Method Comparison of PT, INR, and aPTT Results Obtained Using Two Coagulation Analyzers in Routine Laboratory Practice

Background/Objectives: Prothrombin time (PT), international normalized ratio (INR), and activated partial thromboplastin time (aPTT) are routinely used coagulation assays, and analyzer–reagent differences may affect result comparability. This study evaluated analytical comparability between the Sysmex CS-2500 System using Siemens reagents and the Tokra Medical NOVAE II for PT, INR, and aPTT. Methods: Residual routine citrated plasma specimens were measured on both systems. The CS-2500 was designated as the comparator system. Passing–Bablok regression and Bland–Altman analysis were used as the primary method-comparison approaches; Pearson and Spearman correlations and exploratory categorical agreement were secondary analyses. Results: Fifty paired measurements were analyzed for PT and INR and 54 for aPTT. For all three parameters, the 95% confidence interval (CI) for the Passing–Bablok intercept included 0 and the 95% CI for the slope included 1, providing no statistically supported evidence of constant or proportional bias by regression. Mean paired differences (NOVAE II minus CS-2500) were +1.27 s for PT, +0.023 for INR, and +2.32 s for aPTT. The 95% limits of agreement were −0.15 to +2.69 s for PT, −0.112 to +0.159 for INR, and −1.70 to +6.33 s for aPTT. Categorical agreement was influenced by analyzer-specific reference intervals and by the low prevalence of abnormal results. Conclusions: The two analyzer–reagent systems showed positive analytical associations, but the magnitude and dispersion of paired differences varied by assay. Because no clinical equivalence margins were prespecified and markedly pathological or therapeutic-range samples were sparsely represented, these findings support analytical comparison and local verification but do not establish clinical interchangeability.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2026-09-15
DOI
https://doi.org/10.3390/diagnostics16182989
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
article

Method Comparison of PT, INR, and aPTT Results Obtained Using Two Coagulation Analyzers in Routine Laboratory Practice

Betül Özbek İpteç, Bağnu Dündar, Tuğba Elgün, Asiye Gök Yurttaş et al.
Diagnostics
Clinical Laboratory Practices and Quality Control
article

Method Comparison of PT, INR, and aPTT Results Obtained Using Two Coagulation Analyzers in Routine Laboratory Practice

Betül Özbek İpteç, Bağnu Dündar, Tuğba Elgün, Asiye Gök Yurttaş, Gül İpek Gündoğan, Sevgi Koçyiğit Sevinç
article en

Abstract

Background/Objectives: Prothrombin time (PT), international normalized ratio (INR), and activated partial thromboplastin time (aPTT) are routinely used coagulation assays, and analyzer–reagent differences may affect result comparability. This study evaluated analytical comparability between the Sysmex CS-2500 System using Siemens reagents and the Tokra Medical NOVAE II for PT, INR, and aPTT. Methods: Residual routine citrated plasma specimens were measured on both systems. The CS-2500 was designated as the comparator system. Passing–Bablok regression and Bland–Altman analysis were used as the primary method-comparison approaches; Pearson and Spearman correlations and exploratory categorical agreement were secondary analyses. Results: Fifty paired measurements were analyzed for PT and INR and 54 for aPTT. For all three parameters, the 95% confidence interval (CI) for the Passing–Bablok intercept included 0 and the 95% CI for the slope included 1, providing no statistically supported evidence of constant or proportional bias by regression. Mean paired differences (NOVAE II minus CS-2500) were +1.27 s for PT, +0.023 for INR, and +2.32 s for aPTT. The 95% limits of agreement were −0.15 to +2.69 s for PT, −0.112 to +0.159 for INR, and −1.70 to +6.33 s for aPTT. Categorical agreement was influenced by analyzer-specific reference intervals and by the low prevalence of abnormal results. Conclusions: The two analyzer–reagent systems showed positive analytical associations, but the magnitude and dispersion of paired differences varied by assay. Because no clinical equivalence margins were prespecified and markedly pathological or therapeutic-range samples were sparsely represented, these findings support analytical comparison and local verification but do not establish clinical interchangeability.

DiagnosticsVol. 16(18)
Kütahya Dumlupınar Üniversitesi (TR), Biruni University (TR), Sağlık Bilimleri Üniversitesi (TR), Atlas Üniversitesi, Istanbul Arel University (TR)
Openalex Percentile: Top 11%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.