Real-life evaluation of BioFire® BCID2 and Vitek® REVEALTM for Gram-negative bloodstream infections in critical care

Timely management of sepsis relies on rapid diagnosis of bloodstream infections (BSIs), including early pathogen identification and antimicrobial susceptibility testing (AST). The Vitek ® REVEAL TM is an innovative rapid AST platform that detects volatile organic compounds emitted by Gram-negative bacilli (GNB) directly from positive blood cultures. This study assessed the analytical performance and clinical contribution of the BioFire ® BCID2 molecular panel and the Vitek ® REVEAL TM system in the diagnosis of GNB-BSI among intensive care unit (ICU) patients. We analyzed the prospective phase of the BacteREVEAL study, conducted between March and October 2023 at Nîmes University Hospital. Results from BioFire ® BCID2 and Vitek ® REVEAL TM were compared to standard-of-care methods, including MALDI-TOF identification and direct disk diffusion AST. Performance metrics for Vitek ® REVEAL TM were evaluated according to ISO 20776–2:2007 standards. Among 94 ICU patients with a first episode of GNB-BSI, 50 monomicrobial cases were included. Escherichia coli was the most frequently identified species (n = 30, 60%), followed by Pseudomonas aeruginosa (n = 7, 14%). BioFire ® BCID2 provided accurate identification for all isolates. The innovative tools reduced median turnaround times for identification by 19.0 h and for AST by 14.9 h. The mean time to result for Vitek ® REVEAL TM was 5.7 h. Overall categorical agreement with standard AST was 97.7%, with 8 very major errors, 8 major errors, and 2 minor errors. These findings support the integration of BioFire ® BCID2 and Vitek ® REVEAL TM into routine workflows to accelerate GNB-BSI diagnosis and improve sepsis management.

Authors

Institutions

Publication Details

Journal
PLoS ONE
Published
2026-09-10
DOI
https://doi.org/10.1371/journal.pone.0356199
Primary Topic
Bacterial Identification and Susceptibility Testing
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
article

Real-life evaluation of BioFire® BCID2 and Vitek® REVEALTM for Gram-negative bloodstream infections in critical care

Alix Pantel, Claire Roger, Jean‐Philippe Lavigne, Adeline Boutet-Dubois et al.
PLoS ONE
Bacterial Identification and Susceptibility Testing
article

Real-life evaluation of BioFire® BCID2 and Vitek® REVEALTM for Gram-negative bloodstream infections in critical care

Alix Pantel, Claire Roger, Jean‐Philippe Lavigne, Adeline Boutet-Dubois, Florian Salipante, Massinissa Benyahia, Louis Oudiane, Léo Brochon, Rim Nsiri, Matthieu Curtil Dit Galin
article en

Abstract

Timely management of sepsis relies on rapid diagnosis of bloodstream infections (BSIs), including early pathogen identification and antimicrobial susceptibility testing (AST). The Vitek ® REVEAL TM is an innovative rapid AST platform that detects volatile organic compounds emitted by Gram-negative bacilli (GNB) directly from positive blood cultures. This study assessed the analytical performance and clinical contribution of the BioFire ® BCID2 molecular panel and the Vitek ® REVEAL TM system in the diagnosis of GNB-BSI among intensive care unit (ICU) patients. We analyzed the prospective phase of the BacteREVEAL study, conducted between March and October 2023 at Nîmes University Hospital. Results from BioFire ® BCID2 and Vitek ® REVEAL TM were compared to standard-of-care methods, including MALDI-TOF identification and direct disk diffusion AST. Performance metrics for Vitek ® REVEAL TM were evaluated according to ISO 20776–2:2007 standards. Among 94 ICU patients with a first episode of GNB-BSI, 50 monomicrobial cases were included. Escherichia coli was the most frequently identified species (n = 30, 60%), followed by Pseudomonas aeruginosa (n = 7, 14%). BioFire ® BCID2 provided accurate identification for all isolates. The innovative tools reduced median turnaround times for identification by 19.0 h and for AST by 14.9 h. The mean time to result for Vitek ® REVEAL TM was 5.7 h. Overall categorical agreement with standard AST was 97.7%, with 8 very major errors, 8 major errors, and 2 minor errors. These findings support the integration of BioFire ® BCID2 and Vitek ® REVEAL TM into routine workflows to accelerate GNB-BSI diagnosis and improve sepsis management.

PLoS ONEVol. 21(9)
Inserm (FR), Université de Montpellier (FR), Université de Nîmes (FR)
Responsible consumption and production
Openalex Percentile: Top 14%
Bacterial Identification and Susceptibility Testing
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines