Spider-Inspired S-Loop Amplification Repurposes an Electron Cage for Photoelectrochemical–Fluorescence Detection of Perfluorooctanoic Acid

Abstract Per- and polyfluoroalkyl substances require analytical methods that combine sensitivity with internal result checking. Here, we report an S-Loop CRISPR dual-mode assay (SCDA) for perfluorooctanoic acid (PFOA). PFOA-induced aptamer displacement initiates an AMPLON-derived isothermal S-Loop reaction mediated by an eight-arm poly(ethylene glycol) spider-inspired primer (SP), and the resulting amplicons activate Cas12a trans-cleavage. A FAM–quencher reporter provides the fluorescence (FL) output, whereas cleavage-induced removal of CdS from a CdS/gold nanoparticle/ZnIn2S4 photoelectrode releases an electron cage produced by type-I carrier confinement and visible-light shielding of the underlying photoactive layer, thereby restoring the photoelectrochemical (PEC) signal. At target concentrations of 10–13, 10–10, and 10–7 M, Cas12a reporter cleavage was linear with time over the first 10 min, and PEG20000-SP produced the largest amplification response and highest apparent reporter-signal rising rate. PEC and FL calibrations were linear over 10–15–10–7 and 10–14–10–7 M, respectively, with equations I = 12.3328 + 0.7207 log10C (R2 = 0.9929) and F = 13631.32 + 845.64 log10C (R2 = 0.9926). Estimated blank-based limits of detection were 1.27 × 10–16 M (PEC) and 8.34 × 10–16 M (FL); these estimates are below the lowest tested standards and do not establish quantitation below the calibration ranges. Paired results were compared within the common calibration range using log-concentration differences and relative percent differences. Consensus recoveries were 94.3–103.6% in milk and 95.7–103.0% in river water. SCDA therefore combines sensitive dual-readout with an explicit preliminary screen for discordant measurements.

Authors

Institutions

Publication Details

Journal
Analytical Chemistry
Published
2026-09-10
DOI
https://doi.org/10.1021/acs.analchem.6c05089
Primary Topic
Per- and polyfluoroalkyl substances research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
article

Spider-Inspired S-Loop Amplification Repurposes an Electron Cage for Photoelectrochemical–Fluorescence Detection of Perfluorooctanoic Acid

Yiwei Zheng, Ke‐Jing Huang, Xinyue Zhang, Na Feng et al.
Analytical Chemistry
Per- and polyfluoroalkyl substances research
article

Spider-Inspired S-Loop Amplification Repurposes an Electron Cage for Photoelectrochemical–Fluorescence Detection of Perfluorooctanoic Acid

Yiwei Zheng, Ke‐Jing Huang, Xinyue Zhang, Na Feng, Xuecai Tan, Ruxia Wei, Wenjie Lu, Jiawen Wu, Xu Li, Jingli Huang
article en

Abstract

Abstract Per- and polyfluoroalkyl substances require analytical methods that combine sensitivity with internal result checking. Here, we report an S-Loop CRISPR dual-mode assay (SCDA) for perfluorooctanoic acid (PFOA). PFOA-induced aptamer displacement initiates an AMPLON-derived isothermal S-Loop reaction mediated by an eight-arm poly(ethylene glycol) spider-inspired primer (SP), and the resulting amplicons activate Cas12a trans-cleavage. A FAM–quencher reporter provides the fluorescence (FL) output, whereas cleavage-induced removal of CdS from a CdS/gold nanoparticle/ZnIn2S4 photoelectrode releases an electron cage produced by type-I carrier confinement and visible-light shielding of the underlying photoactive layer, thereby restoring the photoelectrochemical (PEC) signal. At target concentrations of 10–13, 10–10, and 10–7 M, Cas12a reporter cleavage was linear with time over the first 10 min, and PEG20000-SP produced the largest amplification response and highest apparent reporter-signal rising rate. PEC and FL calibrations were linear over 10–15–10–7 and 10–14–10–7 M, respectively, with equations I = 12.3328 + 0.7207 log10C (R2 = 0.9929) and F = 13631.32 + 845.64 log10C (R2 = 0.9926). Estimated blank-based limits of detection were 1.27 × 10–16 M (PEC) and 8.34 × 10–16 M (FL); these estimates are below the lowest tested standards and do not establish quantitation below the calibration ranges. Paired results were compared within the common calibration range using log-concentration differences and relative percent differences. Consensus recoveries were 94.3–103.6% in milk and 95.7–103.0% in river water. SCDA therefore combines sensitive dual-readout with an explicit preliminary screen for discordant measurements.

Analytical Chemistry
Minzu University of China (CN), Guizhou Minzu University (CN)
Clean water and sanitation
Openalex Percentile: Top 18%
Per- and polyfluoroalkyl substances research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.