Validity and reliability of the Vyntus CPX metabolic gas analyzer compared with the Douglas Bag method

Abstract Aims To validate the Vyntus CPX system against Douglas Bag method. Methods Twelve cycle-trained men and women completed an incremental cycling protocol starting at 50 W, increasing by 50 W every fifth minute up to 350 W or to near exhaustion. After a 5 min break a test of peak oxygen uptake ( V̇ O 2 peak) was carried out. Gas exchange parameters were continuously measured using the Vyntus CPX system in mixing chamber mode, while expired air was simultaneously collected in Douglas Bags during the final 60 s of each stage. Results Mean systematic differences were observed for V̇ O 2 (+ 0.108 L/min, 4.6%) and respiratory exchange ratio (RER, –0.032, –3.2%). The Vyntus CPX measured proportionally higher V̇ O 2 (+ 0.032 L/min) and lower V̇ CO 2 (–0.032 L/min) per L/min increase in V̇ O 2 and V̇ CO 2 , respectively, measured by the Douglas Bag. Consequently, the Vyntus CPX proportionally underestimated RER by 0.017 per 0.1-unit increase. From paired measurements on two days, coefficient of variation (CV) for the gas exchange variables were 3.4%–5.7%. After adjusting for biological variability, systematic test–retest differences were observed for V̇ CO 2 (+ 0.066 L/min, 2.0%) and RER (+ 0.021, 2.3%). Proportional bias was evident across all gas exchange variables. Conclusion The Vyntus CPX demonstrated agreement comparable with other studies on similar commercial instrument but showed reduced accuracy at higher levels of V̇ O 2 and V̇ CO 2 . The variability in technological error was somewhat larger than what is generally considered acceptable for gas exchange measurements. The Vyntus CPX showed systematic test–retest differences for V̇ CO 2 and RER, with proportional bias evident for all measured gas exchange variables.

Authors

Institutions

Publication Details

Journal
European Journal of Applied Physiology
Published
2026-09-12
DOI
https://doi.org/10.1007/s00421-026-06429-w
Primary Topic
Cardiovascular and exercise physiology
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validity and reliability of the Vyntus CPX metabolic gas analyzer compared with the Douglas Bag method

Svein Arne Pettersen, Kristoffer Robin Johansen, Kim Arne Heitmann, Jon Ingulf Medbø et al.
European Journal of Applied Physiology
Cardiovascular and exercise physiology
article

Validity and reliability of the Vyntus CPX metabolic gas analyzer compared with the Douglas Bag method

Svein Arne Pettersen, Kristoffer Robin Johansen, Kim Arne Heitmann, Jon Ingulf Medbø, Boye Welde, Karianne Hagerupsen
article en

Abstract

Abstract Aims To validate the Vyntus CPX system against Douglas Bag method. Methods Twelve cycle-trained men and women completed an incremental cycling protocol starting at 50 W, increasing by 50 W every fifth minute up to 350 W or to near exhaustion. After a 5 min break a test of peak oxygen uptake ( V̇ O 2 peak) was carried out. Gas exchange parameters were continuously measured using the Vyntus CPX system in mixing chamber mode, while expired air was simultaneously collected in Douglas Bags during the final 60 s of each stage. Results Mean systematic differences were observed for V̇ O 2 (+ 0.108 L/min, 4.6%) and respiratory exchange ratio (RER, –0.032, –3.2%). The Vyntus CPX measured proportionally higher V̇ O 2 (+ 0.032 L/min) and lower V̇ CO 2 (–0.032 L/min) per L/min increase in V̇ O 2 and V̇ CO 2 , respectively, measured by the Douglas Bag. Consequently, the Vyntus CPX proportionally underestimated RER by 0.017 per 0.1-unit increase. From paired measurements on two days, coefficient of variation (CV) for the gas exchange variables were 3.4%–5.7%. After adjusting for biological variability, systematic test–retest differences were observed for V̇ CO 2 (+ 0.066 L/min, 2.0%) and RER (+ 0.021, 2.3%). Proportional bias was evident across all gas exchange variables. Conclusion The Vyntus CPX demonstrated agreement comparable with other studies on similar commercial instrument but showed reduced accuracy at higher levels of V̇ O 2 and V̇ CO 2 . The variability in technological error was somewhat larger than what is generally considered acceptable for gas exchange measurements. The Vyntus CPX showed systematic test–retest differences for V̇ CO 2 and RER, with proportional bias evident for all measured gas exchange variables.

European Journal of Applied Physiology
Vestlandsforsking (NO), UiT The Arctic University of Norway (NO)
Universitetet i Tromsø
Openalex Percentile: Top 5%
Cardiovascular and exercise physiology
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.