The Effects of Hollow Fiber Membrane Configuration on Hemodynamic Characteristics, Oxygen Transfer Performance and Thrombosis Risk in Oxygenators: A Numerical Research

ABSTRACT Background The effects of hollow fiber membrane (HFM) configuration on hemodynamic characteristics, oxygen transfer performance, and thrombosis risk in oxygenators are not clear, and the present study was accomplished to investigate these effects. Methods Three micro‐scale 3D HFM array models were established, with the staggered angle between HFM layers ( Φ ), the spacing between HFMs ( d ), and the number of HFM layers ( n ) as variables. Computational fluid dynamics (CFD) was employed as a predictive tool to quantify wall shear stress (WSS), oxygen partial pressure (PO 2 ), and saturation (SO 2 ), activated coagulation factor XII concentration (C[FXIIa]), blood residence time (BRT), and pressure drop (PD) under various flow rates. Results It was found that larger Φ resulted in higher outlet‐averaged PO 2 /SO 2 , more concentrated BRT and C[FXIIa] at lower values, and higher PD. Similar trends in oxygen transfer were observed when d was decreased or n was increased. However, those conditions were associated with more extensive high‐BRT and high‐C[FXIIa] regions, which were interpreted as indicating higher thrombosis risk. Conclusions Based on these predictive results, it is suggested that the inter‐layer cross angle should be increased within an acceptable PD range, and that the model porosity should be maintained between 0.4 and 0.6. It is also recommended that the number of layers be reduced for a given volume. Furthermore, it was indicated by the simulations that avoiding extremely large or small flow paths is critical, since larger spacings were predicted to markedly reduce oxygen transfer, while smaller spacings were predicted to increase thrombogenic potential. This study can provide guidance for the design optimization of configurations such as the arrangement of HFMs within the oxygenator.

Authors

Institutions

Publication Details

Journal
Artificial Organs
Published
2026-09-07
DOI
https://doi.org/10.1111/aor.70231
Primary Topic
Hemoglobin structure and function
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The Effects of Hollow Fiber Membrane Configuration on Hemodynamic Characteristics, Oxygen Transfer Performance and Thrombosis Risk in Oxygenators: A Numerical Research

Feilong Hei, Anqiang Sun, Xingji Fu, Zengsheng Chen et al.
Artificial Organs
Hemoglobin structure and function
article

The Effects of Hollow Fiber Membrane Configuration on Hemodynamic Characteristics, Oxygen Transfer Performance and Thrombosis Risk in Oxygenators: A Numerical Research

Feilong Hei, Anqiang Sun, Xingji Fu, Zengsheng Chen, Anna Teng, Xiaofang Yang
article en

Abstract

ABSTRACT Background The effects of hollow fiber membrane (HFM) configuration on hemodynamic characteristics, oxygen transfer performance, and thrombosis risk in oxygenators are not clear, and the present study was accomplished to investigate these effects. Methods Three micro‐scale 3D HFM array models were established, with the staggered angle between HFM layers ( Φ ), the spacing between HFMs ( d ), and the number of HFM layers ( n ) as variables. Computational fluid dynamics (CFD) was employed as a predictive tool to quantify wall shear stress (WSS), oxygen partial pressure (PO 2 ), and saturation (SO 2 ), activated coagulation factor XII concentration (C[FXIIa]), blood residence time (BRT), and pressure drop (PD) under various flow rates. Results It was found that larger Φ resulted in higher outlet‐averaged PO 2 /SO 2 , more concentrated BRT and C[FXIIa] at lower values, and higher PD. Similar trends in oxygen transfer were observed when d was decreased or n was increased. However, those conditions were associated with more extensive high‐BRT and high‐C[FXIIa] regions, which were interpreted as indicating higher thrombosis risk. Conclusions Based on these predictive results, it is suggested that the inter‐layer cross angle should be increased within an acceptable PD range, and that the model porosity should be maintained between 0.4 and 0.6. It is also recommended that the number of layers be reduced for a given volume. Furthermore, it was indicated by the simulations that avoiding extremely large or small flow paths is critical, since larger spacings were predicted to markedly reduce oxygen transfer, while smaller spacings were predicted to increase thrombogenic potential. This study can provide guidance for the design optimization of configurations such as the arrangement of HFMs within the oxygenator.

Artificial Organs
Beijing Anzhen Hospital (CN), Beihang University (CN)
National Natural Science Foundation of China, National Key Research and Development Program of China, Fundamental Research Funds for the Central Universities
Clean water and sanitation
Openalex Percentile: Top 14%
Hemoglobin structure and function
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.