Development and preliminary analytical characterisation of an in-house indirect ELISA for relative measurement of IgG binding to SARS-CoV-2 spike RBD (Omicron BA.1)

We developed an in-house indirect ELISA for relative measurement of IgG binding to the SARS-CoV-2 Spike receptor-binding domain (RBD; Omicron BA.1). Four antigen coating concentrations and three serum dilutions were evaluated using an internal convalescent-serum calibrator assigned arbitrary units (AU/mL). A coating concentration of 2 µg/mL and a serum dilution of 1:100 were selected. The calibration means showed a linear relationship (R² > 0.9999), although serum dilution responses did not demonstrate parallelism with the calibrator. Within-run OD-based coefficients of variation were 4.1–6.5%, and between-run AU-based coefficients of variation were 2.4–8.8% across three days with one operator. Background-derived analytical limits and an operational limit of quantification were characterised as preliminary estimates. In a pilot panel of 10 sera from patients with RT-PCR-confirmed infection and 10 pre-pandemic sera, the cut-off (0.134 OD₄₀₅; mean + 3 SD of the pre-pandemic sera) gave a sensitivity of 90.0% (95% CI 59.​6–98.2%). The same negative sera were used to set the cut-off, thus the specificity was estimated by leave-one-out analysis and was 90.0% (59.6–98.2%), with an accuracy of 90.0% (69.9–97.2%). Apparent specificity at the fixed cut-off was 100%. The ROC AUC was 0.99 (95% CI 0.94–1.00). These findings support further development of a research serology platform, with independent clinical validation and calibration harmonisation needed before broader quantitative or diagnostic use.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2026-11-18
DOI
https://doi.org/10.1038/s41598-026-80943-x
Primary Topic
SARS-CoV-2 detection and testing
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Development and preliminary analytical characterisation of an in-house indirect ELISA for relative measurement of IgG binding to SARS-CoV-2 spike RBD (Omicron BA.1)

Naoual Oukkache, Wafa Khaali, Najat Bouchmaa, Rachid El Fatimy et al.
2 links
Scientific Reports
SARS-CoV-2 detection and testing
article

Development and preliminary analytical characterisation of an in-house indirect ELISA for relative measurement of IgG binding to SARS-CoV-2 spike RBD (Omicron BA.1)

Naoual Oukkache, Wafa Khaali, Najat Bouchmaa, Rachid El Fatimy, Reda Ben Mrid, Hicham Hboub, Salma Chakir, Reda Chahir, Noureddine Hammani, Meriem Baghtoul
article en

Abstract

We developed an in-house indirect ELISA for relative measurement of IgG binding to the SARS-CoV-2 Spike receptor-binding domain (RBD; Omicron BA.1). Four antigen coating concentrations and three serum dilutions were evaluated using an internal convalescent-serum calibrator assigned arbitrary units (AU/mL). A coating concentration of 2 µg/mL and a serum dilution of 1:100 were selected. The calibration means showed a linear relationship (R² > 0.9999), although serum dilution responses did not demonstrate parallelism with the calibrator. Within-run OD-based coefficients of variation were 4.1–6.5%, and between-run AU-based coefficients of variation were 2.4–8.8% across three days with one operator. Background-derived analytical limits and an operational limit of quantification were characterised as preliminary estimates. In a pilot panel of 10 sera from patients with RT-PCR-confirmed infection and 10 pre-pandemic sera, the cut-off (0.134 OD₄₀₅; mean + 3 SD of the pre-pandemic sera) gave a sensitivity of 90.0% (95% CI 59.​6–98.2%). The same negative sera were used to set the cut-off, thus the specificity was estimated by leave-one-out analysis and was 90.0% (59.6–98.2%), with an accuracy of 90.0% (69.9–97.2%). Apparent specificity at the fixed cut-off was 100%. The ROC AUC was 0.99 (95% CI 0.94–1.00). These findings support further development of a research serology platform, with independent clinical validation and calibration harmonisation needed before broader quantitative or diagnostic use.

Scientific Reports
Stab Vida (Portugal) (PT), Université Mohammed VI Polytechnique (MA), Université Hassan 1er (MA), Institut Pasteur du Maroc (MA), Madan Parque (PT), University of Hassan II Casablanca (MA)
Université Mohammed VI Polytechnique, Fondation OCP
Good health and well-being
SARS-CoV-2 detection and testing
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon