Impact of Aortic Annulus Size on Clinical and Hemodynamic Outcomes of Transcatheter Aortic Valve Implantation With the Navitor Valve

ABSTRACT Background Patients with small aortic annuli undergoing transcatheter aortic valve implantation (TAVI) are at risk of higher residual transvalvular gradients. Evidence on the performance of new‐generation intra‐annular self‐expanding valves in this anatomical setting remains limited. Aims To evaluate the impact of annular size on procedural, hemodynamic, and mid‐term clinical outcomes after TAVI with the Navitor valve. Methods Patients were stratified by annular area measured by multi‐slice computed tomography, defining a small annulus as ≤ 430 mm 2 . Primary endpoints at 12 months were: (i) a clinical composite of all‐cause death, disabling stroke, or heart failure rehospitalization, and (ii) a composite endpoint of bioprosthetic valve dysfunction. Secondary endpoints included VARC‐3–defined procedural and early outcomes and valve hemodynamic performance. Results The study population comprised 218 patients, of whom 91 (41.7%) had a small annulus. VARC‐3 technical success was high and comparable between small and non‐small annuli (96.7% vs. 96.1%). Patients with small annuli had higher early post‐procedural mean transvalvular gradients (7.7 ± 3.8 vs. 6.6 ± 3.0 mmHg; p = 0.029) and smaller effective orifice areas index (1.1 ± 0.3 vs. 1.3 ± 0.2 cm 2 /m 2 ; p = 0.037), while the incidence of mean gradients ≥ 20 mmHg remained low (1.1% vs. 0.0%; p = 0.432). At 12 months, the clinical composite endpoint occurred in 8.8% of patients with small annuli and 10.2% with non‐small annuli ( p = 0.759). Bioprosthetic valve dysfunction was infrequent and similar between groups (6.6% vs. 6.3%; p = 0.875). Conclusions TAVI with the Navitor intra‐annular self‐expanding valve was associated with high procedural success, favorable valve hemodynamics, and comparable mid‐term clinical outcomes irrespective of annular size.

Authors

Institutions

Publication Details

Journal
Catheterization and Cardiovascular Interventions
Published
2026-09-17
DOI
https://doi.org/10.1002/ccd.70879
Primary Topic
Cardiac Valve Diseases and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
article

Impact of Aortic Annulus Size on Clinical and Hemodynamic Outcomes of Transcatheter Aortic Valve Implantation With the Navitor Valve

Andrea Berni, Emanuele Gallinoro, Sara Corradetti, Matteo Casenghi et al.
Catheterization and Cardiovascular Interventions
Cardiac Valve Diseases and Treatments
article

Impact of Aortic Annulus Size on Clinical and Hemodynamic Outcomes of Transcatheter Aortic Valve Implantation With the Navitor Valve

Andrea Berni, Emanuele Gallinoro, Sara Corradetti, Matteo Casenghi, Linda Piras, Liesbeth Rosseel, Stefano Rigattieri, Marc Vanderheyden, Marta Belmonte, Eric Wyffels, Pasquale Paolisso, Carlo Barsali, Daniele Barzetti, Emanuele Barbato, Rita Tufo, Emiliano Fiori
article en

Abstract

ABSTRACT Background Patients with small aortic annuli undergoing transcatheter aortic valve implantation (TAVI) are at risk of higher residual transvalvular gradients. Evidence on the performance of new‐generation intra‐annular self‐expanding valves in this anatomical setting remains limited. Aims To evaluate the impact of annular size on procedural, hemodynamic, and mid‐term clinical outcomes after TAVI with the Navitor valve. Methods Patients were stratified by annular area measured by multi‐slice computed tomography, defining a small annulus as ≤ 430 mm 2 . Primary endpoints at 12 months were: (i) a clinical composite of all‐cause death, disabling stroke, or heart failure rehospitalization, and (ii) a composite endpoint of bioprosthetic valve dysfunction. Secondary endpoints included VARC‐3–defined procedural and early outcomes and valve hemodynamic performance. Results The study population comprised 218 patients, of whom 91 (41.7%) had a small annulus. VARC‐3 technical success was high and comparable between small and non‐small annuli (96.7% vs. 96.1%). Patients with small annuli had higher early post‐procedural mean transvalvular gradients (7.7 ± 3.8 vs. 6.6 ± 3.0 mmHg; p = 0.029) and smaller effective orifice areas index (1.1 ± 0.3 vs. 1.3 ± 0.2 cm 2 /m 2 ; p = 0.037), while the incidence of mean gradients ≥ 20 mmHg remained low (1.1% vs. 0.0%; p = 0.432). At 12 months, the clinical composite endpoint occurred in 8.8% of patients with small annuli and 10.2% with non‐small annuli ( p = 0.759). Bioprosthetic valve dysfunction was infrequent and similar between groups (6.6% vs. 6.3%; p = 0.875). Conclusions TAVI with the Navitor intra‐annular self‐expanding valve was associated with high procedural success, favorable valve hemodynamics, and comparable mid‐term clinical outcomes irrespective of annular size.

Catheterization and Cardiovascular Interventions
Onze Lieve Vrouwziekenhuis Hospital (BE), CTO Andrea Alesini (IT), Flanders Environment Agency (BE), Sapienza University of Rome (IT)
Good health and well-being
Openalex Percentile: Top 11%
Cardiac Valve Diseases and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.