Repeatability and reproducibility of corneal endothelial measurements using the topcon SP-1P specular microscope

PURPOSE: To evaluate the repeatability and reproducibility of corneal endothelial parameters measured with the Topcon SP-1P specular microscope in healthy eyes, comparing single and panoramic acquisition modes. METHODS: This reproducibility study included 137 eyes from 79 healthy adults. Endothelial cell density (ECD), average cell area, coefficient of variation (CV), hexagonality, and pachymetry were measured twice by two independent examiners using both single and panoramic modes. Statistical agreement was assessed using intraclass correlation coefficients (ICC), repeatability coefficients (RC), and Bland-Altman analysis. RESULTS: Intra-observer reproducibility for ECD, average cell area, and pachymetry was excellent in both modes (ICC ≥ 0.95). However, the panoramic mode consistently showed improved precision, with higher ICCs and lower RCs for most morphometric variables. Inter-observer agreement was also excellent for ECD (ICC 0.94) and average cell area (ICC 0.95). Significant systematic differences were found between acquisition modalities; specifically, ECD was on average 7.27% lower in single mode compared to panoramic mode (p < 0.001). CONCLUSION: The Topcon SP-1P demonstrates high reproducibility for primary endothelial parameters, particularly when using the panoramic acquisition mode. Since systematic differences exist between modalities, single and panoramic measurements are not directly interchangeable, highlighting the importance of standardizing techniques for longitudinal clinical monitoring.

Authors

Institutions

Publication Details

Journal
Graefe s Archive for Clinical and Experimental Ophthalmology
Published
2026-09-04
DOI
https://doi.org/10.1007/s00417-026-07488-7
Primary Topic
Corneal surgery and disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
article

Repeatability and reproducibility of corneal endothelial measurements using the topcon SP-1P specular microscope

Virgilio Galvis, Alejandro Tello, Juan Urrea, Sylvia J. Villamizar et al.
Graefe s Archive for Clinical and Experimental Ophthalmology
Corneal surgery and disorders
article

Repeatability and reproducibility of corneal endothelial measurements using the topcon SP-1P specular microscope

Virgilio Galvis, Alejandro Tello, Juan Urrea, Sylvia J. Villamizar, Rubén D. Berrospi, Carolina M. Estrella
article en

Abstract

PURPOSE: To evaluate the repeatability and reproducibility of corneal endothelial parameters measured with the Topcon SP-1P specular microscope in healthy eyes, comparing single and panoramic acquisition modes. METHODS: This reproducibility study included 137 eyes from 79 healthy adults. Endothelial cell density (ECD), average cell area, coefficient of variation (CV), hexagonality, and pachymetry were measured twice by two independent examiners using both single and panoramic modes. Statistical agreement was assessed using intraclass correlation coefficients (ICC), repeatability coefficients (RC), and Bland-Altman analysis. RESULTS: Intra-observer reproducibility for ECD, average cell area, and pachymetry was excellent in both modes (ICC ≥ 0.95). However, the panoramic mode consistently showed improved precision, with higher ICCs and lower RCs for most morphometric variables. Inter-observer agreement was also excellent for ECD (ICC 0.94) and average cell area (ICC 0.95). Significant systematic differences were found between acquisition modalities; specifically, ECD was on average 7.27% lower in single mode compared to panoramic mode (p < 0.001). CONCLUSION: The Topcon SP-1P demonstrates high reproducibility for primary endothelial parameters, particularly when using the panoramic acquisition mode. Since systematic differences exist between modalities, single and panoramic measurements are not directly interchangeable, highlighting the importance of standardizing techniques for longitudinal clinical monitoring.

Graefe s Archive for Clinical and Experimental Ophthalmology
Industrial University of Santander (CO), Universidad Autónoma de Bucaramanga (CO), Foscal Hospital (CO)
Openalex Percentile: Top 11%
Corneal surgery and disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.