Respiratory rate estimation by using realignment data from a 1-minute high-temporal-resolution MRI sequence: A feasibility study

Background Breath-hold triggered functional magnetic resonance imaging (bh-fMRI) is a widely available, easily implementable and inexpensive method to quantify cerebrovascular reactivity. Estimation of patients’ respiratory rate is relevant, as it might influence bh-fMRI data quality, but respiratory sensors integrated in the scanner table are not always available. Purpose To propose an approach to estimate patients’ respiratory rate using realignment data of a 1-min high-temporal-resolution cerebral MRI sequence. Material and Methods In a prospective study, 15 patients with Moyamoya angiopathy underwent a 1-min MRI sequence during routine clinical MRI. The respiratory rate was obtained by analyzing patients’ head motion from image realignment data. To obtain robust results, the respiratory rate was calculated using a multi-method approach, incorporating power spectral density and fast Fourier transform and weighted data from multiparametric rigid body transformation. The respiratory rate detected from respiratory sensors integrated in the scanner table was used as reference. Bland–Altman analysis was performed and intra-class correlation coefficients (ICCs) were calculated to compare the respiratory rates. Results Bland–Altman analysis revealed a mean bias of 0.5 breaths/min and limits of agreement (LoA) ranging from −5.3 to 4.4 breaths/min, with no outliers. The overall agreement between the respiratory rate calculated from realignment data and from scanner sensor data was excellent (ICC = 0.866, 95% confidence interval = 0.607–0.955; P < 0.001). Conclusion The calculated mean bias and LoA appear to be clinically acceptable. Estimating respiratory rate from a 1-min high-temporal-resolution MRI sequence may be a feasible method when respiratory sensors integrated in the scanner are not available.

Authors

Publication Details

Journal
Acta Radiologica
Published
2026-09-10
DOI
https://doi.org/10.1177/02841851261484876
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Respiratory rate estimation by using realignment data from a 1-minute high-temporal-resolution MRI sequence: A feasibility study

Leonie Zerweck, Uwe Klose, Till-Karsten Hauser, Ulrike Ernemann
Acta Radiologica
Advanced MRI Techniques and Applications
article

Respiratory rate estimation by using realignment data from a 1-minute high-temporal-resolution MRI sequence: A feasibility study

Leonie Zerweck, Uwe Klose, Till-Karsten Hauser, Ulrike Ernemann
article en

Abstract

Background Breath-hold triggered functional magnetic resonance imaging (bh-fMRI) is a widely available, easily implementable and inexpensive method to quantify cerebrovascular reactivity. Estimation of patients’ respiratory rate is relevant, as it might influence bh-fMRI data quality, but respiratory sensors integrated in the scanner table are not always available. Purpose To propose an approach to estimate patients’ respiratory rate using realignment data of a 1-min high-temporal-resolution cerebral MRI sequence. Material and Methods In a prospective study, 15 patients with Moyamoya angiopathy underwent a 1-min MRI sequence during routine clinical MRI. The respiratory rate was obtained by analyzing patients’ head motion from image realignment data. To obtain robust results, the respiratory rate was calculated using a multi-method approach, incorporating power spectral density and fast Fourier transform and weighted data from multiparametric rigid body transformation. The respiratory rate detected from respiratory sensors integrated in the scanner table was used as reference. Bland–Altman analysis was performed and intra-class correlation coefficients (ICCs) were calculated to compare the respiratory rates. Results Bland–Altman analysis revealed a mean bias of 0.5 breaths/min and limits of agreement (LoA) ranging from −5.3 to 4.4 breaths/min, with no outliers. The overall agreement between the respiratory rate calculated from realignment data and from scanner sensor data was excellent (ICC = 0.866, 95% confidence interval = 0.607–0.955; P < 0.001). Conclusion The calculated mean bias and LoA appear to be clinically acceptable. Estimating respiratory rate from a 1-min high-temporal-resolution MRI sequence may be a feasible method when respiratory sensors integrated in the scanner are not available.

Acta Radiologica
Openalex Percentile: Top 11%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.