Comparison between choroidal thickness measurements from swept-source optical coherence tomography devices in highly myopic Chinese adults

BACKGROUND/AIMS: Reproducibility of choroidal thickness (CT) measurements is imperative when comparing choroidal imaging from different optical coherence tomography (OCT) platforms particularly in high myopia. This study compared CT measurements from commercial swept-source OCT devices in highly myopic Chinese adults using vendor-provided software and a deep learning segmentation model. METHODS: 97 participants underwent imaging with DREAM OCT and Triton OCT. CT was measured across the Early Treatment Diabetic Retinopathy Study (ETDRS) grid using vendor-provided software and a deep learning model, with ocular magnification correction. Agreement was evaluated using mean differences with 95% CIs, intraclass correlation coefficients (ICCs) and Bland-Altman plots. RESULTS: With vendor-provided software, DREAM OCT yielded thicker CT than Triton OCT (global ETDRS: 221.48±66.88 µm vs 156.90±57.64 µm), with mean differences of 61.04-71.22 µm (all p<0.001). Global agreement was moderate (ICC=0.63; 95% CI 0.56 to 0.68). Deep learning-derived CT measurements showed much smaller inter-device differences (DREAM: 173.47±59.54 µm; Triton: 171.21±56.62 µm), yielding a mean global difference of 2.26 µm with limits of agreement from -17.38 to 21.90 µm. Sectoral CT differences were similarly minimised (ICCs>0.90 across all ETDRS regions). The deep learning model demonstrated significantly superior inter-device agreement (global ICC=0.98 vs 0.63; p<0.05). CONCLUSIONS: Vendor-provided software produced systematic differences between DREAM OCT and Triton OCT, whereas the deep learning model produced CT measurements with significantly reduced variability. When evaluating CT measurements from different platforms, deep learning-based approaches may provide more comparable results than vendor-provided software.

Authors

Institutions

Publication Details

Journal
British Journal of Ophthalmology
Published
2026-09-11
DOI
https://doi.org/10.1136/bjo-2026-329740
Primary Topic
Retinal Diseases and Treatments
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison between choroidal thickness measurements from swept-source optical coherence tomography devices in highly myopic Chinese adults

Leopold Schmetterer, Kai Xiong Cheong, Ecosse Luc Lamoureux, Syna Sreng et al.
British Journal of Ophthalmology
Retinal Diseases and Treatments
article

Comparison between choroidal thickness measurements from swept-source optical coherence tomography devices in highly myopic Chinese adults

Leopold Schmetterer, Kai Xiong Cheong, Ecosse Luc Lamoureux, Syna Sreng, Yanfeng Jiang, Seang‐Mei Saw, Weizhong Lan, Zhongping Chen, Xiangzhong Tang, Huanhuan Tan, Yvonne Ng, Quan V. Hoang, Valentina Bellemo, Damon Wong
article en

Abstract

BACKGROUND/AIMS: Reproducibility of choroidal thickness (CT) measurements is imperative when comparing choroidal imaging from different optical coherence tomography (OCT) platforms particularly in high myopia. This study compared CT measurements from commercial swept-source OCT devices in highly myopic Chinese adults using vendor-provided software and a deep learning segmentation model. METHODS: 97 participants underwent imaging with DREAM OCT and Triton OCT. CT was measured across the Early Treatment Diabetic Retinopathy Study (ETDRS) grid using vendor-provided software and a deep learning model, with ocular magnification correction. Agreement was evaluated using mean differences with 95% CIs, intraclass correlation coefficients (ICCs) and Bland-Altman plots. RESULTS: With vendor-provided software, DREAM OCT yielded thicker CT than Triton OCT (global ETDRS: 221.48±66.88 µm vs 156.90±57.64 µm), with mean differences of 61.04-71.22 µm (all p<0.001). Global agreement was moderate (ICC=0.63; 95% CI 0.56 to 0.68). Deep learning-derived CT measurements showed much smaller inter-device differences (DREAM: 173.47±59.54 µm; Triton: 171.21±56.62 µm), yielding a mean global difference of 2.26 µm with limits of agreement from -17.38 to 21.90 µm. Sectoral CT differences were similarly minimised (ICCs>0.90 across all ETDRS regions). The deep learning model demonstrated significantly superior inter-device agreement (global ICC=0.98 vs 0.63; p<0.05). CONCLUSIONS: Vendor-provided software produced systematic differences between DREAM OCT and Triton OCT, whereas the deep learning model produced CT measurements with significantly reduced variability. When evaluating CT measurements from different platforms, deep learning-based approaches may provide more comparable results than vendor-provided software.

British Journal of Ophthalmology
Central South University (CN), National University of Singapore (SG), Nanyang Technological University (SG), Changsha Central Hospital (CN), Singapore Eye Research Institute (SG), Duke-NUS Medical School (SG), National University Health System (SG), He Eye Hospital (CN), Medical University of Vienna (AT), Columbia University (US)
National Research Foundation Singapore, Singapore Eye Research Institute, National Medical Research Council, National Health Innovation Centre Singapore
Openalex Percentile: Top 8%
Retinal Diseases and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.