Novel High‐Quality Fluoroscopic Image Processing System With More Than 50% Radiation Dose Reduction in Endoscopic Retrograde Cholangiopancreatography (With Video)

ABSTRACT Objectives Radiation dose reduction is an important consideration in ERCP. However, the low radiation dose compromises image quality. This study evaluated the utility of radiation‐reduction settings in a fluoroscopic system. Methods Three fluoroscopic settings (high‐dose, low‐dose, and low‐dose with a new image processor; SCORE PRO Advance [ SPA ]) were compared. Results From 2024 to 2025, high‐dose (124‐patients), low‐dose (110‐patients), and SPA (135‐patients) were evaluated. There were no significant differences in baseline characteristics of patients and the details of the ERCP procedure, except for sex distribution. High‐dose (skin dose [SD]:72.6 [mGy]/dose area product [DAP]:233.9 [dGy・cm 2 ]) showed a significantly higher radiation dose than low‐dose (40.0/121.9) and SPA (32.5/109.4) ( p < 0.01). For ERCP, procedure time was significantly longer in low‐dose than in SPA (22 vs. 17 min; p = 0.04). In multivariate analysis, high‐dose (SD: odds ratio [OR] 31.6, p < 0.01/DAP:OR17.7, p < 0.01) and long ERCP procedure time (OR3.33, p < 0.01/OR3.90, p < 0.01) were related to the high radiation dose exposure. SPA was independently associated with shorter ERCP procedure time (OR 0.63; p = 0.04). Conclusions SPA and low‐dose provided comparable reductions in radiation exposure. Furthermore, SPA was associated with a shorter ERCP time than a low dose, possibly because of improved image quality. However, until fluoroscopic systems are upgraded to incorporate SPA, low‐dose may provide clinical utility for reducing radiation exposure. Trial Registration ClinicalTrials.gov identifier: UMIN000061448

Authors

Institutions

Publication Details

Journal
Journal of Hepato-Biliary-Pancreatic Sciences
Published
2026-09-06
DOI
https://doi.org/10.1002/jhbp.70183
Primary Topic
Gallbladder and Bile Duct Disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Novel High‐Quality Fluoroscopic Image Processing System With More Than 50% Radiation Dose Reduction in Endoscopic Retrograde Cholangiopancreatography (With Video)

Hiroki Jinno, Masahito Shimizu, Kensaku Yoshida, Mitsuru Okuno et al.
Journal of Hepato-Biliary-Pancreatic Sciences
Gallbladder and Bile Duct Disorders
article

Novel High‐Quality Fluoroscopic Image Processing System With More Than 50% Radiation Dose Reduction in Endoscopic Retrograde Cholangiopancreatography (With Video)

Hiroki Jinno, Masahito Shimizu, Kensaku Yoshida, Mitsuru Okuno, Hisataka Moriwaki, Ryotaro Nagao, Tsuyoshi Mukai, Fumiya Kataoka, Mizuki Ohashi, Naoki Mita
article en

Abstract

ABSTRACT Objectives Radiation dose reduction is an important consideration in ERCP. However, the low radiation dose compromises image quality. This study evaluated the utility of radiation‐reduction settings in a fluoroscopic system. Methods Three fluoroscopic settings (high‐dose, low‐dose, and low‐dose with a new image processor; SCORE PRO Advance [ SPA ]) were compared. Results From 2024 to 2025, high‐dose (124‐patients), low‐dose (110‐patients), and SPA (135‐patients) were evaluated. There were no significant differences in baseline characteristics of patients and the details of the ERCP procedure, except for sex distribution. High‐dose (skin dose [SD]:72.6 [mGy]/dose area product [DAP]:233.9 [dGy・cm 2 ]) showed a significantly higher radiation dose than low‐dose (40.0/121.9) and SPA (32.5/109.4) ( p < 0.01). For ERCP, procedure time was significantly longer in low‐dose than in SPA (22 vs. 17 min; p = 0.04). In multivariate analysis, high‐dose (SD: odds ratio [OR] 31.6, p < 0.01/DAP:OR17.7, p < 0.01) and long ERCP procedure time (OR3.33, p < 0.01/OR3.90, p < 0.01) were related to the high radiation dose exposure. SPA was independently associated with shorter ERCP procedure time (OR 0.63; p = 0.04). Conclusions SPA and low‐dose provided comparable reductions in radiation exposure. Furthermore, SPA was associated with a shorter ERCP time than a low dose, possibly because of improved image quality. However, until fluoroscopic systems are upgraded to incorporate SPA, low‐dose may provide clinical utility for reducing radiation exposure. Trial Registration ClinicalTrials.gov identifier: UMIN000061448

Journal of Hepato-Biliary-Pancreatic Sciences
Matsunami General Hospital (JP), Gifu University Hospital (JP), Gifu Prefectural General Medical Center (JP)
Good health and well-being
Openalex Percentile: Top 11%
Gallbladder and Bile Duct Disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.