, encompassing (i) target transfer from electroanatomical mapping (EAM) onto planning computed tomography (CT), (ii) simulation of three TP cases from the Standardized Treatment and Outcome Platform for Stereotactic Therapy of Re-Entrant Tachycardia by a Multidisciplinary (STOPSTORM.eu) benchmark, and (iii) evaluation of trigger events and delays for synchrotron beam delivery control using gating. Results CARDIO-RT enabled electroanatomical mapping to computed tomography (EAM-to-CT) target transfer; two experienced cardiologists qualitatively judged the transferred targets as clinically acceptable by visual inspection. Treatment plans demonstrated effective organ at risk (OAR) sparing and clinical adequacy. The planning target volume (PTV) D 95% ranged from 23.3 to 23.9 Gy relative biological effectiveness (RBE) and the clinical target volume electrophysiology (CTV [EP]) D 95% from 30.9 to 31.6 Gy (RBE); underdosage was permitted where anatomical compromise was unavoidable. No CTV(EP) hotspot exceeded 33.5 Gy (RBE). Optical trigger signals are supported for beam gating; the total cardiac gating chain latency was below 30 ms. Conclusion This technical feasibility study demonstrates a prospective preclinical workflow for proton beam STAR, covering EAM-to-CT target transfer, TP across three benchmark clinical cases, and accelerator-side optical gating integration at a synchrotron-based facility. The workflow demonstrates technical feasibility; however, full clinical translation still requires patient positioning and verification, motion characterization, and end-to-end validation. Future work will employ a dynamic anthropomorphic phantom to characterize motion-induced dosimetric changes and assess plan robustness for proton and carbon ion beam delivery.

Strahlentherapie und Onkologie
Danube Private University (AT), Universität für Weiterbildung Krems (AT), Fachhochschule Wiener Neustadt (AT), Azienda Sanitaria Unità Locale di Reggio Emilia (IT), University Hospital Schleswig-Holstein (DE), MedAustron (AT), Universitätsklinikum Wiener Neustadt (AT), Medical University of Vienna (AT), University of Lübeck (DE)
Openalex Percentile: Top 11%
Cardiac Arrhythmias and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.