Measurement properties and operational feasibility of an automated arterial occlusion pressure assessment system for blood flow restriction exercise

OBJECTIVE: To evaluate the validity, reliability, agreement, and operational feasibility of an automated arterial occlusion pressure (AOP) assessment system compared with a manual Doppler-based assessment (MDA). APPROACH: Forty healthy adults (20 men and 20 women) completed upper- and lower-limb AOP assessments using automated and MDA methods in a randomized and counterbalanced design. Concurrent validity, agreement, intraday and interday reliability, operational feasibility, perceptual responses, and procedure-related symptoms were evaluated. Exploratory analyses examined whether sex modified the difference between automated and MDA-derived AOP measurements. MAIN RESULTS: MDA demonstrated good-to-excellent reliability (ICC = 0.84-0.98; CV = 1.38-3.25%). The automated system demonstrated moderate-to-good intraday reliability (ICC = 0.74-0.85) but poor lower-limb interday reliability (ICC = 0.36). Compared with MDA, automated AOP showed systematic bias and wide limits of agreement. No evidence of a sex-related difference in the between-method discrepancy was observed for either upper- or lower-limb AOP. Although automated assessment reduced testing time by 81.9-86.7%, frequent upper-limb measurement failures reduced the automated measurement success rate to approximately 57%, substantially increasing adjusted assessment time and limiting operational feasibility, while discomfort and affective responses were more favorable than with MDA. SIGNIFICANCE: The automated AOP system improved assessment efficiency and participant experience but demonstrated less favorable reliability estimates, limited agreement, and frequent upper-limb measurement failures. Consequently, automated and MDA should not be considered interchangeable for individualized AOP prescription.

Authors

Institutions

Publication Details

Journal
Physiological Measurement
Published
2026-09-18
DOI
https://doi.org/10.1088/1361-6579/aea9ed
Primary Topic
Cardiovascular and exercise physiology
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
article

Measurement properties and operational feasibility of an automated arterial occlusion pressure assessment system for blood flow restriction exercise

Mikhail Santos Cerqueira, Nicholas Rolnick, Joanderson Felipe Soares Silva, Maria Luiza Araujo Cerqueira et al.
Physiological Measurement
Cardiovascular and exercise physiology
article

Measurement properties and operational feasibility of an automated arterial occlusion pressure assessment system for blood flow restriction exercise

Mikhail Santos Cerqueira, Nicholas Rolnick, Joanderson Felipe Soares Silva, Maria Luiza Araujo Cerqueira, Gabriela Mota Pinto Silva, Rafael Pereira
article en

Abstract

OBJECTIVE: To evaluate the validity, reliability, agreement, and operational feasibility of an automated arterial occlusion pressure (AOP) assessment system compared with a manual Doppler-based assessment (MDA). APPROACH: Forty healthy adults (20 men and 20 women) completed upper- and lower-limb AOP assessments using automated and MDA methods in a randomized and counterbalanced design. Concurrent validity, agreement, intraday and interday reliability, operational feasibility, perceptual responses, and procedure-related symptoms were evaluated. Exploratory analyses examined whether sex modified the difference between automated and MDA-derived AOP measurements. MAIN RESULTS: MDA demonstrated good-to-excellent reliability (ICC = 0.84-0.98; CV = 1.38-3.25%). The automated system demonstrated moderate-to-good intraday reliability (ICC = 0.74-0.85) but poor lower-limb interday reliability (ICC = 0.36). Compared with MDA, automated AOP showed systematic bias and wide limits of agreement. No evidence of a sex-related difference in the between-method discrepancy was observed for either upper- or lower-limb AOP. Although automated assessment reduced testing time by 81.9-86.7%, frequent upper-limb measurement failures reduced the automated measurement success rate to approximately 57%, substantially increasing adjusted assessment time and limiting operational feasibility, while discomfort and affective responses were more favorable than with MDA. SIGNIFICANCE: The automated AOP system improved assessment efficiency and participant experience but demonstrated less favorable reliability estimates, limited agreement, and frequent upper-limb measurement failures. Consequently, automated and MDA should not be considered interchangeable for individualized AOP prescription.

Physiological Measurement
Southwest Bahia State University (BR), Bundesverband Freier Radios (DE)
Openalex Percentile: Top 6%
Cardiovascular and exercise physiology
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.