Effects of count rate performance on image quality in an LAFOV PET/CT system

The development of Long Axial Field-of-View (LAFOV) positron emission tomography (PET) combined with computed tomography (CT) warrants reconsideration of the administered activity required for clinical imaging. This study investigated the relationship between body size, administered activity (18F), and image quality using two patient-like phantoms representing normal (78.5 kg) and obese (194.3 kg) body profiles. The phantoms were scanned in selected five-minute periods within a total duration of 27 and 21 hours, respectively, following the decay of the activity. Analysis of the trues, scatter, and randoms demonstrated a close to linear propagation of the trues with activity up to a point where saturation was reached. The normal body profile phantom achieved higher signal-to-noise ratios (SNRs), reaching system saturation at a lower activity (680 MBq) than the large phantom, representing the obese body profile (970 MBq). The SNR2 showed a nonlinear correlation with the noise equivalent count rate (NECR), whereas SNR2 had a linear correlation with the trues. The study demonstrated the count rate capabilities up to and beyond the saturation limit. Furthermore, we illustrated the possibilities of scanning patients with ultralow dose administration in the range of 0.03-0.07 MBq/kg body weight. However, achieving robust clinical diagnostic quality - particularly for low-contrast lesions - at these activity levels requires further clinical investigation. Utilizing these ultralow activity concentrations requires extending the scan duration to compensate for image quality, especially when examining obese patients.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2026-09-04
DOI
https://doi.org/10.1088/1361-6560/aea301
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effects of count rate performance on image quality in an LAFOV PET/CT system

Flemming Littrup Andersen, Nanna Overbeck, Søren Holm, Thomas Lund Andersen et al.
Physics in Medicine and Biology
Medical Imaging Techniques and Applications
article

Effects of count rate performance on image quality in an LAFOV PET/CT system

Flemming Littrup Andersen, Nanna Overbeck, Søren Holm, Thomas Lund Andersen, Julie Verne Henriksen, Zuzanna Czeluśniak
article en

Abstract

The development of Long Axial Field-of-View (LAFOV) positron emission tomography (PET) combined with computed tomography (CT) warrants reconsideration of the administered activity required for clinical imaging. This study investigated the relationship between body size, administered activity (18F), and image quality using two patient-like phantoms representing normal (78.5 kg) and obese (194.3 kg) body profiles. The phantoms were scanned in selected five-minute periods within a total duration of 27 and 21 hours, respectively, following the decay of the activity. Analysis of the trues, scatter, and randoms demonstrated a close to linear propagation of the trues with activity up to a point where saturation was reached. The normal body profile phantom achieved higher signal-to-noise ratios (SNRs), reaching system saturation at a lower activity (680 MBq) than the large phantom, representing the obese body profile (970 MBq). The SNR2 showed a nonlinear correlation with the noise equivalent count rate (NECR), whereas SNR2 had a linear correlation with the trues. The study demonstrated the count rate capabilities up to and beyond the saturation limit. Furthermore, we illustrated the possibilities of scanning patients with ultralow dose administration in the range of 0.03-0.07 MBq/kg body weight. However, achieving robust clinical diagnostic quality - particularly for low-contrast lesions - at these activity levels requires further clinical investigation. Utilizing these ultralow activity concentrations requires extending the scan duration to compensate for image quality, especially when examining obese patients.

Physics in Medicine and Biology
Copenhagen University Hospital (DK), Rigshospitalet (DK)
Openalex Percentile: Top 11%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.