Ultra high-density, 4096-channel intraoperative neurophysiological brain mapping for functional localization of the human central sulcus

OBJECTIVE: Intraoperative localization of the central sulcus (CS) using somatosensory evoked-potential (SSEP) phase reversal is routinely performed with sparse subdural strip electrodes. We evaluated whether a multi-array, 4096-channel surface micro-electrocorticography (μECoG) configuration could provide dense two-dimensional sampling of stimulus-dependent sensorimotor responses within a standard neurosurgical workflow. Approach. Four Precision Neuroscience Layer 7 μECoG arrays (1024 platinum electrodes per array on a flexible polyimide substrate, 400 μm electrode pitch, ~1.5 cm² active area per array) were deployed during resection of a right parafalcine meningioma in a 56-year-old male. Two arrays were placed horizontally on the exposed precentral gyrus; two were inserted under the intact dura overlying the postcentral gyrus, using a custom flexible stylet. SSEPs were recorded across five contralateral stimulation conditions (median nerve, ulnar nerve, index finger, middle finger, ring finger). Per-electrode SSEP responses were classified without anatomical labels, and spatial organization was quantified across stimulation conditions. Stimulus-locked high-gamma activity, digit-response maps, and signal quality across micro- and macroelectrodes were also evaluated. Main results. Aggregate channel yield was 91.3% (3739/4096) at a 2 MΩ impedance criterion (per-array range 86.7-97.0%). Per-electrode amplitude maps resolved continuous phase-reversal contours with stimulus-specific spatial structure; reversal latencies were 19, 21, 25, 26, and 25 ms for median, ulnar, index, middle, and ring stimulation respectively. Phase reversal occurred within a single array in eight of ten sensory-array recordings, with the phase-reversal pattern varying across stimulation conditions. Digit responses showed measurable spatial differentiation within substantially overlapping response fields. Automated classification of the phase-reversal responses recovered the expected motor and sensory organization, consistent with the standard-of-care intraoperative localization performed in the same case. Significance. The deployment shows that a four-array Layer 7 μECoG configuration can be used within a standard neurosurgical exposure and provides dense two-dimensional sampling of stimulus-dependent phase-reversal patterns across the peri-Rolandic recording field. The measured responses were correlated across approximately 3-4 mm of cortex, an order of magnitude coarser than the 400 μm electrode pitch, so the contribution demonstrated here is dense spatial sampling rather than submillimeter physiological resolution. Dense sampling rendered the polarity transition as a continuous two-dimensional boundary across the recording field rather than as a reversal between two adjacent contacts. The platform supports future evaluation of high-density surface μECoG for intraoperative mapping and chronic brain-computer-interface applications. .

Authors

Institutions

Publication Details

Journal
Journal of Neural Engineering
Published
2026-09-14
DOI
https://doi.org/10.1088/1741-2552/aea70f
Primary Topic
Neuroscience and Neural Engineering
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Ultra high-density, 4096-channel intraoperative neurophysiological brain mapping for functional localization of the human central sulcus

Elton Ho, Benjamin I. Rapoport, Daniel D. Cummins, Gelana Tostaeva et al.
Journal of Neural Engineering
Neuroscience and Neural Engineering
article

Ultra high-density, 4096-channel intraoperative neurophysiological brain mapping for functional localization of the human central sulcus

Elton Ho, Benjamin I. Rapoport, Daniel D. Cummins, Gelana Tostaeva, Alexandra M. Fink, Joshua B. Bederson, Craig H. Mermel, Jacqueline Dister, Katrina Barth, Adam Poole, Lizbeth Nunez, Ignacio Saez
article en

Abstract

OBJECTIVE: Intraoperative localization of the central sulcus (CS) using somatosensory evoked-potential (SSEP) phase reversal is routinely performed with sparse subdural strip electrodes. We evaluated whether a multi-array, 4096-channel surface micro-electrocorticography (μECoG) configuration could provide dense two-dimensional sampling of stimulus-dependent sensorimotor responses within a standard neurosurgical workflow. Approach. Four Precision Neuroscience Layer 7 μECoG arrays (1024 platinum electrodes per array on a flexible polyimide substrate, 400 μm electrode pitch, ~1.5 cm² active area per array) were deployed during resection of a right parafalcine meningioma in a 56-year-old male. Two arrays were placed horizontally on the exposed precentral gyrus; two were inserted under the intact dura overlying the postcentral gyrus, using a custom flexible stylet. SSEPs were recorded across five contralateral stimulation conditions (median nerve, ulnar nerve, index finger, middle finger, ring finger). Per-electrode SSEP responses were classified without anatomical labels, and spatial organization was quantified across stimulation conditions. Stimulus-locked high-gamma activity, digit-response maps, and signal quality across micro- and macroelectrodes were also evaluated. Main results. Aggregate channel yield was 91.3% (3739/4096) at a 2 MΩ impedance criterion (per-array range 86.7-97.0%). Per-electrode amplitude maps resolved continuous phase-reversal contours with stimulus-specific spatial structure; reversal latencies were 19, 21, 25, 26, and 25 ms for median, ulnar, index, middle, and ring stimulation respectively. Phase reversal occurred within a single array in eight of ten sensory-array recordings, with the phase-reversal pattern varying across stimulation conditions. Digit responses showed measurable spatial differentiation within substantially overlapping response fields. Automated classification of the phase-reversal responses recovered the expected motor and sensory organization, consistent with the standard-of-care intraoperative localization performed in the same case. Significance. The deployment shows that a four-array Layer 7 μECoG configuration can be used within a standard neurosurgical exposure and provides dense two-dimensional sampling of stimulus-dependent phase-reversal patterns across the peri-Rolandic recording field. The measured responses were correlated across approximately 3-4 mm of cortex, an order of magnitude coarser than the 400 μm electrode pitch, so the contribution demonstrated here is dense spatial sampling rather than submillimeter physiological resolution. Dense sampling rendered the polarity transition as a continuous two-dimensional boundary across the recording field rather than as a reversal between two adjacent contacts. The platform supports future evaluation of high-density surface μECoG for intraoperative mapping and chronic brain-computer-interface applications. .

Journal of Neural Engineering
Icahn School of Medicine at Mount Sinai (US)
Openalex Percentile: Top 17%
Neuroscience and Neural Engineering
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.