Performance of the i-SENS CareSens Air CGM System in Compliance with the European CGM (eCGM) Clinical Performance Criteria: A Pooled Analysis

The eCGM designation is a recently proposed framework for the approval of continuous glucose monitoring (CGM) systems in Europe. The aim of this analysis was to assess whether the performance of the CareSens Air CGM system (i-SENS, Inc., Republic of Korea), derived from a pooled dataset comprising three clinical investigations, complies with eCGM requirements. A total of 164 participants (86% type 1 diabetes, 50% female, 66% on multiple daily injections) across the three studies contributed 17,655 data points to the data analysis. Meal and/or insulin challenges were implemented to obtain sufficient data points at comparator glucose concentrations <70 mg/dL (10.9%) and >300 mg/dL (5.7%). Measurement accuracy was assessed by applying eCGM accuracy requirements. The CareSens Air CGM system complies with eCGM accuracy requirements in this data analysis. Approximately 91.0% of the CGM readings <70 mg/dL were found within ±15 mg/dL from the paired comparator value. At eu- (70–180 mg/dL) and hyperglycemic (>180 mg/dL) CGM readings, 78.6% and 84.4% of CGM readings exhibited differences ≤ ±15%. Overall, 90.3% of CGM readings were found within ±20% of the paired comparator values across the measuring range. Accuracy remained stable throughout the sensor lifetime. The CareSens Air CGM system is suitable for nonadjunctive insulin dosing decisions and integration with automated insulin delivery systems. The three studies whose data were used in this pooled analysis were registered in the Republic of Korea’s Clinical Research Information Service (KCT0009215) and the German Clinical Trials Register (DRKS00028739 and DRKS00037132).

Authors

Institutions

Publication Details

Journal
Diabetes Therapy
Published
2026-09-15
DOI
https://doi.org/10.1007/s13300-026-01917-w
Primary Topic
Diabetes Management and Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
article

Performance of the i-SENS CareSens Air CGM System in Compliance with the European CGM (eCGM) Clinical Performance Criteria: A Pooled Analysis

Manuel Eichenlaub, Won Sang Yoo, Cheol‐Young Park, Şükrü Öter et al.
Diabetes Therapy
Diabetes Management and Research
article

Performance of the i-SENS CareSens Air CGM System in Compliance with the European CGM (eCGM) Clinical Performance Criteria: A Pooled Analysis

Manuel Eichenlaub, Won Sang Yoo, Cheol‐Young Park, Şükrü Öter, Kyung‐Soo Kim, Nina Jendrike, Lorena Morent, Seung-Hwan Lee, Guido Freckmann
article en

Abstract

The eCGM designation is a recently proposed framework for the approval of continuous glucose monitoring (CGM) systems in Europe. The aim of this analysis was to assess whether the performance of the CareSens Air CGM system (i-SENS, Inc., Republic of Korea), derived from a pooled dataset comprising three clinical investigations, complies with eCGM requirements. A total of 164 participants (86% type 1 diabetes, 50% female, 66% on multiple daily injections) across the three studies contributed 17,655 data points to the data analysis. Meal and/or insulin challenges were implemented to obtain sufficient data points at comparator glucose concentrations <70 mg/dL (10.9%) and >300 mg/dL (5.7%). Measurement accuracy was assessed by applying eCGM accuracy requirements. The CareSens Air CGM system complies with eCGM accuracy requirements in this data analysis. Approximately 91.0% of the CGM readings <70 mg/dL were found within ±15 mg/dL from the paired comparator value. At eu- (70–180 mg/dL) and hyperglycemic (>180 mg/dL) CGM readings, 78.6% and 84.4% of CGM readings exhibited differences ≤ ±15%. Overall, 90.3% of CGM readings were found within ±20% of the paired comparator values across the measuring range. Accuracy remained stable throughout the sensor lifetime. The CareSens Air CGM system is suitable for nonadjunctive insulin dosing decisions and integration with automated insulin delivery systems. The three studies whose data were used in this pooled analysis were registered in the Republic of Korea’s Clinical Research Information Service (KCT0009215) and the German Clinical Trials Register (DRKS00028739 and DRKS00037132).

Diabetes Therapy
Universität Ulm (DE), Kangbuk Samsung Hospital (KR), CHA University Bundang Medical Center (KR), CHA University (KR), The Catholic University of Korea Seoul St. Mary's Hospital (KR), Catholic University of Korea (KR), Dankook University (KR)
Openalex Percentile: Top 10%
Diabetes Management and Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.