Cardiac and respiratory signal separation in electrical impedance tomography (EIT)

Electrical Impedance Tomography (EIT) is a promising non-invasive imaging modality for respiratory and cardiac monitoring. However, separating cardiac signals from the dominant respiratory component remains challenging because both physiological processes are simultaneously embedded in the measured impedance signals. This study investigates the influence of EIT measurement patterns on cardiac-respiratory signal separation using four methods: Continuous Wavelet Transform (CWT), Empirical Mode Decomposition (EMD), Independent Component Analysis (ICA), and Principal Component Analysis (PCA). Numerical simulations were performed in EIDORS using a realistic thoracic geometry reconstructed from the POPI dataset with a GREIT-based image reconstruction framework. Four measurement configurations were evaluated: adjacent-adjacent, adjacent-Skip 4, Skip 4-adjacent, and Skip 4-Skip 4. Signal separation performance was assessed using the Cross Correlation (CC) between the extracted and reference signals, while the normalized Root Mean Square Error (RMSE) was additionally used to evaluate the reconstructed cardiac waveform. Among all evaluated methods, PCA consistently achieved the highest waveform similarity, reaching a Lung CC of 98.12\\% and a Heart CC of 99.56\\% under the Skip 4-Skip 4 measurement pattern. Visual inspection of the reconstructed images further demonstrated clearer separation of respiratory and cardiac conductivity distributions using this configuration. These results indicate that combining PCA with a wide electrode measurement pattern improves cardiac-respiratory signal separation in simulated EIT measurements.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2026-09-08
DOI
https://doi.org/10.1088/2057-1976/aea425
Primary Topic
Electrical and Bioimpedance Tomography
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
article

Cardiac and respiratory signal separation in electrical impedance tomography (EIT)

Suharyana Suharyana, Marlin Ramadhan Baidillah, Maulina Nurul Nurul Hidayah
Biomedical Physics & Engineering Express
Electrical and Bioimpedance Tomography
article

Cardiac and respiratory signal separation in electrical impedance tomography (EIT)

Suharyana Suharyana, Marlin Ramadhan Baidillah, Maulina Nurul Nurul Hidayah
article en

Abstract

Electrical Impedance Tomography (EIT) is a promising non-invasive imaging modality for respiratory and cardiac monitoring. However, separating cardiac signals from the dominant respiratory component remains challenging because both physiological processes are simultaneously embedded in the measured impedance signals. This study investigates the influence of EIT measurement patterns on cardiac-respiratory signal separation using four methods: Continuous Wavelet Transform (CWT), Empirical Mode Decomposition (EMD), Independent Component Analysis (ICA), and Principal Component Analysis (PCA). Numerical simulations were performed in EIDORS using a realistic thoracic geometry reconstructed from the POPI dataset with a GREIT-based image reconstruction framework. Four measurement configurations were evaluated: adjacent-adjacent, adjacent-Skip 4, Skip 4-adjacent, and Skip 4-Skip 4. Signal separation performance was assessed using the Cross Correlation (CC) between the extracted and reference signals, while the normalized Root Mean Square Error (RMSE) was additionally used to evaluate the reconstructed cardiac waveform. Among all evaluated methods, PCA consistently achieved the highest waveform similarity, reaching a Lung CC of 98.12\% and a Heart CC of 99.56\% under the Skip 4-Skip 4 measurement pattern. Visual inspection of the reconstructed images further demonstrated clearer separation of respiratory and cardiac conductivity distributions using this configuration. These results indicate that combining PCA with a wide electrode measurement pattern improves cardiac-respiratory signal separation in simulated EIT measurements.

Biomedical Physics & Engineering Express
Sebelas Maret University (ID), National Nuclear Energy Agency of Indonesia (ID), National Research and Innovation Agency (ID)
Good health and well-being
Openalex Percentile: Top 20%
Electrical and Bioimpedance Tomography
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.