Reduction Quality Determines Failure Behavior in Stable Intertrochanteric Fractures: A Mesh-Free Simulation Comparing DHS and TFNA.

Introduction: Stable 2-part intertrochanteric hip fractures (Arbeitsgemeinschaft für Osteosynthesefragen/Orthoapedic Trauma Association [AO/OTA] 31A1.2) are commonly treated with dynamic hip screw (DHS) or trochanteric fixation nail advanced (TFNA). Although nails are increasingly favored, superiority in stable patterns remains unproven with DHS remains widely used. This study applied mesh-free simulation to investigate construct mechanics and failure behavior in stable fracture models, with implant migration and construct stability as surrogate measures for cutout risk. Methods: Particle models of AO/OTA 31A1.2 fractures were obtained from cadaveric computed tomography scans with bone mineral density mapping. Using volumetric particle modelling (100-400 μm resolution), either DHS (316L stainless steel lag screw/barrel) or TFNA (Ti6Al4V extra low interstitial nail/blade) was inserted at a tip-apex distance of <25 mm. A standardized axial compression test approximating single-leg stance was simulated at 2 m/s along the Z-axis. Two implant conditions were tested: sliding (allowing screw/blade migration) and non-sliding (rigid). Two fracture reductions were modelled: anatomical (0 mm gap) and proximal gap (5 mm). Outcomes were maximum load, displacement at failure, load-displacement curves, and von Mises stress distribution. Results: With anatomical reduction, DHS and TFNA showed nearly identical curves up to 3-4 mm displacement; DHS resisted higher loads beyond. In non-sliding, DHS reached a maximum load of 4,066 N vs. 3,945 N for TFNA; in sliding, DHS reached 4,156 N vs. 3,790 N for TFNA. With gap, early performance was similar, but DHS reached higher loads (2,161-2,504 N) vs. TFNA (1,643-1,877 N). Displacement at failure was comparable (approximately 5 mm). DHS exhibited a more gradual postpeak decline, whereas TFNA showed a steeper loss of resistance. Conclusion: DHS and TFNA provide comparable initial stability but diverge with increasing displacement and reduction imperfection. Rather than establishing implant superiority, these findings highlight differences in construct mechanics and tolerance to suboptimal reduction. Clinical Relevance: DHS may better tolerate reduction imperfections in stable intertrochanteric fractures, potentially guiding implant selection when anatomical reduction is challenging. Results should be interpreted as mechanistic insights requiring clinical validation. Level of Evidence: No level of evidence. See Instructions for Authors for a complete description of levels of evidence.

Authors

Institutions

Publication Details

Journal
PubMed
Published
2026-09-05
DOI
https://doi.org/10.2106/jbjs.oa.26.00131
Primary Topic
Hip and Femur Fractures
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Reduction Quality Determines Failure Behavior in Stable Intertrochanteric Fractures: A Mesh-Free Simulation Comparing DHS and TFNA.

Marilyn Janice Oentaryo, Si Jian Hui, Nazrul Nashi, Christian Fang et al.
PubMed
Hip and Femur Fractures
article

Reduction Quality Determines Failure Behavior in Stable Intertrochanteric Fractures: A Mesh-Free Simulation Comparing DHS and TFNA.

Marilyn Janice Oentaryo, Si Jian Hui, Nazrul Nashi, Christian Fang, Erica Ueda Boles, Sloan Kulper
article en

Abstract

Introduction: Stable 2-part intertrochanteric hip fractures (Arbeitsgemeinschaft für Osteosynthesefragen/Orthoapedic Trauma Association [AO/OTA] 31A1.2) are commonly treated with dynamic hip screw (DHS) or trochanteric fixation nail advanced (TFNA). Although nails are increasingly favored, superiority in stable patterns remains unproven with DHS remains widely used. This study applied mesh-free simulation to investigate construct mechanics and failure behavior in stable fracture models, with implant migration and construct stability as surrogate measures for cutout risk. Methods: Particle models of AO/OTA 31A1.2 fractures were obtained from cadaveric computed tomography scans with bone mineral density mapping. Using volumetric particle modelling (100-400 μm resolution), either DHS (316L stainless steel lag screw/barrel) or TFNA (Ti6Al4V extra low interstitial nail/blade) was inserted at a tip-apex distance of <25 mm. A standardized axial compression test approximating single-leg stance was simulated at 2 m/s along the Z-axis. Two implant conditions were tested: sliding (allowing screw/blade migration) and non-sliding (rigid). Two fracture reductions were modelled: anatomical (0 mm gap) and proximal gap (5 mm). Outcomes were maximum load, displacement at failure, load-displacement curves, and von Mises stress distribution. Results: With anatomical reduction, DHS and TFNA showed nearly identical curves up to 3-4 mm displacement; DHS resisted higher loads beyond. In non-sliding, DHS reached a maximum load of 4,066 N vs. 3,945 N for TFNA; in sliding, DHS reached 4,156 N vs. 3,790 N for TFNA. With gap, early performance was similar, but DHS reached higher loads (2,161-2,504 N) vs. TFNA (1,643-1,877 N). Displacement at failure was comparable (approximately 5 mm). DHS exhibited a more gradual postpeak decline, whereas TFNA showed a steeper loss of resistance. Conclusion: DHS and TFNA provide comparable initial stability but diverge with increasing displacement and reduction imperfection. Rather than establishing implant superiority, these findings highlight differences in construct mechanics and tolerance to suboptimal reduction. Clinical Relevance: DHS may better tolerate reduction imperfections in stable intertrochanteric fractures, potentially guiding implant selection when anatomical reduction is challenging. Results should be interpreted as mechanistic insights requiring clinical validation. Level of Evidence: No level of evidence. See Instructions for Authors for a complete description of levels of evidence.

PubMedVol. 11(3)
Hai Kang Life (China) (CN), National University Health System (SG), University of Hong Kong (HK)
Openalex Percentile: Top 8%
Hip and Femur Fractures
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.