Pre-analytical and analytical validation of calcitonin measurement in fine needle aspiration cytology washout

OBJECTIVES: Current guidelines advise calcitonin measurements in washout from fine needle aspiration cytology (FNAC) upon inconclusive cytological examination of medullary thyroid cancer (MTC). (Pre-)analytical validation studies of calcitonin measurements in FNAC washout are however missing. In this study, we examined (pre-)analytical aspects of calcitonin measurements using three washout solvents, aiming to provide insights for implementation of calcitonin measurements in FNAC washout as a supplementary test for diagnosing and monitoring patients with MTC. METHODS: Validation studies were performed using CytoLyt, NaCl and MultiDiluent as washout solvents. Precision (intra- and inter-assay coefficients of variation (CV)), accuracy (linearity and recovery) and storage stability of each solvent were assessed after spiking with calcitonin standard. Calcitonin measurements were performed using the serum CALCT immunoassay (Siemens, Atellica). Decision criteria were set based on CLSI and manufacturer's manual. RESULTS: =0.997) and median recovery was 148 % with limited variation. Furthermore, calcitonin concentrations remained stable (<10 % change) after storage. Measurements in CytoLyt were precise and accurate, but unstable upon storage. Measurements in NaCl only met the criteria for intra-assay CV and linearity. CONCLUSIONS: This study showed that MultiDiluent was the optimal solvent to use for calcitonin measurements in a non-serum matrix using the Atellica calcitonin immunoassay. CytoLyt performed suboptimal, while the performance of NaCl was poor. Further research is needed to validate the recommended work-up using patient-derived material.

Authors

Institutions

Publication Details

Journal
Clinical Chemistry and Laboratory Medicine (CCLM)
Published
2026-09-10
DOI
https://doi.org/10.1515/cclm-2026-1057
Primary Topic
Thyroid Cancer Diagnosis and Treatment
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Pre-analytical and analytical validation of calcitonin measurement in fine needle aspiration cytology washout

Anita Boelen, Pim de Graaf, Amber Mater, Annemieke C. Heijboer et al.
Clinical Chemistry and Laboratory Medicine (CCLM)
Thyroid Cancer Diagnosis and Treatment
article

Pre-analytical and analytical validation of calcitonin measurement in fine needle aspiration cytology washout

Anita Boelen, Pim de Graaf, Amber Mater, Annemieke C. Heijboer, Jacquelien J. Hillebrand, Koen M.A. Dreijerink
article en

Abstract

OBJECTIVES: Current guidelines advise calcitonin measurements in washout from fine needle aspiration cytology (FNAC) upon inconclusive cytological examination of medullary thyroid cancer (MTC). (Pre-)analytical validation studies of calcitonin measurements in FNAC washout are however missing. In this study, we examined (pre-)analytical aspects of calcitonin measurements using three washout solvents, aiming to provide insights for implementation of calcitonin measurements in FNAC washout as a supplementary test for diagnosing and monitoring patients with MTC. METHODS: Validation studies were performed using CytoLyt, NaCl and MultiDiluent as washout solvents. Precision (intra- and inter-assay coefficients of variation (CV)), accuracy (linearity and recovery) and storage stability of each solvent were assessed after spiking with calcitonin standard. Calcitonin measurements were performed using the serum CALCT immunoassay (Siemens, Atellica). Decision criteria were set based on CLSI and manufacturer's manual. RESULTS: =0.997) and median recovery was 148 % with limited variation. Furthermore, calcitonin concentrations remained stable (<10 % change) after storage. Measurements in CytoLyt were precise and accurate, but unstable upon storage. Measurements in NaCl only met the criteria for intra-assay CV and linearity. CONCLUSIONS: This study showed that MultiDiluent was the optimal solvent to use for calcitonin measurements in a non-serum matrix using the Atellica calcitonin immunoassay. CytoLyt performed suboptimal, while the performance of NaCl was poor. Further research is needed to validate the recommended work-up using patient-derived material.

Clinical Chemistry and Laboratory Medicine (CCLM)
Amsterdam Neuroscience (NL), Amsterdam University Medical Centers (NL), Vrije Universiteit Amsterdam (NL), University of Amsterdam (NL)
Openalex Percentile: Top 10%
Thyroid Cancer Diagnosis and Treatment
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.