The Structure–Function Link of Ellipsoid Zone Integrity and Outer Retinal Features with Visual Acuity in Dry AMD

Background/Objectives: The purpose of this study was to evaluate associations between quantitative ellipsoid zone (EZ) integrity metrics derived from spectral-domain OCT (SD-OCT) and visual outcomes, including clinically meaningful visual acuity (VA) loss, in dry age-related macular degeneration (AMD). Methods: This single-center retrospective image-analysis cohort included 351 eyes from 351 patients with intermediate AMD or atrophic changes. Macular SD-OCT scans were analyzed using a certified-reader-validated machine learning-based segmentation platform to delineate the EZ, retinal pigment epithelium (RPE), and Bruch’s membrane (BM). EZ metrics included EZ-RPE thickness (i.e., outer segment thickness) and EZ attenuation, defined as partial attenuation (EZ-RPE thickness of ≤20 µm, abnormal outer segment thinning)) or total attenuation/loss (0 µm). Drusen volume was quantified from the RPE-BM compartment. VA was converted to ETDRS letters. Associations between baseline OCT metrics and baseline VA, Year 2 VA, and ≥15-letter VA loss were assessed using nonparametric testing, linear regression, and ANCOVA. Results: Greater EZ-RPE thickness and lower EZ attenuation were associated with better baseline VA (all p < 0.001), and EZ metrics remained significant in cross-sectional regression models. Among 188 eyes with Year 2 follow-up, 38 (20%) experienced ≥15-letter VA loss. These eyes had significantly lower baseline EZ-RPE thickness (i.e., abnormal outer segment thinning) and greater EZ attenuation than eyes without such loss (all p < 0.001). In ANCOVA models, EZ metrics remained associated with Year 2 VA (all p < 0.001), whereas drusen volume was not. Conclusions: Quantitative EZ integrity metrics were associated with cross-sectional and longitudinal visual outcomes in dry AMD. EZ-RPE thickness and EZ attenuation may serve as clinically relevant OCT biomarkers reflecting photoreceptor integrity and risk of meaningful VA loss.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2026-09-10
DOI
https://doi.org/10.3390/diagnostics16182925
Primary Topic
Retinal Diseases and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
article

The Structure–Function Link of Ellipsoid Zone Integrity and Outer Retinal Features with Visual Acuity in Dry AMD

Ming Hu, Rachel A. Downes, Karen Matar, Asmita Indurkar et al.
Diagnostics
Retinal Diseases and Treatments
article

The Structure–Function Link of Ellipsoid Zone Integrity and Outer Retinal Features with Visual Acuity in Dry AMD

Ming Hu, Rachel A. Downes, Karen Matar, Asmita Indurkar, Masaharu Mizuno, Reem Amine, Justis P. Ehlers, Katherine E. Talcott, Jamie Reese, Paulo Simoes
article en

Abstract

Background/Objectives: The purpose of this study was to evaluate associations between quantitative ellipsoid zone (EZ) integrity metrics derived from spectral-domain OCT (SD-OCT) and visual outcomes, including clinically meaningful visual acuity (VA) loss, in dry age-related macular degeneration (AMD). Methods: This single-center retrospective image-analysis cohort included 351 eyes from 351 patients with intermediate AMD or atrophic changes. Macular SD-OCT scans were analyzed using a certified-reader-validated machine learning-based segmentation platform to delineate the EZ, retinal pigment epithelium (RPE), and Bruch’s membrane (BM). EZ metrics included EZ-RPE thickness (i.e., outer segment thickness) and EZ attenuation, defined as partial attenuation (EZ-RPE thickness of ≤20 µm, abnormal outer segment thinning)) or total attenuation/loss (0 µm). Drusen volume was quantified from the RPE-BM compartment. VA was converted to ETDRS letters. Associations between baseline OCT metrics and baseline VA, Year 2 VA, and ≥15-letter VA loss were assessed using nonparametric testing, linear regression, and ANCOVA. Results: Greater EZ-RPE thickness and lower EZ attenuation were associated with better baseline VA (all p < 0.001), and EZ metrics remained significant in cross-sectional regression models. Among 188 eyes with Year 2 follow-up, 38 (20%) experienced ≥15-letter VA loss. These eyes had significantly lower baseline EZ-RPE thickness (i.e., abnormal outer segment thinning) and greater EZ attenuation than eyes without such loss (all p < 0.001). In ANCOVA models, EZ metrics remained associated with Year 2 VA (all p < 0.001), whereas drusen volume was not. Conclusions: Quantitative EZ integrity metrics were associated with cross-sectional and longitudinal visual outcomes in dry AMD. EZ-RPE thickness and EZ attenuation may serve as clinically relevant OCT biomarkers reflecting photoreceptor integrity and risk of meaningful VA loss.

DiagnosticsVol. 16(18)
Cleveland Clinic (US), Advanced Imaging Research (United States) (US), Cleveland Eye Clinic (US), Center for Excellence in Education (US)
Quality Education
Openalex Percentile: Top 8%
Retinal Diseases and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.