Effects of spiral deflation and warming on cuff diameter, insertion force, extubation, and sealing performance in polyvinyl chloride-cuffed tracheal tubes

Cuffed tracheal tubes are widely used in paediatric airway management, but passage through the relatively narrow subglottic airway may increase insertion resistance and risk mucosal trauma. This study evaluated whether spiral deflation (i.e. folding the cuff inward like an umbrella) and warming the tube could reduce cuff diameter and insertion force in a simulated trachea. Four commercially available polyvinyl chloride-cuffed tubes with an internal diameter of 5.0 mm (PORTEX ® (Smiths Medical International Ltd, Ashford, UK), Sheridan ® (Teleflex Inc., Morrisville, NC, USA), Shiley™ (Medtronic, Minneapolis, MN, USA), and VentiSeal™ (Flexicare Medical Ltd, Mountain Ash, UK)) were examined. Spiral deflation significantly reduced cuff diameter, with reductions of (mean difference, MD) −3.6 mm (95% confidence intervals, CI −4.5 to −2.7, P <0.001), −2.8 mm (−4.0 to −1.6, P <0.001), −3.5 mm (−5.0 to −2.1, P =0.001), and −4.9 mm (−6.0 to −3.9, P <0.001), respectively. Spiral deflation also decreased insertion force, with reductions of (MD (CI)) −1.01 N (−1.52 to −0.50, P =0.008), −0.34 N (−0.63 to −0.05, P =0.115), −0.49 N (−0.61 to −0.37, P <0.001), and −2.33 N (−3.83 to −0.84, P =0.028), respectively. Additional warming at 40°C further lowered insertion force, with reductions of (MD (CI)) −0.12 N (−0.19 to −0.05, P =0.020), −0.06 N (−0.08 to −0.03, P =0.002), −0.07 N (−0.11 to −0.03, P =0.010), and −0.48 N (−0.86 to −0.11, P =0.068), respectively. Spiral deflation with warming did not affect extubation force or sealing performance. Overall, spiral deflation effectively reduced cuff diameter by approximately 25% and lowered insertion force, while warming further enhanced ease of insertion. These approaches did not compromise sealing or extubation performance.

Authors

Institutions

Publication Details

Journal
Anaesthesia and Intensive Care
Published
2026-09-09
DOI
https://doi.org/10.1177/0310057x261478645
Primary Topic
Airway Management and Intubation Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effects of spiral deflation and warming on cuff diameter, insertion force, extubation, and sealing performance in polyvinyl chloride-cuffed tracheal tubes

Michiko Kinoshita, Hiroki Yonezawa, Katsuya Tanaka, Chiaki Murakami et al.
Anaesthesia and Intensive Care
Airway Management and Intubation Techniques
article

Effects of spiral deflation and warming on cuff diameter, insertion force, extubation, and sealing performance in polyvinyl chloride-cuffed tracheal tubes

Michiko Kinoshita, Hiroki Yonezawa, Katsuya Tanaka, Chiaki Murakami, Takako Ikezaki
article en

Abstract

Cuffed tracheal tubes are widely used in paediatric airway management, but passage through the relatively narrow subglottic airway may increase insertion resistance and risk mucosal trauma. This study evaluated whether spiral deflation (i.e. folding the cuff inward like an umbrella) and warming the tube could reduce cuff diameter and insertion force in a simulated trachea. Four commercially available polyvinyl chloride-cuffed tubes with an internal diameter of 5.0 mm (PORTEX ® (Smiths Medical International Ltd, Ashford, UK), Sheridan ® (Teleflex Inc., Morrisville, NC, USA), Shiley™ (Medtronic, Minneapolis, MN, USA), and VentiSeal™ (Flexicare Medical Ltd, Mountain Ash, UK)) were examined. Spiral deflation significantly reduced cuff diameter, with reductions of (mean difference, MD) −3.6 mm (95% confidence intervals, CI −4.5 to −2.7, P <0.001), −2.8 mm (−4.0 to −1.6, P <0.001), −3.5 mm (−5.0 to −2.1, P =0.001), and −4.9 mm (−6.0 to −3.9, P <0.001), respectively. Spiral deflation also decreased insertion force, with reductions of (MD (CI)) −1.01 N (−1.52 to −0.50, P =0.008), −0.34 N (−0.63 to −0.05, P =0.115), −0.49 N (−0.61 to −0.37, P <0.001), and −2.33 N (−3.83 to −0.84, P =0.028), respectively. Additional warming at 40°C further lowered insertion force, with reductions of (MD (CI)) −0.12 N (−0.19 to −0.05, P =0.020), −0.06 N (−0.08 to −0.03, P =0.002), −0.07 N (−0.11 to −0.03, P =0.010), and −0.48 N (−0.86 to −0.11, P =0.068), respectively. Spiral deflation with warming did not affect extubation force or sealing performance. Overall, spiral deflation effectively reduced cuff diameter by approximately 25% and lowered insertion force, while warming further enhanced ease of insertion. These approaches did not compromise sealing or extubation performance.

Anaesthesia and Intensive Care
Tokushima Prefectural Central Hospital (JP), Tokushima University Hospital (JP), Tokushima University (JP)
Openalex Percentile: Top 8%
Airway Management and Intubation Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.